 "manufacturer": "E+E Elektronik",
 "manufacturer_slug": "e-e-elektronik",
 "model": "EE260 humidity and temperature probe for meteorology (RS-485 Modbus RTU)",
 "model_slug": "ee260",
 "short_model": "EE260",
 "aliases": [
  "EE260",
  "E+E EE260",
  "EE 260"
 ],
 "equipment_type": "Sensor (humidity / temperature probe)",
 "applies_to": "EE260 meteorological RH/T probe with RS-485 Modbus RTU, per Quick Guide QG_EE260 v1.3 (10-2025).",
 "summary": "The EE260 is an M12-connected humidity and temperature probe for weather stations, with two analogue outputs and an RS-485 Modbus RTU port. It publishes every measured and calculated humidity quantity twice: as 32-bit floats and as scaled 16-bit integers, in metric and US units. This map lists both blocks.",
 "transport": [
  "RTU"
 ],
 "defaults": {
  "unit_id": 235,
  "baud": 9600,
  "parity": "Even",
  "data_bits": 8,
  "stop_bits": 1,
  "tcp_port": null
 },
 "addressing": {
  "convention": "1-based",
  "note": "The quick guide prints a decimal register number that starts at 1 and a hex register address that starts at 0 (1003 = 0x3EA). Send the hex address on the wire: register number minus 1."
 },
 "default_word_order": "CDAB",
 "firmware": null,
 "source": {
  "title": "Quick Guide EE260 - Humidity and Temperature Probe for Meteorological Applications",
  "doc_number": "QG_EE260 v1.3 (194928)",
  "revision_date": "2025-10",
  "url": "https://downloads.epluse.com/fileadmin/data/product/ee260/QG_EE260.pdf",
  "pages": "Modbus setup p. 1; register map p. 2 (float and integer tables, footnotes 1-3)"
 },
 "notes": [
  "Float word order comes from E+E application note AN0103 Rev. 1.4 (11/2020), section 7.2, which the quick guide points to: the two byte pairs of a float are swapped (bytes 3-4 first, then 1-2), i.e. low word first (CDAB). The note's write example (1013.25 mbar sent as 49 9A 44 7D) also puts the high word 44 7D second. AN0103: https://downloads.epluse.com/fileadmin/data/product/ee071/AN0103.pdf",
  "Integer scaling: the quick guide's scale column is a divisor (scale 100: a reading of 2550 = 25.5; scale 50: 2550 = 51). The scale field here is the matching multiplier, so the ppm row with divisor 0.1 becomes x10.",
  "Measured values can be read with function 03 or 04 (both are listed); they are filed here as holding registers.",
  "Left out: the slave-address register (0x00) and communication-settings register (0x01). AN0103 says they are write-only and does not fully define the settings byte coding.",
  "Read each float with one request for both registers, so both halves come from the same measurement (AN0103 advice)."
 ],
 "registers": [
  {
   "table": "holding_register",
   "address": 1002,
   "documented": "1003 (0x3EA)",
   "name": "Temperature T (°C)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1004,
   "documented": "1005 (0x3EC)",
   "name": "Temperature T (°F)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1020,
   "documented": "1021 (0x3FC)",
   "name": "Relative humidity RH (%RH)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "%RH",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1100,
   "documented": "1101 (0x44C)",
   "name": "Water vapour partial pressure e (mbar)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "mbar",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1102,
   "documented": "1103 (0x44E)",
   "name": "Water vapour partial pressure e (psi)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "psi",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1104,
   "documented": "1105 (0x450)",
   "name": "Dew point temperature Td (°C)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1106,
   "documented": "1107 (0x452)",
   "name": "Dew point temperature Td (°F)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1108,
   "documented": "1109 (0x454)",
   "name": "Wet bulb temperature Tw (°C)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1110,
   "documented": "1111 (0x456)",
   "name": "Wet bulb temperature Tw (°F)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1112,
   "documented": "1113 (0x458)",
   "name": "Absolute humidity dv (g/m3)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "g/m3",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1114,
   "documented": "1115 (0x45A)",
   "name": "Absolute humidity dv (gr/ft3)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "gr/ft3",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1120,
   "documented": "1121 (0x460)",
   "name": "Mixing ratio r (g/kg)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "g/kg",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1122,
   "documented": "1123 (0x462)",
   "name": "Mixing ratio r (gr/lb)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "gr/lb",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1124,
   "documented": "1125 (0x464)",
   "name": "Specific enthalpy h (kJ/kg)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "kJ/kg",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1128,
   "documented": "1129 (0x468)",
   "name": "Specific enthalpy h (BTU/lb)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "BTU/lb",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1130,
   "documented": "1131 (0x46A)",
   "name": "Frost point temperature Tf (°C)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1132,
   "documented": "1133 (0x46C)",
   "name": "Frost point temperature Tf (°F)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1150,
   "documented": "1151 (0x47E)",
   "name": "Volume concentration Wv (ppm)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "ppm",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1220,
   "documented": "1221 (0x4C4)",
   "name": "Saturation vapour pressure over water ew (mbar)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "mbar",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1222,
   "documented": "1223 (0x4C6)",
   "name": "Saturation vapour pressure over water ew (psi)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "psi",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1224,
   "documented": "1225 (0x4C8)",
   "name": "Saturation vapour pressure over ice ei (mbar)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "mbar",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1226,
   "documented": "1227 (0x4CA)",
   "name": "Saturation vapour pressure over ice ei (psi)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "psi",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1236,
   "documented": "1237 (0x4D4)",
   "name": "Ice bulb temperature Ti (°C)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1238,
   "documented": "1239 (0x4D6)",
   "name": "Ice bulb temperature Ti (°F)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1246,
   "documented": "1247 (0x4DE)",
   "name": "Specific humidity qv (g/kg)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "g/kg",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1248,
   "documented": "1249 (0x4E0)",
   "name": "Specific humidity qv (gr/lb)",
   "description": "IEEE 754 float.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "gr/lb",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4001,
   "documented": "4002 (0xFA1)",
   "name": "Temperature T (°C, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4002,
   "documented": "4003 (0xFA2)",
   "name": "Temperature T (°F, integer)",
   "description": "Raw value divided by 50 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.02,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4010,
   "documented": "4011 (0xFAA)",
   "name": "Relative humidity RH (%RH, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "%RH",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4050,
   "documented": "4051 (0xFD2)",
   "name": "Water vapour partial pressure e (mbar, integer)",
   "description": "Raw value divided by 10 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.1,
   "units": "mbar",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4051,
   "documented": "4052 (0xFD3)",
   "name": "Water vapour partial pressure e (psi, integer)",
   "description": "Raw value divided by 1000 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.001,
   "units": "psi",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4052,
   "documented": "4053 (0xFD4)",
   "name": "Dew point temperature Td (°C, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4053,
   "documented": "4054 (0xFD5)",
   "name": "Dew point temperature Td (°F, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4054,
   "documented": "4055 (0xFD6)",
   "name": "Wet bulb temperature Tw (°C, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4055,
   "documented": "4056 (0xFD7)",
   "name": "Wet bulb temperature Tw (°F, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4056,
   "documented": "4057 (0xFD8)",
   "name": "Absolute humidity dv (g/m3, integer)",
   "description": "Raw value divided by 10 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.1,
   "units": "g/m3",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4057,
   "documented": "4058 (0xFD9)",
   "name": "Absolute humidity dv (gr/ft3, integer)",
   "description": "Raw value divided by 10 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.1,
   "units": "gr/ft3",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4060,
   "documented": "4061 (0xFDC)",
   "name": "Mixing ratio r (g/kg, integer)",
   "description": "Raw value divided by 10 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.1,
   "units": "g/kg",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4061,
   "documented": "4062 (0xFDD)",
   "name": "Mixing ratio r (gr/lb, integer)",
   "description": "Raw value divided by 10 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.1,
   "units": "gr/lb",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4062,
   "documented": "4063 (0xFDE)",
   "name": "Specific enthalpy h (kJ/kg, integer)",
   "description": "Raw value divided by 1 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "kJ/kg",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4064,
   "documented": "4065 (0xFE0)",
   "name": "Specific enthalpy h (BTU/lb, integer)",
   "description": "Raw value divided by 1 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "BTU/lb",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4065,
   "documented": "4066 (0xFE1)",
   "name": "Frost point temperature Tf (°C, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4066,
   "documented": "4067 (0xFE2)",
   "name": "Frost point temperature Tf (°F, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4075,
   "documented": "4076 (0xFEB)",
   "name": "Volume concentration Wv (ppm, integer)",
   "description": "Raw value divided by 0.1 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 10,
   "units": "ppm",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4110,
   "documented": "4111 (0x100E)",
   "name": "Saturation vapour pressure over water ew (mbar, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "mbar",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4111,
   "documented": "4112 (0x100F)",
   "name": "Saturation vapour pressure over water ew (psi, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "psi",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4112,
   "documented": "4113 (0x1010)",
   "name": "Saturation vapour pressure over ice ei (mbar, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "mbar",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4113,
   "documented": "4114 (0x1011)",
   "name": "Saturation vapour pressure over ice ei (psi, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "psi",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4118,
   "documented": "4119 (0x1016)",
   "name": "Ice bulb temperature Ti (°C, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4119,
   "documented": "4120 (0x1017)",
   "name": "Ice bulb temperature Ti (°F, integer)",
   "description": "Raw value divided by 100 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.01,
   "units": "°F",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4123,
   "documented": "4124 (0x101B)",
   "name": "Specific humidity qv (g/kg, integer)",
   "description": "Raw value divided by 10 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.1,
   "units": "g/kg",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4124,
   "documented": "4125 (0x101C)",
   "name": "Specific humidity qv (gr/lb, integer)",
   "description": "Raw value divided by 10 (scale column).",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.1,
   "units": "gr/lb",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  }
 ],
 "integration": {
  "summary": "Connect at 9600 baud, 8E1, address 235, and read two registers from 0x3EA for temperature (float, low word first) and from 0x3FC for relative humidity. If the master cannot swap words, read the scaled integers at 0xFA1 (T x100) and 0xFAA (RH x100).",
  "minimum_registers": [
   {
    "table": "holding_register",
    "address": 1002,
    "why": "Temperature, float °C"
   },
   {
    "table": "holding_register",
    "address": 1020,
    "why": "Relative humidity, float %RH"
   },
   {
    "table": "holding_register",
    "address": 1104,
    "why": "Dew point, float °C"
   },
   {
    "table": "holding_register",
    "address": 4001,
    "why": "Temperature, integer °C x100"
   },
   {
    "table": "holding_register",
    "address": 4010,
    "why": "Relative humidity, integer x100"
   }
  ],
  "enable": "Address, baud rate, parity and stop bits are set with the free PCS10 software through the HA011018 adapter and HA011020 cable.",
  "watchdog": null,
  "quirks": [
   "Floats are low word first (CDAB).",
   "Integer °F temperature uses divisor 50, not 100.",
   "Register numbers in the guide are 1-based; subtract 1."
  ],
  "setup": [
   "Factory settings: 9600 baud, 8 data bits, even parity, 1 stop bit, address 235.",
   "Selectable: 9600-115200 baud, none/odd/even parity, 1 or 2 stop bits, address 1-247.",
   "Supply 7-30 V DC; RS-485 A (D+) is pin 6, B (D-) is pin 2 on the M12 plug."
  ],
  "gotchas": [
   "The factory address is 235, not 1.",
   "Use a shielded cable to stay inside the EMC figures."
  ]
 },
 "page": "https://easymodbus.com/modbus-register-maps/e-e-elektronik/ee260",
 "address_note": "address = 0-based Modbus wire (PDU) address; documented = as printed in the source.",
 "note": "Re-typed from the manufacturer's public documentation (see source) and verified against it. Registers vary by firmware and options; verify against the live device before writing."
}