 "manufacturer": "Endress+Hauser",
 "manufacturer_slug": "endress-hauser",
 "model": "Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP)",
 "model_slug": "liquiline-cm44x",
 "short_model": "Liquiline CM44x",
 "aliases": [
  "CM442",
  "CM444",
  "CM448",
  "CM44P",
  "CM44R",
  "Liquiline CM442",
  "Liquiline CM444",
  "Liquiline CM448"
 ],
 "equipment_type": "Sensor (water analysis transmitter)",
 "applies_to": "Liquiline CM44x with the Modbus RS485 or Modbus TCP option, software 1.11.00 and later (SD01189C/07/EN/07.22). The CSFxx sampler and CA80xx analyzer blocks of the same guide are not included.",
 "summary": "The Liquiline CM44x is Endress+Hauser's multichannel transmitter for Memosens pH, ORP, conductivity, oxygen, turbidity and similar sensors. With the Modbus option it exposes 16 configurable analog and 8 digital device variables, 4 analog and 8 digital write-back variables, diagnostics and per-sensor information blocks. This map covers those common blocks, with the sensor block shown for sensor 1.",
 "transport": [
  "RTU",
  "ASCII",
  "TCP"
 ],
 "defaults": {
  "unit_id": 247,
  "baud": 19200,
  "parity": "Even",
  "data_bits": 8,
  "stop_bits": 1,
  "tcp_port": 502
 },
 "addressing": {
  "convention": "0-based",
  "note": "The guide states that its register numbers start at 0 exactly as sent on the wire, and that Modicon-style PLCs must add 1. So AI 1 value (register 0) is 40001 in Modicon terms."
 },
 "default_word_order": "CDAB",
 "firmware": "Software 1.11.00 and later",
 "source": {
  "title": "Liquiline CM44x / Liquistation CSFxx / Liquiline System CA80xx - Data transmission via Modbus (Special Documentation)",
  "doc_number": "SD01189C/07/EN/07.22",
  "revision_date": "2022-07",
  "url": "https://bdih-download.endress.com/files/DLA/005056A500261EECB9FE4A363A4B4FD5/SD01189CEN_07.22.pdf",
  "pages": "p. 8 (numbering), pp. 9-11 (byte order, data types, device-variable layout and status), pp. 16-18 (settings), pp. 26-27 (FC08 bits, NE 107), pp. 44-61 (register tables 7.3.2.1-7.3.2.5)"
 },
 "notes": [
  "Word order depends on the device's Byte order setting. At the factory default 1-0-3-2 a FLOAT is sent mantissa-low word first and sign/exponent word second, i.e. CDAB, which is what this file uses.",
  "The guide says every data type longer than 2 bytes other than FLOAT is sent like a string, most significant byte first at the default setting, so UNSIGNED32 values are listed as ABCD.",
  "UNSIGNED8 values occupy one full register.",
  "The AI/DI device variables carry whatever source is assigned to them in the transmitter menu (Setup/Outputs/Modbus); their unit code (table 7.1) tells what the value is. Units are therefore blank here.",
  "Status bytes: 0 good, 1 uncertain, 2 bad, 3 not assigned. AO and DO values written by the PLC are used only when their status is written as 0.",
  "Sensor blocks repeat per channel: sensor N uses registers N000-N999 (sensor 1 = 1000-1999). Only sensor 1 is listed; add (N-1) x 1000 for the others.",
  "DATETIME fields (7 bytes in 4 registers, packed date/time) are left out because the format is a packed structure, not a plain number.",
  "Current output values 96-110 are labelled in ampere in the table although the names say mA; check scaling against the analog output on site.",
  "A parameter has to be read or written in full; reading half of a FLOAT or string returns a slave-device-failure exception.",
  "The guide says function codes 03 and 04 return the same data, but it also gives separate input and holding tables that reuse the same numbers: diagnostics (input) and configuration (holding) both start at 500, and the sensor block (input) and sensor setup (holding) both start at X000. Because the guide does not settle which content a read returns at those numbers, every register that sits on such a clash is left out: the whole configuration block 500-736 (communication settings, byte order, watchdog, device tag, PID setpoints) and input registers 500-521, 594-625, 690-737 and sensor X000-X018.",
  "Timestamps (UNSIGNED32) count seconds since 1 January 1970."
 ],
 "registers": [
  {
   "table": "input_register",
   "address": 522,
   "documented": "522",
   "name": "MODBUS_FRAMING_ERRORS",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 524,
   "documented": "524",
   "name": "MODBUS_CRC_ERRORS",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 526,
   "documented": "526",
   "name": "MODBUS_PARITY_ERRORS",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 528,
   "documented": "528",
   "name": "MODBUS_01_EXCEPTIONS",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 530,
   "documented": "530",
   "name": "MODBUS_02_EXCEPTIONS",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 532,
   "documented": "532",
   "name": "MODBUS_03_EXCEPTIONS",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 534,
   "documented": "534",
   "name": "MODBUS_04_EXCEPTIONS",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 536,
   "documented": "536",
   "name": "ETHERNET_RECEIVED_FRAMES",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 538,
   "documented": "538",
   "name": "ETHERNET_SEND_FRAMES",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 540,
   "documented": "540",
   "name": "ETHERNET_VALID_FRAMES",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 542,
   "documented": "542",
   "name": "ETHERNET_INVALID_FRAMES",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 544,
   "documented": "544",
   "name": "ETHERNET_CRC_ERRORS",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 546,
   "documented": "546",
   "name": "DEVICE_SW_VERSION",
   "description": "16 characters.",
   "data_type": "STRING",
   "length": 8,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 554,
   "documented": "554",
   "name": "DEVICE_BOOTLOADER_VERSION",
   "description": "16 characters.",
   "data_type": "STRING",
   "length": 8,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 562,
   "documented": "562",
   "name": "DEVICE_SERIAL",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 578,
   "documented": "578",
   "name": "DISPLAY_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 626,
   "documented": "626",
   "name": "BASE_MODUL_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 642,
   "documented": "642",
   "name": "EXTENSION_MODUL_1_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 658,
   "documented": "658",
   "name": "EXTENSION_MODUL_2_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 674,
   "documented": "674",
   "name": "EXTENSION_MODUL_3_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 738,
   "documented": "738",
   "name": "EXTENSION_MODUL_7_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 754,
   "documented": "754",
   "name": "EXTENSION_MODUL_8_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 770,
   "documented": "770",
   "name": "SENSOR_1_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 786,
   "documented": "786",
   "name": "SENSOR_2_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 802,
   "documented": "802",
   "name": "SENSOR_3_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 818,
   "documented": "818",
   "name": "SENSOR_4_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 834,
   "documented": "834",
   "name": "SENSOR_5_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 850,
   "documented": "850",
   "name": "SENSOR_6_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 866,
   "documented": "866",
   "name": "SENSOR_7_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 882,
   "documented": "882",
   "name": "SENSOR_8_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 898,
   "documented": "898",
   "name": "FMSY1_MODUL_SER_NUM",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 941,
   "documented": "941",
   "name": "RS485_TERMINATION_STATE",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Off",
    "1": "On"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 942,
   "documented": "942",
   "name": "ETHERNET_MAC",
   "description": "6 characters.",
   "data_type": "STRING",
   "length": 3,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 945,
   "documented": "945",
   "name": "CLEANINGACTIONSTATE_1",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Off",
    "1": "Waiting",
    "2": "Waiting",
    "3": "Pre-rinse",
    "4": "Cleaning",
    "5": "Postrinsing",
    "6": "Failsafe",
    "7": "Manual mode",
    "8": "Waiting for execution"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 946,
   "documented": "946",
   "name": "CLEANINGACTIONSTATE_2",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Off",
    "1": "Waiting",
    "2": "Waiting",
    "3": "Pre-rinse",
    "4": "Cleaning",
    "5": "Postrinsing",
    "6": "Failsafe",
    "7": "Manual mode",
    "8": "Waiting for execution"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 947,
   "documented": "947",
   "name": "CLEANINGACTIONSTATE_3",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Off",
    "1": "Waiting",
    "2": "Waiting",
    "3": "Pre-rinse",
    "4": "Cleaning",
    "5": "Postrinsing",
    "6": "Failsafe",
    "7": "Manual mode",
    "8": "Waiting for execution"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 948,
   "documented": "948",
   "name": "CLEANINGACTIONSTATE_4",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Off",
    "1": "Waiting",
    "2": "Waiting",
    "3": "Pre-rinse",
    "4": "Cleaning",
    "5": "Postrinsing",
    "6": "Failsafe",
    "7": "Manual mode",
    "8": "Waiting for execution"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 953,
   "documented": "953",
   "name": "HEARTBEAT_DEVICE_STATUS",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 955,
   "documented": "955",
   "name": "HEARTBEAT_DEVICE_HEALTH_CONDITION",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "%",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 957,
   "documented": "957",
   "name": "MAINTENANCE_TIMER_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 959,
   "documented": "959",
   "name": "HEARTBEAT_INDEX_DEVICE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 965,
   "documented": "965",
   "name": "KEY_PERF_INDICATOR_DEVICE_AVAILABILITY",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "%",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 975,
   "documented": "975",
   "name": "KEY_PERF_INDICATOR_DEVICE_NUMBER_OF_FAILURES",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1019,
   "documented": "X019 (sensor 1)",
   "name": "SENSOR_SERIAL_CH0",
   "description": "Sensor 1 block. 16 characters.",
   "data_type": "STRING",
   "length": 8,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1027,
   "documented": "X027 (sensor 1)",
   "name": "SENSOR_HW_VERSION_CH0",
   "description": "Sensor 1 block. 16 characters.",
   "data_type": "STRING",
   "length": 8,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1035,
   "documented": "X035 (sensor 1)",
   "name": "SENSOR_SW_VERSION_CH0",
   "description": "Sensor 1 block. 16 characters.",
   "data_type": "STRING",
   "length": 8,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1043,
   "documented": "X043 (sensor 1)",
   "name": "SENSOR_MANUFACTURER_CH0",
   "description": "Sensor 1 block. 32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1059,
   "documented": "X059 (sensor 1)",
   "name": "SENSOR_ORDERCODE_CH0",
   "description": "Sensor 1 block. 32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1083,
   "documented": "X083 (sensor 1)",
   "name": "SENSOR_EXTREME_SPECIFIC_MIN_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1085,
   "documented": "X085 (sensor 1)",
   "name": "SENSOR_EXTREME_TEMP_MIN_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1087,
   "documented": "X087 (sensor 1)",
   "name": "SENSOR_EXTREME_SPECIFIC_MAX_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1089,
   "documented": "X089 (sensor 1)",
   "name": "SENSOR_EXTREME_TEMP_MAX_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1091,
   "documented": "X091 (sensor 1)",
   "name": "SENSOR_CURRENT_DIAG_SOURCETYPE_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "General",
    "1": "Temperature Input",
    "2": "Binary Input",
    "3": "Binary Output",
    "4": "Current Input",
    "5": "Current Output",
    "6": "Relay",
    "7": "Memosens Sensor",
    "8": "PID Controller",
    "9": "Cleaning",
    "10": "Limit switch",
    "11": "Binary formula result",
    "12": "Binary Input Module",
    "13": "Binary Output Module",
    "14": "Sample point"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1092,
   "documented": "X092 (sensor 1)",
   "name": "SENSOR_CURRENT_DIAG_MODUL_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Slot 1",
    "1": "Slot 1",
    "2": "Slot 2",
    "3": "Slot 3",
    "4": "Slot 4",
    "5": "Slot 5",
    "6": "Slot 6",
    "7": "Slot 7",
    "8": "Sampler",
    "9": "Software",
    "10": "Analyzer"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1093,
   "documented": "X093 (sensor 1)",
   "name": "SENSOR_CURRENT_DIAG_PORT_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1094,
   "documented": "X094 (sensor 1)",
   "name": "SENSOR_CURRENT_DIAG_NE107_STATUS_CH0",
   "description": "Sensor 1 block. NAMUR NE 107 category bits.",
   "data_type": "BITMAP16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": {
    "0": "Failure (F)",
    "1": "Function check (C)",
    "2": "Maintenance required (M)",
    "3": "Out of specification (S)"
   }
  },
  {
   "table": "input_register",
   "address": 1095,
   "documented": "X095 (sensor 1)",
   "name": "SENSOR_CURRENT_DIAG_CODE_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1096,
   "documented": "X096 (sensor 1)",
   "name": "SENSOR_CURRENT_DIAG_TIMESTAMP_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1098,
   "documented": "X098 (sensor 1)",
   "name": "SENSOR_NUMBER_ADDITIONAL_DIAG_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1099,
   "documented": "X099 (sensor 1)",
   "name": "SENSOR_PAST_DIAG_SOURCETYPE_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "General",
    "1": "Temperature Input",
    "2": "Binary Input",
    "3": "Binary Output",
    "4": "Current Input",
    "5": "Current Output",
    "6": "Relay",
    "7": "Memosens Sensor",
    "8": "PID Controller",
    "9": "Cleaning",
    "10": "Limit switch",
    "11": "Binary formula result",
    "12": "Binary Input Module",
    "13": "Binary Output Module",
    "14": "Sample preparation"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1100,
   "documented": "X100 (sensor 1)",
   "name": "SENSOR_PAST_DIAG_MODUL_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Slot 1",
    "1": "Slot 1",
    "2": "Slot 2",
    "3": "Slot 3",
    "4": "Slot 4",
    "5": "Slot 5",
    "6": "Slot 6",
    "7": "Slot 7",
    "8": "Sampler",
    "9": "Software",
    "10": "Analyzer"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1101,
   "documented": "X101 (sensor 1)",
   "name": "SENSOR_PAST_DIAG_PORT_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1102,
   "documented": "X102 (sensor 1)",
   "name": "SENSOR_PAST_DIAG_NE107_STATUS_CH0",
   "description": "Sensor 1 block. NAMUR NE 107 category bits.",
   "data_type": "BITMAP16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": {
    "0": "Failure (F)",
    "1": "Function check (C)",
    "2": "Maintenance required (M)",
    "3": "Out of specification (S)"
   }
  },
  {
   "table": "input_register",
   "address": 1103,
   "documented": "X103 (sensor 1)",
   "name": "SENSOR_PAST_DIAG_CODE_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1104,
   "documented": "X104 (sensor 1)",
   "name": "SENSOR_PAST_DIAG_TIMESTAMP_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1106,
   "documented": "X106 (sensor 1)",
   "name": "OP_TIME_TOTAL_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "s",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1108,
   "documented": "X108 (sensor 1)",
   "name": "OP_TIME_ABOVE_SPEC_TEMP_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "s",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1110,
   "documented": "X110 (sensor 1)",
   "name": "OP_TIME_BELOW_SPEC_TEMP_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "s",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1112,
   "documented": "X112 (sensor 1)",
   "name": "OP_TIME_LAMP_LIFE_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1114,
   "documented": "X114 (sensor 1)",
   "name": "OP_TIME_STERILISATIONS_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1115,
   "documented": "X115 (sensor 1)",
   "name": "OP_TIME_CIP_CYCLES_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1116,
   "documented": "X116 (sensor 1)",
   "name": "OP_TIME_CAP_CALIBRATIONS_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1117,
   "documented": "X117 (sensor 1)",
   "name": "OP_TIME_CAP_CALIB_TIMER_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "s",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1119,
   "documented": "X119 (sensor 1)",
   "name": "OP_TIME_CAP_STERILISATIONS_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1120,
   "documented": "X120 (sensor 1)",
   "name": "OP_TIME_CAP_FILTER_CHANGED_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "s",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1122,
   "documented": "X122 (sensor 1)",
   "name": "CAL_COUNT_0_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1123,
   "documented": "X123 (sensor 1)",
   "name": "CAL_COUNT_1_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1124,
   "documented": "X124 (sensor 1)",
   "name": "CAL_COUNT_2_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1125,
   "documented": "X125 (sensor 1)",
   "name": "CAL_COUNT_3_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1142,
   "documented": "X142 (sensor 1)",
   "name": "CAL_METHOD_0_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "None",
    "1": "Numeric input",
    "2": "1-point calibration",
    "3": "2-point calibration",
    "4": "Multipoint Calibration",
    "5": "Table",
    "6": "Grab sample",
    "7": "Zero point",
    "8": "Air 100%% rh",
    "9": "H2O air-saturated",
    "10": "Air variable",
    "11": "Zero point sample",
    "12": "Slope sample",
    "13": "Offset",
    "14": "Slope",
    "15": "with temp. comp",
    "16": "without temp. comp",
    "17": "Standard addition",
    "18": "Numeric input",
    "19": "Numeric input zero point",
    "20": "Numeric input",
    "21": "Factory calib.",
    "22": "Test gas calibration",
    "23": "Value acquisition",
    "24": "3-point calibration",
    "25": "Ext. Reference",
    "26": "Int. Reference",
    "27": "in water",
    "28": "in air"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1143,
   "documented": "X143 (sensor 1)",
   "name": "CAL_METHOD_1_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "None",
    "1": "Numeric input",
    "2": "1-point calibration",
    "3": "2-point calibration",
    "4": "Multipoint Calibration",
    "5": "Table",
    "6": "Grab sample",
    "7": "Zero point",
    "8": "Air 100%% rh",
    "9": "H2O air-saturated",
    "10": "Air variable",
    "11": "Zero point sample",
    "12": "Slope sample",
    "13": "Offset",
    "14": "Slope",
    "15": "with temp. comp",
    "16": "without temp. comp",
    "17": "Standard addition",
    "18": "Numeric input",
    "19": "Numeric input zero point",
    "20": "Numeric input",
    "21": "Factory calib.",
    "22": "Test gas calibration",
    "23": "Value acquisition",
    "24": "3-point calibration",
    "25": "Ext. Reference",
    "26": "Int. Reference",
    "27": "in water",
    "28": "in air"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1144,
   "documented": "X144 (sensor 1)",
   "name": "CAL_METHOD_2_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "None",
    "1": "Numeric input",
    "2": "1-point calibration",
    "3": "2-point calibration",
    "4": "Multipoint Calibration",
    "5": "Table",
    "6": "Grab sample",
    "7": "Zero point",
    "8": "Air 100%% rh",
    "9": "H2O air-saturated",
    "10": "Air variable",
    "11": "Zero point sample",
    "12": "Slope sample",
    "13": "Offset",
    "14": "Slope",
    "15": "with temp. comp",
    "16": "without temp. comp",
    "17": "Standard addition",
    "18": "Numeric input",
    "19": "Numeric input zero point",
    "20": "Numeric input",
    "21": "Factory calib.",
    "22": "Test gas calibration",
    "23": "Value acquisition",
    "24": "3-point calibration",
    "25": "Ext. Reference",
    "26": "Int. Reference",
    "27": "in water",
    "28": "in air"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1145,
   "documented": "X145 (sensor 1)",
   "name": "CAL_METHOD_3_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "None",
    "1": "Numeric input",
    "2": "1-point calibration",
    "3": "2-point calibration",
    "4": "Multipoint Calibration",
    "5": "Table",
    "6": "Grab sample",
    "7": "Zero point",
    "8": "Air 100%% rh",
    "9": "H2O air-saturated",
    "10": "Air variable",
    "11": "Zero point sample",
    "12": "Slope sample",
    "13": "Offset",
    "14": "Slope",
    "15": "with temp. comp",
    "16": "without temp. comp",
    "17": "Standard addition",
    "18": "Numeric input",
    "19": "Numeric input zero point",
    "20": "Numeric input",
    "21": "Factory calib.",
    "22": "Test gas calibration",
    "23": "Value acquisition",
    "24": "3-point calibration",
    "25": "Ext. Reference",
    "26": "Int. Reference",
    "27": "in water",
    "28": "in air"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1146,
   "documented": "X146 (sensor 1)",
   "name": "CAL_COUNT_ZERO_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1151,
   "documented": "X151 (sensor 1)",
   "name": "CAL_METHOD_ZERO_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "None",
    "1": "Numeric input",
    "2": "1-point calibration",
    "3": "2-point calibration",
    "4": "Multipoint Calibration",
    "5": "Table",
    "6": "Grab sample",
    "7": "Zero point",
    "8": "Air 100%% rh",
    "9": "H2O air-saturated",
    "10": "Air variable",
    "11": "Zero point sample",
    "12": "Slope sample",
    "13": "Offset",
    "14": "Slope",
    "15": "with temp. comp",
    "16": "without temp. comp",
    "17": "Standard addition",
    "18": "Numeric input",
    "19": "Numeric input zero point",
    "20": "Numeric input 2-point",
    "21": "Factory calib.",
    "22": "Test gas calibration",
    "23": "Value acquisition",
    "24": "3-point calibration",
    "25": "Ext. Reference",
    "26": "Int. Reference",
    "27": "in water",
    "28": "in air"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1152,
   "documented": "X152 (sensor 1)",
   "name": "CAL_COUNT_TEMP_CH0",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1157,
   "documented": "X157 (sensor 1)",
   "name": "CAL_METHOD_TEMP_CH0",
   "description": "Sensor 1 block. 8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "None",
    "1": "Numeric input",
    "2": "1-point calibration",
    "3": "2-point calibration",
    "4": "Multipoint Calibration",
    "5": "Table",
    "6": "Grab sample",
    "7": "Zero point",
    "8": "Air 100%% rh",
    "9": "H2O air-saturated",
    "10": "Air variable",
    "11": "Zero point sample",
    "12": "Slope sample",
    "13": "Offset",
    "14": "Slope",
    "15": "with temp. comp",
    "16": "without temp. comp",
    "17": "Standard addition",
    "18": "Numeric input",
    "19": "Numeric input zero point",
    "20": "Numeric input",
    "21": "Factory calib.",
    "22": "Test gas calibration",
    "23": "Value acquisition",
    "24": "3-point calibration",
    "25": "Ext. Reference",
    "26": "Int. Reference",
    "27": "in water",
    "28": "in air"
   },
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1158,
   "documented": "X158 (sensor 1)",
   "name": "HEARTBEAT_SENSOR_STATUS_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "%",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1160,
   "documented": "X160 (sensor 1)",
   "name": "HEARTBEAT_SENSOR_HEALTH_CONDITION_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "%",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1162,
   "documented": "X162 (sensor 1)",
   "name": "MAINTENANCE_TIMER_VALUE_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "%",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1164,
   "documented": "X164 (sensor 1)",
   "name": "HEARTBEAT_INDEX_SENSOR_CH0",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "%",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1166,
   "documented": "X166 (sensor 1)",
   "name": "KEY_PERF_INDICATOR_SENSOR_AVAILABILITY_CH1",
   "description": "Sensor 1 block.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "%",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1176,
   "documented": "X176 (sensor 1)",
   "name": "KEY_PERF_INDICATOR_SENSOR_NUMBER_OF_FAILURES_CH1",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 1189,
   "documented": "X189 (sensor 1)",
   "name": "KEY_PERF_INDICATOR_SENSOR_NUMBER_OF_CALIBRATIONS_CH1",
   "description": "Sensor 1 block.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 9000,
   "documented": "9000",
   "name": "BUS_ADDRESS",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 9001,
   "documented": "9001",
   "name": "DEVICE_TAG",
   "description": "32 characters.",
   "data_type": "STRING",
   "length": 16,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "input_register",
   "address": 9017,
   "documented": "9017",
   "name": "DEVICE_ID",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Liquiline CM442",
    "1": "Liquiline CM448",
    "2": "Liquiport CSP44",
    "3": "Liquistation CSF48",
    "4": "Liquistation CSF22",
    "5": "Liquistation CSF33",
    "6": "Liquiline System CA80",
    "7": "Liquistation CSF34",
    "8": "Liquistation CSF39",
    "9": "Liquiline CM44P"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 0,
   "documented": "0",
   "name": "MODBUS_AI_1_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 2,
   "documented": "2",
   "name": "MODBUS_AI_1_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 3,
   "documented": "3",
   "name": "MODBUS_AI_1_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 4,
   "documented": "4",
   "name": "MODBUS_AI_2_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 6,
   "documented": "6",
   "name": "MODBUS_AI_2_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 7,
   "documented": "7",
   "name": "MODBUS_AI_2_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 8,
   "documented": "8",
   "name": "MODBUS_AI_3_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 10,
   "documented": "10",
   "name": "MODBUS_AI_3_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 11,
   "documented": "11",
   "name": "MODBUS_AI_3_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 12,
   "documented": "12",
   "name": "MODBUS_AI_4_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 14,
   "documented": "14",
   "name": "MODBUS_AI_4_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 15,
   "documented": "15",
   "name": "MODBUS_AI_4_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 16,
   "documented": "16",
   "name": "MODBUS_AI_5_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 18,
   "documented": "18",
   "name": "MODBUS_AI_5_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 19,
   "documented": "19",
   "name": "MODBUS_AI_5_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 20,
   "documented": "20",
   "name": "MODBUS_AI_6_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 22,
   "documented": "22",
   "name": "MODBUS_AI_6_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 23,
   "documented": "23",
   "name": "MODBUS_AI_6_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 24,
   "documented": "24",
   "name": "MODBUS_AI_7_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 26,
   "documented": "26",
   "name": "MODBUS_AI_7_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 27,
   "documented": "27",
   "name": "MODBUS_AI_7_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 28,
   "documented": "28",
   "name": "MODBUS_AI_8_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 30,
   "documented": "30",
   "name": "MODBUS_AI_8_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 31,
   "documented": "31",
   "name": "MODBUS_AI_8_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 32,
   "documented": "32",
   "name": "MODBUS_AI_9_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 34,
   "documented": "34",
   "name": "MODBUS_AI_9_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 35,
   "documented": "35",
   "name": "MODBUS_AI_9_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 36,
   "documented": "36",
   "name": "MODBUS_AI_10_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 38,
   "documented": "38",
   "name": "MODBUS_AI_10_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 39,
   "documented": "39",
   "name": "MODBUS_AI_10_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 40,
   "documented": "40",
   "name": "MODBUS_AI_11_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 42,
   "documented": "42",
   "name": "MODBUS_AI_11_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 43,
   "documented": "43",
   "name": "MODBUS_AI_11_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 44,
   "documented": "44",
   "name": "MODBUS_AI_12_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 46,
   "documented": "46",
   "name": "MODBUS_AI_12_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 47,
   "documented": "47",
   "name": "MODBUS_AI_12_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 48,
   "documented": "48",
   "name": "MODBUS_AI_13_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 50,
   "documented": "50",
   "name": "MODBUS_AI_13_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 51,
   "documented": "51",
   "name": "MODBUS_AI_13_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 52,
   "documented": "52",
   "name": "MODBUS_AI_14_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 54,
   "documented": "54",
   "name": "MODBUS_AI_14_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 55,
   "documented": "55",
   "name": "MODBUS_AI_14_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 56,
   "documented": "56",
   "name": "MODBUS_AI_15_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 58,
   "documented": "58",
   "name": "MODBUS_AI_15_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 59,
   "documented": "59",
   "name": "MODBUS_AI_15_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 60,
   "documented": "60",
   "name": "MODBUS_AI_16_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 62,
   "documented": "62",
   "name": "MODBUS_AI_16_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 63,
   "documented": "63",
   "name": "MODBUS_AI_16_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 64,
   "documented": "64",
   "name": "CURRENTOUT_01_PV_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 66,
   "documented": "66",
   "name": "CURRENTOUT_01_PV_STATUS",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 67,
   "documented": "67",
   "name": "CURRENTOUT_01_PV_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 68,
   "documented": "68",
   "name": "CURRENTOUT_02_PV_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 70,
   "documented": "70",
   "name": "CURRENTOUT_02_PV_STATUS",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 71,
   "documented": "71",
   "name": "CURRENTOUT_02_PV_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 72,
   "documented": "72",
   "name": "CURRENTOUT_03_PV_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 74,
   "documented": "74",
   "name": "CURRENTOUT_03_PV_STATUS",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 75,
   "documented": "75",
   "name": "CURRENTOUT_03_PV_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 76,
   "documented": "76",
   "name": "CURRENTOUT_04_PV_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 78,
   "documented": "78",
   "name": "CURRENTOUT_04_PV_STATUS",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 79,
   "documented": "79",
   "name": "CURRENTOUT_04_PV_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 80,
   "documented": "80",
   "name": "CURRENTOUT_05_PV_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 82,
   "documented": "82",
   "name": "CURRENTOUT_05_PV_STATUS",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 83,
   "documented": "83",
   "name": "CURRENTOUT_05_PV_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 84,
   "documented": "84",
   "name": "CURRENTOUT_06_PV_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 86,
   "documented": "86",
   "name": "CURRENTOUT_06_PV_STATUS",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 87,
   "documented": "87",
   "name": "CURRENTOUT_06_PV_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 88,
   "documented": "88",
   "name": "CURRENTOUT_07_PV_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 90,
   "documented": "90",
   "name": "CURRENTOUT_07_PV_STATUS",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 91,
   "documented": "91",
   "name": "CURRENTOUT_07_PV_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 92,
   "documented": "92",
   "name": "CURRENTOUT_08_PV_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 94,
   "documented": "94",
   "name": "CURRENTOUT_08_PV_STATUS",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 95,
   "documented": "95",
   "name": "CURRENTOUT_08_PV_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 96,
   "documented": "96",
   "name": "CURRENTOUT_01_MA_VALUE",
   "description": "Current output signal; the table prints the unit as ampere.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 98,
   "documented": "98",
   "name": "CURRENTOUT_02_MA_VALUE",
   "description": "Current output signal; the table prints the unit as ampere.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 100,
   "documented": "100",
   "name": "CURRENTOUT_03_MA_VALUE",
   "description": "Current output signal; the table prints the unit as ampere.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 102,
   "documented": "102",
   "name": "CURRENTOUT_04_MA_VALUE",
   "description": "Current output signal; the table prints the unit as ampere.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 104,
   "documented": "104",
   "name": "CURRENTOUT_05_MA_VALUE",
   "description": "Current output signal; the table prints the unit as ampere.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 106,
   "documented": "106",
   "name": "CURRENTOUT_06_MA_VALUE",
   "description": "Current output signal; the table prints the unit as ampere.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 108,
   "documented": "108",
   "name": "CURRENTOUT_07_MA_VALUE",
   "description": "Current output signal; the table prints the unit as ampere.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 110,
   "documented": "110",
   "name": "CURRENTOUT_08_MA_VALUE",
   "description": "Current output signal; the table prints the unit as ampere.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 112,
   "documented": "112",
   "name": "MODBUS_DI_1_VALUE",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "False",
    "1": "True"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 113,
   "documented": "113",
   "name": "MODBUS_DI_1_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 114,
   "documented": "114",
   "name": "MODBUS_DI_2_VALUE",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "False",
    "1": "True"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 115,
   "documented": "115",
   "name": "MODBUS_DI_2_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 116,
   "documented": "116",
   "name": "MODBUS_DI_3_VALUE",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "False",
    "1": "True"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 117,
   "documented": "117",
   "name": "MODBUS_DI_3_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 118,
   "documented": "118",
   "name": "MODBUS_DI_4_VALUE",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "False",
    "1": "True"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 119,
   "documented": "119",
   "name": "MODBUS_DI_4_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 120,
   "documented": "120",
   "name": "MODBUS_DI_5_VALUE",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "False",
    "1": "True"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 121,
   "documented": "121",
   "name": "MODBUS_DI_5_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 122,
   "documented": "122",
   "name": "MODBUS_DI_6_VALUE",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "False",
    "1": "True"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 123,
   "documented": "123",
   "name": "MODBUS_DI_6_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 124,
   "documented": "124",
   "name": "MODBUS_DI_7_VALUE",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "False",
    "1": "True"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 125,
   "documented": "125",
   "name": "MODBUS_DI_7_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 126,
   "documented": "126",
   "name": "MODBUS_DI_8_VALUE",
   "description": "8-bit value in one register.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "False",
    "1": "True"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 127,
   "documented": "127",
   "name": "MODBUS_DI_8_STATUS",
   "description": "Quality of the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "R",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 250,
   "documented": "250",
   "name": "MODBUS_DO_01_VALUE",
   "description": "0 = false, 1-255 = true.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 251,
   "documented": "251",
   "name": "MODBUS_DO_01_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 252,
   "documented": "252",
   "name": "MODBUS_DO_02_VALUE",
   "description": "0 = false, 1-255 = true.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 253,
   "documented": "253",
   "name": "MODBUS_DO_02_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 254,
   "documented": "254",
   "name": "MODBUS_DO_03_VALUE",
   "description": "0 = false, 1-255 = true.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 255,
   "documented": "255",
   "name": "MODBUS_DO_03_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 256,
   "documented": "256",
   "name": "MODBUS_DO_04_VALUE",
   "description": "0 = false, 1-255 = true.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 257,
   "documented": "257",
   "name": "MODBUS_DO_04_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 258,
   "documented": "258",
   "name": "MODBUS_DO_05_VALUE",
   "description": "0 = false, 1-255 = true.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 259,
   "documented": "259",
   "name": "MODBUS_DO_05_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 260,
   "documented": "260",
   "name": "MODBUS_DO_06_VALUE",
   "description": "0 = false, 1-255 = true.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 261,
   "documented": "261",
   "name": "MODBUS_DO_06_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 262,
   "documented": "262",
   "name": "MODBUS_DO_07_VALUE",
   "description": "0 = false, 1-255 = true.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 263,
   "documented": "263",
   "name": "MODBUS_DO_07_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 264,
   "documented": "264",
   "name": "MODBUS_DO_08_VALUE",
   "description": "0 = false, 1-255 = true.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 265,
   "documented": "265",
   "name": "MODBUS_DO_08_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 266,
   "documented": "266",
   "name": "MODBUS_AO_01_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 268,
   "documented": "268",
   "name": "MODBUS_AO_01_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 269,
   "documented": "269",
   "name": "MODBUS_AO_01_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 270,
   "documented": "270",
   "name": "MODBUS_AO_02_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 272,
   "documented": "272",
   "name": "MODBUS_AO_02_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 273,
   "documented": "273",
   "name": "MODBUS_AO_02_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 274,
   "documented": "274",
   "name": "MODBUS_AO_03_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 276,
   "documented": "276",
   "name": "MODBUS_AO_03_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 277,
   "documented": "277",
   "name": "MODBUS_AO_03_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 278,
   "documented": "278",
   "name": "MODBUS_AO_04_VALUE",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 280,
   "documented": "280",
   "name": "MODBUS_AO_04_STATUS",
   "description": "Write 0 (good) or the device ignores the value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Good",
    "1": "Uncertain",
    "2": "Bad",
    "3": "Not assigned"
   },
   "bits": null
  },
  {
   "table": "holding_register",
   "address": 281,
   "documented": "281",
   "name": "MODBUS_AO_04_UNIT",
   "description": "Unit code (table 7.1 of the guide).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "offset": 0,
   "scale_factor_ref": null,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null
  }
 ],
 "integration": {
  "summary": "Assign the measured values you need to AI 1-16 in the transmitter menu, then read registers 0-63 (value, status, unit for each AI) with function 03 or 04 and decode floats as CDAB. Check each status byte is 0 before using the value.",
  "minimum_registers": [
   {
    "table": "holding_register",
    "address": 0,
    "why": "AI 1 value (e.g. main measured value)"
   },
   {
    "table": "holding_register",
    "address": 2,
    "why": "AI 1 status"
   },
   {
    "table": "holding_register",
    "address": 3,
    "why": "AI 1 unit code"
   },
   {
    "table": "holding_register",
    "address": 4,
    "why": "AI 2 value (e.g. temperature)"
   },
   {
    "table": "holding_register",
    "address": 6,
    "why": "AI 2 status"
   },
   {
    "table": "holding_register",
    "address": 112,
    "why": "DI 1 value (e.g. limit contact)"
   }
  ],
  "enable": "Menu Setup/General settings/Extended setup/Modbus/Settings: bus address (default 247, or set by the module's address switch), RTU/ASCII, baud (default 19200), parity (default even, 1 stop bit), byte order and watchdog. For TCP set the IP under .../Ethernet/Settings (DHCP by default) and use port 502. Assign sources to the device variables under Setup/Outputs.",
  "watchdog": "Modbus watchdog default 5 s (0-999 s, 0 = off), set in the Modbus settings menu. If no valid request arrives in that time the device raises diagnostic S969 and lights the red BF LED until the next valid request.",
  "quirks": [
   "Floats are CDAB at the factory byte order but 32-bit integers and strings are most-significant-byte first.",
   "A hardware address set on the module's address switch overrides the menu address."
  ],
  "setup": [
   "RS-485 termination is set on the module; input register 941 reports whether it is on.",
   "The 485/ETH module has a service switch that forces IP 192.168.1.212/24 while it is on."
  ],
  "gotchas": [
   "Values written to AO/DO are ignored unless their status register is written as 0 (good) in the same transfer.",
   "Up to 125 registers per read and 120 per write."
  ]
 },
 "page": "https://easymodbus.com/modbus-register-maps/endress-hauser/liquiline-cm44x",
 "address_note": "address = 0-based Modbus wire (PDU) address; documented = as printed in the source.",
 "note": "Re-typed from the manufacturer's public documentation (see source) and verified against it. Registers vary by firmware and options; verify against the live device before writing."
}