 "manufacturer": "JUMO",
 "manufacturer_slug": "jumo",
 "model": "tecLine digital disinfection sensors (Cl2, TC, ClO2/O3, H2O2/PAA, Br, Cl2 OM) (Modbus RTU)",
 "model_slug": "tecline-20263x",
 "short_model": "tecLine 20263x",
 "aliases": [
  "202630",
  "202631",
  "202634",
  "202636",
  "202637",
  "202681",
  "tecLine Cl2",
  "tecLine TC",
  "tecLine ClO2",
  "tecLine O3",
  "tecLine H2O2",
  "tecLine PAA",
  "tecLine Br",
  "tecLine Cl2 OM"
 ],
 "equipment_type": "Sensor (disinfectant / chlorine)",
 "applies_to": "JUMO tecLine 202630, 202631, 202634, 202636, 202637 and 202681 sensors with RS-485 Modbus RTU (interface description V2.00, 2022-03-24).",
 "summary": "JUMO tecLine sensors are amperometric membrane sensors for free and total chlorine, chlorine dioxide, ozone, hydrogen peroxide, peracetic acid and bromine, with a built-in RS-485 Modbus RTU interface. This map covers concentration, cell current and temperature, the unit and decimal settings, calibration zero/slope with history, device identification and serial settings.",
 "transport": [
  "RTU"
 ],
 "defaults": {
  "unit_id": null,
  "baud": 38400,
  "parity": "None",
  "data_bits": 8,
  "stop_bits": 1,
  "tcp_port": null
 },
 "addressing": {
  "convention": "hex 0-based",
  "note": "Addresses are printed in hex and sent unchanged; the example reads the temperature with start address 0x0004."
 },
 "default_word_order": "CDAB",
 "firmware": null,
 "source": {
  "title": "JUMO tecLine Cl2 (202630), TC (202631), ClO2+O3 (202634), H2O2+PAA (202636), Br (202637), Cl2 OM (202681) - Modbus interface description",
  "doc_number": "20263000T92Z001K000 V2.00/EN/00718260",
  "revision_date": "2022-03-24",
  "url": "https://www.onetemp.com.au/wp-content/uploads/2024/02/Jumo-tecLine-Modbus-Interface-Description-for-CI2-TC-CIO2O3-H2O2PAA-Br-CI2-OM.pdf",
  "unofficial": true,
  "pages": "Float format ch. 2.5 and 2.6.3 (pp. 10-13); serial settings ch. 5.1 (p. 17); address tables ch. 6 (pp. 18-19)"
 },
 "notes": [
  "Copy hosted by the Australian distributor Onetemp; JUMO does not offer a public direct link to this document. The file is the unaltered JUMO interface description.",
  "Floats are sent low word first (CDAB): the guide says the bytes come as 3-4-1-2, and its temperature example returns BA 2F 41 C0 = 24.09 °C.",
  "Default slave address depends on type: 20 Cl2, 30 TC, 50 O3, 80 ClO2, 60 H2O2, 70 PAA, 90 Br, 100 Cl2 OM. Default serial format reads 38400 baud, none parity, 1 stop bit in the settings table.",
  "Left out: the calibration date/time stamps (0x020A and history), which are unsigned 32-bit integers with no stated word order; writing the stamp at 0x020A is what activates a new calibration, so use JUMO DSM software or test word order first.",
  "Left out: history entry 3, because the table prints the same addresses (0x0212/0x0214) as entry 0.",
  "Unit code 1 is printed as an unreadable symbol in the table and is left out of the enum.",
  "Function 03 and 04 both read all parameters; writes use 06 or 16."
 ],
 "registers": [
  {
   "table": "holding_register",
   "address": 0,
   "documented": "0x0000",
   "name": "Concentration",
   "description": "In the unit set at 0x0200.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2,
   "documented": "0x0002",
   "name": "Cell current",
   "description": "Sensor current referred to 25 °C.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4,
   "documented": "0x0004",
   "name": "Temperature",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 512,
   "documented": "0x0200",
   "name": "Unit",
   "description": "Unit of the concentration value.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0": "%",
    "2": "g/l",
    "3": "ppm",
    "4": "mg/l",
    "5": "ppb",
    "6": "µg/l"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 513,
   "documented": "0x0201",
   "name": "Decimal places",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0": "0000",
    "1": "000.0",
    "2": "00.00",
    "3": "0.000"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 518,
   "documented": "0x0206",
   "name": "X_Zero (active calibration)",
   "description": "Zero-point current.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 520,
   "documented": "0x0208",
   "name": "X_Span (active calibration)",
   "description": "Slope.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA/unit",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 528,
   "documented": "0x0210",
   "name": "X_Zero history 0",
   "description": "Latest calibration.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 530,
   "documented": "0x0212",
   "name": "X_Span history 0",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA/unit",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 534,
   "documented": "0x0216",
   "name": "X_Zero history 1",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 536,
   "documented": "0x0218",
   "name": "X_Span history 1",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA/unit",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 540,
   "documented": "0x021C",
   "name": "X_Zero history 2",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 542,
   "documented": "0x021E",
   "name": "X_Span history 2",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA/unit",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 552,
   "documented": "0x0228",
   "name": "X_Zero history 4",
   "description": "Oldest calibration kept.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 554,
   "documented": "0x022A",
   "name": "X_Span history 4",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA/unit",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 558,
   "documented": "0x022E",
   "name": "Measuring range",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 776,
   "documented": "0x0308",
   "name": "Hardware version",
   "description": "E.g. 1130 = 1.130.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 777,
   "documented": "0x0309",
   "name": "Firmware version",
   "description": "E.g. 1503 = 1.503.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 778,
   "documented": "0x030A",
   "name": "Nominal slope",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "nA/unit",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 780,
   "documented": "0x030C",
   "name": "Fabrication number",
   "description": "20 characters.",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 791,
   "documented": "0x0317",
   "name": "Part number",
   "description": "10 characters.",
   "data_type": "STRING",
   "length": 5,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1024,
   "documented": "0x0400",
   "name": "Slave address",
   "description": "1-247; default depends on sensor type.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1025,
   "documented": "0x0401",
   "name": "Baud rate",
   "description": "Default 4 (38400).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "2400",
    "1": "4800",
    "2": "9600",
    "3": "19200",
    "4": "38400",
    "5": "57600",
    "6": "115200"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1026,
   "documented": "0x0402",
   "name": "Parity / stop bits",
   "description": "Default 3 (none/1).",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "None, 2 stop",
    "1": "Even, 1 stop",
    "2": "Odd, 1 stop",
    "3": "None, 1 stop"
   },
   "bits": null,
   "offset": 0
  }
 ],
 "integration": {
  "summary": "Read 6 registers from 0x0000 (function 03) for concentration, cell current and temperature as CDAB floats; read 0x0200 once for the unit.",
  "minimum_registers": [
   {
    "table": "holding_register",
    "address": 0,
    "why": "Concentration"
   },
   {
    "table": "holding_register",
    "address": 4,
    "why": "Temperature"
   },
   {
    "table": "holding_register",
    "address": 2,
    "why": "Cell current (for calibration)"
   },
   {
    "table": "holding_register",
    "address": 512,
    "why": "Unit code"
   }
  ],
  "enable": "Modbus RTU is the native interface; change address and serial format with JUMO DSM software or registers 0x0400-0x0402.",
  "watchdog": null,
  "quirks": [
   "Floats are low word first (CDAB).",
   "Only RTU mode is supported."
  ],
  "setup": [
   "M12 pins: 2 = +24 V, 3 = GND, 4 = RS-485 B, 5 = RS-485 A.",
   "Check the type-specific default address before commissioning several sensors on one bus."
  ],
  "gotchas": [
   "Sensors ship uncalibrated; a calibration only becomes active when the date/time stamp is written.",
   "Replace the membrane/electrolyte when the slope falls below 30 % of the nominal slope (0x030A)."
  ]
 },
 "page": "https://easymodbus.com/modbus-register-maps/jumo/tecline-20263x",
 "address_note": "address = 0-based Modbus wire (PDU) address; documented = as printed in the source.",
 "note": "Re-typed from the manufacturer's public documentation (see source) and verified against it. Registers vary by firmware and options; verify against the live device before writing."
}