 "manufacturer": "Siemens",
 "manufacturer_slug": "siemens",
 "model": "SITRANS F C MASS 6000 with USM II Modbus RTU module",
 "model_slug": "sitrans-f-c-mass-6000-modbus",
 "short_model": "MASS 6000",
 "aliases": [
  "MASS 6000",
  "SITRANS FC MASS 6000"
 ],
 "equipment_type": "Sensor (flow transmitter)",
 "applies_to": "SITRANS F C MASS 6000 Coriolis flow transmitters fitted with the USM II Modbus RTU RS-485 add-on module (operating instructions 03/2021).",
 "summary": "The MASS 6000 Coriolis transmitter gets a Modbus RTU slave by fitting the USM II RS-485 add-on module. Data is held in holding registers and coils only. This map covers mass and volume flow, density, sensor temperature, fractions, totalizers, driver and identity registers, the MASS 6000 settings tables, sensor diagnostics, system status and the control coils.",
 "transport": [
  "RTU"
 ],
 "defaults": {
  "unit_id": 1,
  "baud": 19200,
  "parity": "Even",
  "data_bits": 8,
  "stop_bits": 1,
  "tcp_port": null
 },
 "addressing": {
  "convention": "0-based",
  "note": "Every row prints both the Modbus register (4:0xxxx / 0:0xxxx, 1-based) and the Modbus address (0-based) sent on the wire; this map uses the address column. The manual's own example reads absolute massflow 4:03001 with query address 0x0BB8 (3000)."
 },
 "default_word_order": "ABCD",
 "firmware": null,
 "source": {
  "title": "SITRANS F Modbus RTU RS 485 (USM II) communication module, Operating Instructions",
  "doc_number": "A5E00753974-AB",
  "revision_date": "2021-03",
  "url": "https://cache.industry.siemens.com/dl/files/906/44880906/att_70364/v1/A5E00753974-ABen_USMII_MODBUS_AOM_OI_en-US.pdf",
  "pages": "process data p. 38; driver/application settings pp. 38-40; MASS 6000 settings tables A-9, A-11, A-13, A-15, A-17, A-19, A-21, A-23, A-26 pp. 41-58; status p. 57; coils pp. 64-66; float definition and example pp. 71-72; factory settings p. 21"
 },
 "notes": [
  "Word order: the manual states the module is big-endian, most significant byte first, and its float table and example frame (4:03001 read as 40 C3 52 93 = 6.10383 kg/s) put the high word in register N. Floats, doubles and double words are therefore listed as ABCD.",
  "'Byte' values sit in the low byte of one register and are listed as UINT16 (0-255).",
  "Units of the process values and totalizers follow the unit registers (2904-2917, ID tables in Appendix B.1/B.2). The 'point' registers set the number of decimals (0-6) shown on the display.",
  "Adding 10000 to the address of a writable register returns its minimum allowed value, adding 20000 returns its maximum (these mirror registers are not listed).",
  "Values printed as (1/100%) are fractions where 1.0 = 100 % (the manual's example: 0.02 = 2 %).",
  "Left out: capability bits (603, 3-byte array) and the 54-byte error pending / error log lists (2702, 2729); their record layout is in Appendix B.4.",
  "Error numbers are listed in the SITRANS F M / F C transmitter operating instructions, not in this manual.",
  "Process values 3000-3012 are in fixed SI units (kg/s, m3/s, kg/m3, °C) as printed; the totalizers are kg or m3 depending on the totalizer selection."
 ],
 "registers": [
  {
   "table": "coil",
   "address": 0,
   "documented": "0:00001",
   "name": "Restart MODBUS communication",
   "description": "Write 1 to restart; reads 0.",
   "data_type": "BIT",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "coil",
   "address": 8,
   "documented": "0:00009",
   "name": "Start auto zero adjust",
   "description": "Write 1 to start; reads 1 while busy.",
   "data_type": "BIT",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "coil",
   "address": 9,
   "documented": "0:00010",
   "name": "Reset totalizer 1",
   "description": "Write 1 to reset; reads 0.",
   "data_type": "BIT",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "coil",
   "address": 10,
   "documented": "0:00011",
   "name": "Hold totalizer 1",
   "description": "1 = hold, 0 = run; reads back held state.",
   "data_type": "BIT",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "coil",
   "address": 11,
   "documented": "0:00012",
   "name": "Reset totalizer 2",
   "description": "Write 1 to reset; reads 0.",
   "data_type": "BIT",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "coil",
   "address": 12,
   "documented": "0:00013",
   "name": "Hold totalizer 2",
   "description": "1 = hold, 0 = run; reads back held state.",
   "data_type": "BIT",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "coil",
   "address": 13,
   "documented": "0:00014",
   "name": "Start batch",
   "description": "1 = start, 0 = stop batch.",
   "data_type": "BIT",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "coil",
   "address": 14,
   "documented": "0:00015",
   "name": "Pause batch",
   "description": "1 = pause, 0 = resume batch.",
   "data_type": "BIT",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 500,
   "documented": "4:00501",
   "name": "No. of ParityErrors",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 501,
   "documented": "4:00502",
   "name": "No. of FramingErrors",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 503,
   "documented": "4:00504",
   "name": "No. of CRC Errors",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 504,
   "documented": "4:00505",
   "name": "No. of Ok Messages (received)",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 507,
   "documented": "4:00508",
   "name": "InterFrameSpace",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "us",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 511,
   "documented": "4:00512",
   "name": "Response Timeout",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "ms",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 512,
   "documented": "4:00513",
   "name": "Response Delay",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "ms",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 513,
   "documented": "4:00514",
   "name": "10xInterFrameSpace (10 x chars)",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 514,
   "documented": "4:00515",
   "name": "Baudrate as double word",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 528,
   "documented": "4:00529",
   "name": "Device address (1 - 247)",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 529,
   "documented": "4:00530",
   "name": "Baudrate",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "1200",
    "1": "2400",
    "2": "4800",
    "3": "9600",
    "4": "19200",
    "5": "38400",
    "6": "57600",
    "7": "76800",
    "8": "115200"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 530,
   "documented": "4:00531",
   "name": "Parity/framing",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "8, E, 1",
    "1": "8, O, 1",
    "2": "8, N, 2",
    "3": "8, N, 1"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 600,
   "documented": "4:00601",
   "name": "Slave ID",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0x2A": "Siemens"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 601,
   "documented": "4:00602",
   "name": "Run indicator",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0x00": "not running",
    "0xFF": "running"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 602,
   "documented": "4:00603",
   "name": "Product code",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "24": "MAG 6000",
    "25": "MASS 6000"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 605,
   "documented": "4:00606",
   "name": "Flowmeter revision",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 606,
   "documented": "4:00607",
   "name": "MODBUS revision",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 607,
   "documented": "4:00608",
   "name": "Manufacturer name",
   "description": "",
   "data_type": "STRING",
   "length": 6,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 613,
   "documented": "4:00614",
   "name": "Product name",
   "description": "",
   "data_type": "STRING",
   "length": 9,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 622,
   "documented": "4:00623",
   "name": "MODBUS module name",
   "description": "",
   "data_type": "STRING",
   "length": 6,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 640,
   "documented": "4:00641",
   "name": "TAG",
   "description": "",
   "data_type": "STRING",
   "length": 9,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 649,
   "documented": "4:00650",
   "name": "MODBUS module SW version",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 659,
   "documented": "4:00660",
   "name": "MODBUS module HW version",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 680,
   "documented": "4:00681",
   "name": "Last Coil Error (CoilAddr)",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 681,
   "documented": "4:00682",
   "name": "Last Coil Error (ErrorNo)",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 682,
   "documented": "4:00683",
   "name": "Last HoldReg Error (HoldReg Addr)",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 683,
   "documented": "4:00684",
   "name": "Last HoldReg Error (ErrorNo)",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2000,
   "documented": "4:02001",
   "name": "Flow direction",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Negative",
    "1": "Positive"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2001,
   "documented": "4:02002",
   "name": "Massflow max.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2003,
   "documented": "4:02004",
   "name": "Volumeflow max.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "m3/s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2009,
   "documented": "4:02010",
   "name": "Density min.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/m3",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2011,
   "documented": "4:02012",
   "name": "Density max.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/m3",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2013,
   "documented": "4:02014",
   "name": "Sensor temperature min.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "°C",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2015,
   "documented": "4:02016",
   "name": "Sensor temperature max.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "°C",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2017,
   "documented": "4:02018",
   "name": "Fraction A max.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2019,
   "documented": "4:02020",
   "name": "Fraction B max.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2021,
   "documented": "4:02022",
   "name": "% fraction A min.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2023,
   "documented": "4:02024",
   "name": "% fraction A max.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2025,
   "documented": "4:02026",
   "name": "Low flow cut off",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2027,
   "documented": "4:02028",
   "name": "Empty pipe limit",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/m3",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2029,
   "documented": "4:02030",
   "name": "Empty pipe on/off",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "On"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2030,
   "documented": "4:02031",
   "name": "Noise filter",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "1": "min.",
    "5": "max."
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2031,
   "documented": "4:02032",
   "name": "Error level",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "1": "Warning",
    "2": "Permanent",
    "3": "Fatal"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2032,
   "documented": "4:02033",
   "name": "Zero adjust mode",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Auto",
    "1": "Manual"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2033,
   "documented": "4:02034",
   "name": "Manual zero adjust",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2035,
   "documented": "4:02036",
   "name": "Zero adjust time",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2036,
   "documented": "4:02037",
   "name": "Zero sigma",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2038,
   "documented": "4:02039",
   "name": "Zero sigma limit",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2100,
   "documented": "4:02101",
   "name": "Totalizer 1 selection",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "1": "Massflow",
    "2": "Fraction A",
    "3": "Fraction B",
    "4": "Volumeflow"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2101,
   "documented": "4:02102",
   "name": "Totalizer 1 direction",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Reverse",
    "1": "Forward",
    "2": "Net"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2102,
   "documented": "4:02103",
   "name": "Totalizer 2 selection",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "1": "Massflow",
    "2": "Fraction A",
    "3": "Fraction B",
    "4": "Volumeflow"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2103,
   "documented": "4:02104",
   "name": "Totalizer 2 direction",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Reverse",
    "1": "Forward",
    "2": "Net"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2200,
   "documented": "4:02201",
   "name": "Current Output 1 selection",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "Massflow",
    "2": "Fraction A",
    "3": "Fraction B",
    "4": "Volumeflow",
    "5": "Sensor temp.",
    "6": "Density",
    "7": "% fraction A"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2201,
   "documented": "4:02202",
   "name": "Current Output 1 direction",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "1": "Unidirectional",
    "2": "Bidirectional"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2202,
   "documented": "4:02203",
   "name": "Current Output 1 range",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "0 - 20 mA",
    "1": "4 - 20 mA",
    "2": "4 - 20 mA + Alarm"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2203,
   "documented": "4:02204",
   "name": "Current Output 1 time constants",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2205,
   "documented": "4:02206",
   "name": "Digital Output 1 function",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "Pulse",
    "2": "Frequency",
    "3": "Error Level",
    "4": "Error Number",
    "5": "Direction/Limit",
    "6": "Batch"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2206,
   "documented": "4:02207",
   "name": "Pulse 1 selection",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Uninitialized",
    "1": "Massflow",
    "2": "Fraction A",
    "3": "Fraction B",
    "4": "Volumeflow"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2207,
   "documented": "4:02208",
   "name": "Pulse 1 direction",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "1": "Unidirectional",
    "2": "Bidirectional"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2208,
   "documented": "4:02209",
   "name": "Pulse 1 width",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "64 us",
    "1": "130 us",
    "2": "260 us",
    "3": "510 us",
    "4": "1.0 ms",
    "5": "2.0 ms",
    "6": "4.1 ms",
    "7": "8.2 ms",
    "8": "16 ms",
    "9": "33 ms",
    "10": "66 ms",
    "11": "130 ms",
    "12": "260 ms",
    "13": "520 ms",
    "14": "1.0 s",
    "15": "2.1 s",
    "16": "4.2 s"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2210,
   "documented": "4:02211",
   "name": "Frequency 1 selection",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off (read only)",
    "1": "Massflow",
    "2": "Fraction A",
    "3": "Fraction B",
    "4": "Volumeflow",
    "5": "Sensor temp.",
    "6": "Density",
    "7": "% fraction A"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2211,
   "documented": "4:02212",
   "name": "Frequency 1 Direction",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "1": "Unidirectional",
    "2": "Bidirectional"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2212,
   "documented": "4:02213",
   "name": "Frequency 1 Fmax.",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "10 kHz",
    "1": "5 kHz",
    "2": "1 kHz",
    "3": "500 Hz"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2213,
   "documented": "4:02214",
   "name": "Frequency 1 Timeconstants",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2215,
   "documented": "4:02216",
   "name": "Limit Digital selection",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off (read only)",
    "1": "Massflow",
    "2": "Fraction A",
    "3": "Fraction B",
    "4": "Volumeflow",
    "5": "Sensor temp.",
    "6": "Density",
    "7": "% fraction A"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2216,
   "documented": "4:02217",
   "name": "Limit Digital mode",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "1 setpoint",
    "1": "2 setpoints"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2217,
   "documented": "4:02218",
   "name": "Limit Digital Setpoint min.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "1/100%",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2219,
   "documented": "4:02220",
   "name": "Limit Digital Setpoint max.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "1/100%",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2221,
   "documented": "4:02222",
   "name": "Limit Digital Hysteresis",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "1/100%",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2223,
   "documented": "4:02224",
   "name": "Relay Output 1 function",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "3": "Error Level",
    "4": "Error Number",
    "5": "Direction/Limit"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2224,
   "documented": "4:02225",
   "name": "Limit Relay selection",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "Massflow",
    "2": "Fraction A",
    "3": "Fraction B",
    "4": "Volumeflow",
    "5": "Sensor temp.",
    "6": "Density",
    "7": "% fraction A"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2225,
   "documented": "4:02226",
   "name": "Limit Relay mode",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "1 setpoint",
    "1": "2 setpoints"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2226,
   "documented": "4:02227",
   "name": "Limit Relay Setpoint min.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "1/100%",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2228,
   "documented": "4:02229",
   "name": "Limit Relay Setpoint max.",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "1/100%",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2230,
   "documented": "4:02231",
   "name": "Limit Relay Hysteresis",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "1/100%",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2232,
   "documented": "4:02233",
   "name": "Error number (0-255)",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2233,
   "documented": "4:02234",
   "name": "Batch selection",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off (read only)",
    "1": "Massflow",
    "2": "Fraction A",
    "3": "Fraction B",
    "4": "Volumeflow"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2234,
   "documented": "4:02235",
   "name": "Batch Quantity",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg or m3",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2236,
   "documented": "4:02237",
   "name": "Batch Compensation",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "m3",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2238,
   "documented": "4:02239",
   "name": "Batch Lead constant",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2240,
   "documented": "4:02241",
   "name": "Batch Time error on/off",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "On"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2241,
   "documented": "4:02242",
   "name": "Max. Batch time",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2243,
   "documented": "4:02244",
   "name": "Batch overrun on/off",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "On"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2244,
   "documented": "4:02245",
   "name": "Batch overrun error",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg or m3",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2246,
   "documented": "4:02247",
   "name": "Batch counter up/down",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Down",
    "1": "Up"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2249,
   "documented": "4:02250",
   "name": "Batch cycle counter",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2290,
   "documented": "4:02291",
   "name": "Mass or Volume per pulse",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "m3",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2292,
   "documented": "4:02293",
   "name": "Pulse Polarity",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Negative",
    "1": "Positive"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2295,
   "documented": "4:02296",
   "name": "Quadratur",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "On"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2300,
   "documented": "4:02301",
   "name": "Input Function",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "Start Batch",
    "2": "Hold/Continue",
    "3": "Stop Batch",
    "4": "Zero Adjust",
    "5": "Totalizer reset",
    "6": "Force output",
    "7": "Freeze output"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2301,
   "documented": "4:02302",
   "name": "Totalizer reset",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Totalizer 1",
    "1": "Totalizer 2",
    "2": "Totalizer 1 + 2"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2302,
   "documented": "4:02303",
   "name": "Force output",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "1/100%",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2400,
   "documented": "4:02401",
   "name": "Sensor size",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "m",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2402,
   "documented": "4:02403",
   "name": "Calibration factor",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "us2/kg",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2404,
   "documented": "4:02405",
   "name": "Correction factor",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2406,
   "documented": "4:02407",
   "name": "Sensor TC (1/100%",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2408,
   "documented": "4:02409",
   "name": "Density parm. A",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/m3",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2410,
   "documented": "4:02411",
   "name": "Density parm. B (*10E6)",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2412,
   "documented": "4:02413",
   "name": "Density TC (1/100%",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2414,
   "documented": "4:02415",
   "name": "Density offset",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/m3",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2416,
   "documented": "4:02417",
   "name": "Density factor",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2418,
   "documented": "4:02419",
   "name": "Table slope",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2421,
   "documented": "4:02422",
   "name": "Fraction offset",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2500,
   "documented": "4:02501",
   "name": "Converter type",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2510,
   "documented": "4:02511",
   "name": "Converter code no.",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2520,
   "documented": "4:02521",
   "name": "Converter serial no.",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2530,
   "documented": "4:02531",
   "name": "Converter SW version",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2540,
   "documented": "4:02541",
   "name": "Sensor type",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2550,
   "documented": "4:02551",
   "name": "Sensor code no.",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2560,
   "documented": "4:02561",
   "name": "Sensor serial no.",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2570,
   "documented": "4:02571",
   "name": "Sensor size",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2580,
   "documented": "4:02581",
   "name": "Customer code no.",
   "description": "",
   "data_type": "STRING",
   "length": 10,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2700,
   "documented": "4:02701",
   "name": "Operating time",
   "description": "",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "s",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2756,
   "documented": "4:02757",
   "name": "Driver signal",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "A",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2758,
   "documented": "4:02759",
   "name": "Pickup 1 amplitude",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "V",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2760,
   "documented": "4:02761",
   "name": "Pickup 2 amplitude",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "V",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2762,
   "documented": "4:02763",
   "name": "Sensor frequency",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "Hz",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2900,
   "documented": "4:02901",
   "name": "Language",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "English",
    "1": "Deutsch",
    "2": "Français",
    "3": "Dansk",
    "4": "Svenska",
    "5": "Suomeksi",
    "6": "Russian",
    "7": "Español",
    "8": "Italiano",
    "9": "Português"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2904,
   "documented": "4:02905",
   "name": "Massflow unit",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2905,
   "documented": "4:02906",
   "name": "Massflow point",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2906,
   "documented": "4:02907",
   "name": "Volumeflow unit",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2907,
   "documented": "4:02908",
   "name": "Volumeflow point",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2908,
   "documented": "4:02909",
   "name": "Fraction A + B unit",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2909,
   "documented": "4:02910",
   "name": "Fraction A point",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2911,
   "documented": "4:02912",
   "name": "Fraction B point",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2912,
   "documented": "4:02913",
   "name": "Totalizer 1 unit",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2913,
   "documented": "4:02914",
   "name": "Totalizer 1 point",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2914,
   "documented": "4:02915",
   "name": "Totalizer 2 unit",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2915,
   "documented": "4:02916",
   "name": "Totalizer 2 point",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2916,
   "documented": "4:02917",
   "name": "Density unit",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2917,
   "documented": "4:02918",
   "name": "Density point",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3000,
   "documented": "4:03001",
   "name": "Absolute Massflow",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3002,
   "documented": "4:03003",
   "name": "Absolute Volumeflow",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "m3/s",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3004,
   "documented": "4:03005",
   "name": "Absolute Density",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/m3",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3006,
   "documented": "4:03007",
   "name": "Sensor temperature",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3008,
   "documented": "4:03009",
   "name": "Fraction A",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3010,
   "documented": "4:03011",
   "name": "Fraction B",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg/s",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3012,
   "documented": "4:03013",
   "name": "% fraction A",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "%",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3014,
   "documented": "4:03015",
   "name": "Totalizer 1",
   "description": "",
   "data_type": "FLOAT64",
   "length": 4,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg or m3",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3018,
   "documented": "4:03019",
   "name": "Totalizer 2/Batch",
   "description": "",
   "data_type": "FLOAT64",
   "length": 4,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg or m3",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3022,
   "documented": "4:03023",
   "name": "Totalizer 1 (float)",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg or m3",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3024,
   "documented": "4:03025",
   "name": "Totalizer 2/Batch (float)",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "kg or m3",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3099,
   "documented": "4:03100",
   "name": "Update rate set",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3200,
   "documented": "4:03201",
   "name": "System status",
   "description": "",
   "data_type": "BITMAP16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": {
    "0": "RunIndicator",
    "1": "ErrorPendingOn",
    "2": "ErrPendChanged",
    "3": "Batching",
    "4": "BatchPaused",
    "5": "Zeroadjusting",
    "6": "Totalizer1Hold",
    "7": "Totalizer2Hold"
   },
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3201,
   "documented": "4:03202",
   "name": "Error pending 1",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0xFFFF": "no error"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3202,
   "documented": "4:03203",
   "name": "Error pending 2",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3203,
   "documented": "4:03204",
   "name": "Error pending 3",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3204,
   "documented": "4:03205",
   "name": "Error pending 4",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3205,
   "documented": "4:03206",
   "name": "Error pending 5",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3206,
   "documented": "4:03207",
   "name": "Error pending 6",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3207,
   "documented": "4:03208",
   "name": "Error pending 7",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3208,
   "documented": "4:03209",
   "name": "Error pending 8",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3209,
   "documented": "4:03210",
   "name": "Error pending 9",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3210,
   "documented": "4:03211",
   "name": "Batch status",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0": "Idle",
    "1": "Ready",
    "2": "Running",
    "3": "Paused"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3211,
   "documented": "4:03212",
   "name": "Zero adjust progress",
   "description": "Progress of the zero adjustment, 0-100 %.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "ABCD",
   "scale": 1,
   "units": "%",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  }
 ],
 "integration": {
  "enable": "Fit the USM II Modbus RTU add-on module in the transmitter. The module adds a 'MODBUS RTU module' menu (between Reset mode and Service mode) for address, baud, parity/framing, timeouts and TAG; the same settings are writable at addresses 528-530.",
  "watchdog": null,
  "setup": [
   "Factory settings: address 1, 19200 baud, even parity, 1 stop bit; response delay 0, interframe space 3.5 characters.",
   "Baud 1200-115200; framing 8E1, 8O1, 8N2 or 8N1.",
   "RS-485 2-wire on screw terminals; up to 31 devices per segment without repeaters.",
   "Modbus settings live in the module; after swapping the module they must be written again from the master."
  ],
  "summary": "Read holding registers (FC 03) from address 3000: seven floats give mass flow, volume flow, density, temperature and fractions, followed by the double totalizers at 3014 and 3018. All values are big-endian, high word first.",
  "minimum_registers": [
   {
    "table": "holding_register",
    "address": 3000,
    "why": "Mass flow"
   },
   {
    "table": "holding_register",
    "address": 3002,
    "why": "Volume flow"
   },
   {
    "table": "holding_register",
    "address": 3004,
    "why": "Density"
   },
   {
    "table": "holding_register",
    "address": 3006,
    "why": "Sensor temperature"
   },
   {
    "table": "holding_register",
    "address": 3014,
    "why": "Totalizer 1 (double)"
   },
   {
    "table": "holding_register",
    "address": 3200,
    "why": "System status bits"
   }
  ],
  "quirks": [
   "Use the address column (0-based); 4:03001 is address 3000, as the manual's example frame shows.",
   "Max 26 registers per read and 25 per write.",
   "No input registers or discrete inputs: everything is holding registers and coils.",
   "3099 picks the process-data update rate set (2 Hz up to 30 Hz mass flow); the manual prints the five sets as a numbered list without stating the register value of each, so no value table is given."
  ],
  "gotchas": [
   "Zero adjust progress (3211) is a float 0-100 % on MASS 6000 but a byte state on MAG 6000.",
   "If the reply is not ready within the response timeout the module returns exception 6 (busy); repeat the request."
  ]
 },
 "page": "https://easymodbus.com/modbus-register-maps/siemens/sitrans-f-c-mass-6000-modbus",
 "address_note": "address = 0-based Modbus wire (PDU) address; documented = as printed in the source.",
 "note": "Re-typed from the manufacturer's public documentation (see source) and verified against it. Registers vary by firmware and options; verify against the live device before writing."
}