 "manufacturer": "Vaisala",
 "manufacturer_slug": "vaisala",
 "model": "GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)",
 "model_slug": "gmp251",
 "short_model": "GMP251",
 "aliases": [
  "GMP251",
  "CARBOCAP GMP251",
  "Vaisala GMP251"
 ],
 "equipment_type": "Sensor (CO2 probe)",
 "applies_to": "GMP251 probe in Modbus RTU mode, user guide M211799EN-N.",
 "summary": "The GMP251 is a Vaisala CARBOCAP probe for percent-level CO2 (incubators, life science). Over RS-485 it is a Modbus RTU slave. This map covers the CO2 and temperature readings, the compensation registers, serial settings and status/error registers.",
 "transport": [
  "RTU"
 ],
 "defaults": {
  "unit_id": 240,
  "baud": 19200,
  "parity": "None",
  "data_bits": 8,
  "stop_bits": 2,
  "tcp_port": null
 },
 "addressing": {
  "convention": "1-based",
  "note": "The guide numbers registers in decimal starting at 1 and also prints the 0-based hex address used in the request; register number 1 is address 0x0000 on the wire."
 },
 "default_word_order": "CDAB",
 "firmware": "Error code register needs probe firmware 1.4.3 or later",
 "source": {
  "title": "GMP251 and GMP80P User Guide",
  "doc_number": "M211799EN-N",
  "revision_date": "2026-11",
  "url": "https://docs.vaisala.com/r/M211799EN-N/en-US",
  "pages": "Appendix A Modbus reference, pp. 131-140 (Tables 62-70)"
 },
 "notes": [
  "All values are holding registers read with function 03 and written with function 16.",
  "32-bit floats and the 32-bit error code put the least significant word at the listed register (CDAB); the guide confirms this with a worked read frame (bytes D4 7A 43 E8 = 465.66 ppm).",
  "Missing data returns quiet NaN (float) or 0x8000 (integer) instead of an exception.",
  "Power-up compensation registers are stored in EEPROM (about 30000 writes); write the volatile registers for frequent updates.",
  "The same guide covers GMP80P, which is not included here; the register list in the Modbus chapter is written for the GMP251 probe.",
  "The guide lists the CO2 registers in ppm."
 ],
 "registers": [
  {
   "table": "holding_register",
   "address": 0,
   "documented": "1",
   "name": "Measured CO2",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "ppm",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2,
   "documented": "3",
   "name": "Compensation T",
   "description": "Temperature value the probe is using for compensation.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 4,
   "documented": "5",
   "name": "Measured T",
   "description": "Temperature measured by the probe.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 256,
   "documented": "257",
   "name": "Measured CO2 (integer)",
   "description": "Integer copy, valid up to 32000 ppm.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "ppm",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 257,
   "documented": "258",
   "name": "Measured CO2 (integer, x10)",
   "description": "Integer copy scaled down by 10; multiply raw by 10.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 10,
   "units": "ppm",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 512,
   "documented": "513",
   "name": "Power-up pressure compensation",
   "description": "Stored in EEPROM; 500-1100 hPa, default 1013.25.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "hPa",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 514,
   "documented": "515",
   "name": "Power-up temperature compensation",
   "description": "Stored in EEPROM; -40...+60, default 25.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 516,
   "documented": "517",
   "name": "Power-up humidity compensation",
   "description": "Stored in EEPROM; 0-100, default 0.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "%RH",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 518,
   "documented": "519",
   "name": "Power-up oxygen compensation",
   "description": "Stored in EEPROM; 0-100, default 0.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "%O2",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 520,
   "documented": "521",
   "name": "Volatile pressure compensation",
   "description": "Cleared at reset; use for frequent updates.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "hPa",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 522,
   "documented": "523",
   "name": "Volatile temperature compensation",
   "description": "Cleared at reset; use for frequent updates.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "°C",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 524,
   "documented": "525",
   "name": "Volatile humidity compensation",
   "description": "Cleared at reset; use for frequent updates.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "%RH",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 526,
   "documented": "527",
   "name": "Volatile oxygen compensation",
   "description": "Cleared at reset; use for frequent updates.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "%O2",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 768,
   "documented": "769",
   "name": "Modbus address",
   "description": "1-247, default 240.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 769,
   "documented": "770",
   "name": "Serial speed",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "4800",
    "1": "9600",
    "2": "19200",
    "3": "38400",
    "4": "57600",
    "5": "115200"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 770,
   "documented": "771",
   "name": "Serial parity",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "None",
    "1": "Even",
    "2": "Odd"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 771,
   "documented": "772",
   "name": "Serial stop bits",
   "description": "1 or 2, default 2.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 772,
   "documented": "773",
   "name": "Pressure compensation mode",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "On"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 773,
   "documented": "774",
   "name": "Temperature compensation mode",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "Given",
    "2": "Internal"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 774,
   "documented": "775",
   "name": "Humidity compensation mode",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "On"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 775,
   "documented": "776",
   "name": "Oxygen compensation mode",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "On"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 776,
   "documented": "777",
   "name": "CO2 filtering factor",
   "description": "0-100; 100 = no filtering, 50 = factor 0.5.",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2048,
   "documented": "2049",
   "name": "Device status",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0": "OK",
    "1": "Critical error",
    "2": "Error",
    "4": "Warning"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2049,
   "documented": "2050",
   "name": "CO2 status",
   "description": "",
   "data_type": "UINT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0": "OK",
    "2": "CO2 reading not reliable (start-up)",
    "256": "Measurement not ready (start-up)"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2051,
   "documented": "2052",
   "name": "Error code",
   "description": "Sum of active error bits; probe firmware 1.4.3 and later.",
   "data_type": "UINT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": {
    "0": "Program memory CRC critical error",
    "1": "Parameter memory CRC critical error",
    "4": "Low supply voltage",
    "5": "Internal 30 V error",
    "6": "Low RX signal",
    "7": "Internal 8 V error",
    "8": "RX signal cut",
    "12": "Out of measurement range",
    "13": "Sensor heater error",
    "14": "IR temperature error",
    "15": "FPI slope error",
    "16": "Internal 2.5 V error",
    "17": "Internal 1.7 V error",
    "18": "Low IR current"
   },
   "offset": 0
  }
 ],
 "integration": {
  "summary": "Read holding registers 1-6 in one request (function 03) for CO2 and temperature as CDAB floats. Write compensation values to the volatile registers 521-528 if they change often.",
  "minimum_registers": [
   {
    "table": "holding_register",
    "address": 0,
    "why": "CO2 concentration"
   },
   {
    "table": "holding_register",
    "address": 4,
    "why": "Probe temperature"
   },
   {
    "table": "holding_register",
    "address": 2048,
    "why": "Device status"
   },
   {
    "table": "holding_register",
    "address": 2051,
    "why": "Active error bits"
   },
   {
    "table": "holding_register",
    "address": 520,
    "why": "Live pressure compensation"
   }
  ],
  "enable": "Modbus RTU over RS-485 is one of the probe's output modes, chosen at order or later by configuration. From Vaisala Industrial Protocol, switch with the serial command smode modbus. If pin 5 is tied to GND (pin 3) the probe is locked in analog mode and Modbus is not available.",
  "watchdog": null,
  "quirks": [
   "Float word order is low word first (CDAB); use \"word-swapped\" in masters that expect high word first.",
   "Read each 32-bit value in a single transaction."
  ],
  "setup": [
   "Defaults: address 240, 19200 baud, no parity, 8 data bits, 2 stop bits.",
   "Change address and serial settings with registers 769-772, or in Vaisala Insight PC software."
  ],
  "gotchas": [
   "Register 258 must be multiplied by 10.",
   "Relative humidity compensation only works with temperature and pressure compensation also enabled."
  ]
 },
 "page": "https://easymodbus.com/modbus-register-maps/vaisala/gmp251",
 "address_note": "address = 0-based Modbus wire (PDU) address; documented = as printed in the source.",
 "note": "Re-typed from the manufacturer's public documentation (see source) and verified against it. Registers vary by firmware and options; verify against the live device before writing."
}