 "manufacturer": "Vaisala",
 "manufacturer_slug": "vaisala",
 "model": "Indigo510 / Indigo520 transmitters (Modbus TCP)",
 "model_slug": "indigo510-indigo520",
 "short_model": "Indigo520",
 "aliases": [
  "Indigo510",
  "Indigo520",
  "Indigo500 series"
 ],
 "equipment_type": "Sensor transmitter (host for Vaisala probes)",
 "applies_to": "Indigo510 and Indigo520 transmitter registers (unit ID 240) over Modbus TCP (user guide M212287EN-AJ).",
 "summary": "Indigo510 and Indigo520 are Vaisala transmitters that host one or two smart probes (humidity, CO2, dew point, H2O2 and others) and expose them over Modbus TCP, analog outputs and relays. This map covers the transmitter itself at unit ID 240: optional barometer and analog input, analog output and relay configuration, network and clock settings.",
 "transport": [
  "TCP"
 ],
 "defaults": {
  "unit_id": 240,
  "baud": null,
  "parity": null,
  "data_bits": null,
  "stop_bits": null,
  "tcp_port": 502
 },
 "addressing": {
  "convention": "1-based",
  "note": "The guide numbers registers in decimal starting at 1 and also prints the 0-based hex address used in the request; register number 1 is address 0x0000 on the wire."
 },
 "default_word_order": "CDAB",
 "firmware": null,
 "source": {
  "title": "Indigo510, Indigo520 User Guide",
  "doc_number": "M212287EN-AJ",
  "revision_date": "2026-10",
  "url": "https://docs.vaisala.com/r/M212287EN-AJ/en-US",
  "pages": "Chapter 10 Modbus pp. 177-178; Appendix A Modbus reference pp. 203-219 (Tables 37-48)"
 },
 "notes": [
  "Transmitter registers are holding registers on unit ID 240: read with function 03, write with function 16.",
  "32-bit floats put the least significant word at the listed register (CDAB). The guide states this, and its worked example (30.56 %RH read from a probe at unit ID 241) confirms it.",
  "Unavailable data reads as quiet NaN (float) or 0x8000 (16-bit integer) rather than an exception.",
  "Probe readings are not in this map: read the probe on unit ID 241 (terminal 1) or 242 (terminal 2) using that probe's own register map.",
  "Modbus TCP read and write access are both disabled by default and must be enabled on the touchscreen or web interface.",
  "Indigo510 has 2 analog outputs; Indigo520 has 4 analog outputs, an analog input and 2 relays (the analog input, outputs and relays are not available on PoE-powered Indigo520 units). Registers for hardware a unit lacks are not useful on it.",
  "Left out: analog output clipping limits, NTP address and DNS server strings, the time-zone enum (91 values) and the test registers.",
  "Network strings occupy 8 registers for up to 15 characters."
 ],
 "registers": [
  {
   "table": "holding_register",
   "address": 42,
   "documented": "43",
   "name": "Barometric pressure",
   "description": "Optional barometer module.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "hPa",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 48,
   "documented": "49",
   "name": "Height-corrected pressure",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "hPa",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 52,
   "documented": "53",
   "name": "Analog input level (actual)",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "mA",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 54,
   "documented": "55",
   "name": "Analog input level (validated)",
   "description": "NaN outside 3.8-20.5 mA.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "mA",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 277,
   "documented": "278",
   "name": "Barometric pressure (integer)",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.1,
   "units": "hPa",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 280,
   "documented": "281",
   "name": "Height-corrected pressure (integer)",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.1,
   "units": "hPa",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 282,
   "documented": "283",
   "name": "Analog input level (integer)",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.001,
   "units": "mA",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 283,
   "documented": "284",
   "name": "Analog input validated (integer)",
   "description": "-32768 outside 3.8-20.5 mA.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 0.001,
   "units": "mA",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 513,
   "documented": "514",
   "name": "Barometer status",
   "description": "0 = OK.",
   "data_type": "BITMAP16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": {
    "5": "Value locked",
    "7": "Sensor failure"
   },
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 514,
   "documented": "515",
   "name": "Barometer error flags",
   "description": "0 = OK; flags add up.",
   "data_type": "BITMAP16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": {
    "2": "Pressure out of range",
    "3": "Pressure measurement error",
    "4": "Manufactory error",
    "5": "Module type not defined",
    "8": "F measurement",
    "10": "Temperature measurement",
    "11": "Info 2nd bank checksum",
    "12": "Info 1st bank checksum",
    "13": "Checksum error of current settings",
    "14": "User setting bank checksum",
    "15": "Factory bank setting checksum"
   },
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 784,
   "documented": "785",
   "name": "HCP height",
   "description": "Height correction, -30...+30 m.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "m",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1024,
   "documented": "1025",
   "name": "Date/time mode",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "1": "NTP",
    "2": "Manual"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1058,
   "documented": "1059",
   "name": "Year",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1059,
   "documented": "1060",
   "name": "Month",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1060,
   "documented": "1061",
   "name": "Day",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1061,
   "documented": "1062",
   "name": "Hour",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1062,
   "documented": "1063",
   "name": "Minute",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1063,
   "documented": "1064",
   "name": "Second",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1536,
   "documented": "1537",
   "name": "IP setting type",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "1": "DHCP",
    "2": "Static IP"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1537,
   "documented": "1538",
   "name": "IP address",
   "description": "15-character ASCII text.",
   "data_type": "STRING",
   "length": 8,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1545,
   "documented": "1546",
   "name": "Gateway",
   "description": "15-character ASCII text.",
   "data_type": "STRING",
   "length": 8,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1553,
   "documented": "1554",
   "name": "Subnet mask",
   "description": "15-character ASCII text.",
   "data_type": "STRING",
   "length": 8,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1577,
   "documented": "1578",
   "name": "Apply network settings",
   "description": "Write 1 to apply.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "W",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1792,
   "documented": "1793",
   "name": "AOUT1 output mode",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "0-20 mA",
    "2": "4-20 mA",
    "3": "0-1 V",
    "4": "0-5 V",
    "5": "0-10 V"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1793,
   "documented": "1794",
   "name": "AOUT1 output parameter",
   "description": "Register (unit 240 space) that drives the output.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1794,
   "documented": "1795",
   "name": "AOUT1 scale low end",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1796,
   "documented": "1797",
   "name": "AOUT1 scale high end",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1798,
   "documented": "1799",
   "name": "AOUT1 error output",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1808,
   "documented": "1809",
   "name": "AOUT1 output level",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 1810,
   "documented": "1811",
   "name": "AOUT1 force output level",
   "description": "NaN = not forced.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2048,
   "documented": "2049",
   "name": "AOUT2 output mode",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "0-20 mA",
    "2": "4-20 mA",
    "3": "0-1 V",
    "4": "0-5 V",
    "5": "0-10 V"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2049,
   "documented": "2050",
   "name": "AOUT2 output parameter",
   "description": "Register (unit 240 space) that drives the output.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2050,
   "documented": "2051",
   "name": "AOUT2 scale low end",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2052,
   "documented": "2053",
   "name": "AOUT2 scale high end",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2054,
   "documented": "2055",
   "name": "AOUT2 error output",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2064,
   "documented": "2065",
   "name": "AOUT2 output level",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2066,
   "documented": "2067",
   "name": "AOUT2 force output level",
   "description": "NaN = not forced.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2304,
   "documented": "2305",
   "name": "AOUT3 output mode",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "0-20 mA",
    "2": "4-20 mA",
    "3": "0-1 V",
    "4": "0-5 V",
    "5": "0-10 V"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2305,
   "documented": "2306",
   "name": "AOUT3 output parameter",
   "description": "Register (unit 240 space) that drives the output.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2306,
   "documented": "2307",
   "name": "AOUT3 scale low end",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2308,
   "documented": "2309",
   "name": "AOUT3 scale high end",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2310,
   "documented": "2311",
   "name": "AOUT3 error output",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2320,
   "documented": "2321",
   "name": "AOUT3 output level",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2322,
   "documented": "2323",
   "name": "AOUT3 force output level",
   "description": "NaN = not forced.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2560,
   "documented": "2561",
   "name": "AOUT4 output mode",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "0-20 mA",
    "2": "4-20 mA",
    "3": "0-1 V",
    "4": "0-5 V",
    "5": "0-10 V"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2561,
   "documented": "2562",
   "name": "AOUT4 output parameter",
   "description": "Register (unit 240 space) that drives the output.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2562,
   "documented": "2563",
   "name": "AOUT4 scale low end",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2564,
   "documented": "2565",
   "name": "AOUT4 scale high end",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2566,
   "documented": "2567",
   "name": "AOUT4 error output",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2576,
   "documented": "2577",
   "name": "AOUT4 output level",
   "description": "",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "R",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 2578,
   "documented": "2579",
   "name": "AOUT4 force output level",
   "description": "NaN = not forced.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3072,
   "documented": "3073",
   "name": "Relay 1 mode",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "Above setpoint",
    "2": "Below setpoint"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3073,
   "documented": "3074",
   "name": "Relay 1 parameter",
   "description": "Register (unit 240 space) that drives the relay.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3074,
   "documented": "3075",
   "name": "Relay 1 active above",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3076,
   "documented": "3077",
   "name": "Relay 1 active below",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3078,
   "documented": "3079",
   "name": "Relay 1 hysteresis",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3081,
   "documented": "3082",
   "name": "Relay 1 output state",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0": "Disabled",
    "1": "Active",
    "2": "Inactive"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3082,
   "documented": "3083",
   "name": "Relay 1 force output",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Normal mode",
    "1": "Force active",
    "2": "Force inactive"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3328,
   "documented": "3329",
   "name": "Relay 2 mode",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Off",
    "1": "Above setpoint",
    "2": "Below setpoint"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3329,
   "documented": "3330",
   "name": "Relay 2 parameter",
   "description": "Register (unit 240 space) that drives the relay.",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3330,
   "documented": "3331",
   "name": "Relay 2 active above",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3332,
   "documented": "3333",
   "name": "Relay 2 active below",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3334,
   "documented": "3335",
   "name": "Relay 2 hysteresis",
   "description": "NaN if not set.",
   "data_type": "FLOAT32",
   "length": 2,
   "word_order": "CDAB",
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": null,
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3337,
   "documented": "3338",
   "name": "Relay 2 output state",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "R",
   "values": {
    "0": "Disabled",
    "1": "Active",
    "2": "Inactive"
   },
   "bits": null,
   "offset": 0
  },
  {
   "table": "holding_register",
   "address": 3338,
   "documented": "3339",
   "name": "Relay 2 force output",
   "description": "",
   "data_type": "INT16",
   "length": 1,
   "word_order": null,
   "scale": 1,
   "units": "",
   "access": "RW",
   "values": {
    "0": "Normal mode",
    "1": "Force active",
    "2": "Force inactive"
   },
   "bits": null,
   "offset": 0
  }
 ],
 "integration": {
  "summary": "Enable Modbus TCP read (and write if needed) on the transmitter, then poll unit ID 240 for these registers and 241/242 for the connected probes.",
  "minimum_registers": [
   {
    "table": "holding_register",
    "address": 42,
    "why": "Barometric pressure (if fitted)"
   },
   {
    "table": "holding_register",
    "address": 54,
    "why": "Validated analog input"
   },
   {
    "table": "holding_register",
    "address": 1808,
    "why": "AOUT1 output level"
   },
   {
    "table": "holding_register",
    "address": 3081,
    "why": "Relay 1 state"
   },
   {
    "table": "holding_register",
    "address": 514,
    "why": "Barometer errors"
   }
  ],
  "enable": "Turn on Modbus TCP/IP read and, if needed, write in the touchscreen or web interface (both are off by default).",
  "watchdog": null,
  "quirks": [
   "Floats are low word first (CDAB); pick \"word-swapped\" float in masters that default to high word first.",
   "Read or write each 32-bit value in one transaction."
  ],
  "setup": [
   "Unit ID 240 = transmitter, 241 = probe on terminal 1, 242 = probe on terminal 2; Modbus TCP port 502.",
   "Set DHCP or a static IP with registers 1537-1561 (write 1 to 1578 to apply) or on the touchscreen/web interface."
  ],
  "gotchas": [
   "Use only one Modbus TCP client at a time and one transaction at a time; nested requests are not handled reliably.",
   "Response time is about 2 s; raise the client timeout if requests time out.",
   "Vaisala warns not to leave Modbus write enabled on an open network; turn it off when not needed."
  ]
 },
 "page": "https://easymodbus.com/modbus-register-maps/vaisala/indigo510-indigo520",
 "address_note": "address = 0-based Modbus wire (PDU) address; documented = as printed in the source.",
 "note": "Re-typed from the manufacturer's public documentation (see source) and verified against it. Registers vary by firmware and options; verify against the live device before writing."
}