Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators Modbus register map

Updated 17 November 2026

Short answer

Radwag's PUE C32, HX7 and HX5.EX indicators act as Modbus slaves over RS-232, RS-485 (not on the C32) or Ethernet. Input registers give mass, tare, unit, status and thresholds per platform; holding registers take zero/tare/print commands and parameter writes such as a tare value, thresholds, outputs and product codes.

This page lists 36 Modbus registers for the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators — 15 writable — with the table, 0-based wire address, data type, word order, scale, units and access of each, re-typed from Radwag’s own free documentation (linked below) and checked against it.

Manufacturer
Radwag
Model
PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators
Equipment type
Sensor (weighing indicator)
Registers
36 (15 writable)
Transport
Modbus RTU, Modbus TCP
Default unit ID
Not stated
Documented addressing
Modicon 3xxxx/4xxxx (drop the prefix, subtract 1)
32-bit word order
ABCD high word first (big-endian)
Also known as
PUE C32, PUE HX7, PUE HX5.EX
Source document
Software manual MODBUS - Communication protocol: PUE C32, PUE HX7, PUE HX5.EX Weighing Terminal · ITKP-56-01-05-24-EN · rev. 2024-05
Download CSV all columns JSON Manufacturer PDF ↗

Integration notes: Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators

Integration in brief

Read input registers (FC04) from offset 0: platform 1 mass is a big-endian float at offset 0 and its status bits at offset 5. To zero or tare, write the matching bit to the Command register at offset 500, then clear it before the next use.

Minimum registers to map

Addresses are 0-based wire addresses. Enough to monitor and run the unit; map more as the job needs.

RegisterNameTypeWhyAccess
IR 0Platform 1 massFLOAT32 ABCDPlatform 1 massRead
IR 5Platform 1 statusBITMAP16Platform 1 status (stable, zero, errors)Read
IR 4Platform 1 unitBITMAP16Displayed unitRead
IR 2Platform 1 tareFLOAT32 ABCDPlatform 1 tareRead
HR 500CommandBITMAP16Zero / tare / print command bitsRead/write

Enable Modbus control & writes

Turn Modbus on in the indicator menu under SETUP / Peripherals / Modbus; settings are described in the indicator software manual.

Byte order & addressing

Network setup

Watch out for

Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators Modbus registers

Address = 0-based wire address (the documented number is shown under it when it differs) · Order: ABCD = high word first, CDAB = low word first · Scale: engineering value = raw × scale.

36 of 36 shown
Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators Modbus register map
TableAddressNameDescriptionTypeOrderScaleUnitsAccessValues / bits
IR0doc 30001Platform 1 massMass on platform 1 in the currently displayed unit.FLOAT32ABCDRead
IR2doc 30003Platform 1 tareTare of platform 1 in the adjustment (calibration) unit.FLOAT32ABCDRead
IR4doc 30005Platform 1 unitBit set shows the unit now displayed.BITMAP16Readbit 0 = gram [g]; bit 1 = kilogram [kg]; bit 2 = carat [ct]; bit 3 = pound [lb]; bit 4 = ounce [oz]; bit 5 = Newton [N]
IR5doc 30006Platform 1 statusBITMAP16Readbit 0 = Measurement correct (no error); bit 1 = Measurement stable; bit 2 = Indicating zero; bit 3 = Tared; bit 4 = In weighing range II; bit 5 = In weighing range III; bit 6 = NULL error; bit 7 = LH error; bit 8 = FULL error
IR6doc 30007Platform 1 LO thresholdIn the adjustment unit.FLOAT32ABCDRead
IR8doc 30009Platform 2 massPUE HX7 only.FLOAT32ABCDRead
IR10doc 30011Platform 2 tarePUE HX7 only; adjustment unit.FLOAT32ABCDRead
IR12doc 30013Platform 2 unitPUE HX7 only.BITMAP16Readbit 0 = gram [g]; bit 1 = kilogram [kg]; bit 2 = carat [ct]; bit 3 = pound [lb]; bit 4 = ounce [oz]; bit 5 = Newton [N]
IR13doc 30014Platform 2 statusPUE HX7 only.BITMAP16Readbit 0 = Measurement correct (no error); bit 1 = Measurement stable; bit 2 = Indicating zero; bit 3 = Tared; bit 4 = In weighing range II; bit 5 = In weighing range III; bit 6 = NULL error; bit 7 = LH error; bit 8 = FULL error
IR14doc 30015Platform 2 LO thresholdPUE HX7 only.FLOAT32ABCDRead
IR32doc 30033Process statusDosing / formulation process state.UINT16Read0 = Process disabled; 1 = Process stopped; 2 = Process aborted; 3 = Process finished
IR33doc 30034Input statusLowest 4 bits are the indicator inputs 1-4.BITMAP16Readbit 0 = Input 1; bit 1 = Input 2; bit 2 = Input 3; bit 3 = Input 4
IR34doc 30035MIN thresholdIn the current unit.FLOAT32ABCDRead
IR36doc 30037MAX thresholdIn the current unit.FLOAT32ABCDRead
IR42doc 30043Lot numberNumeric lot number (non-digits are dropped).INT32ABCDRead
IR44doc 30045OperatorCode of the logged-in operator.UINT16Read
IR45doc 30046ProductCode of the selected product.UINT16Read
IR46doc 30047CustomerCode of the selected customer.UINT16Read
IR47doc 30048PackagingCode of the selected packaging.UINT16Read
IR50doc 30051FormulationCode of the selected formulation.UINT16Read
IR51doc 30052Dosing processCode of the selected dosing process.UINT16Read
HR500doc 40501CommandSet a bit to run that action once; clear it before repeating.BITMAP16Read/writebit 0 = Zero the platform; bit 1 = Tare the platform; bit 2 = Delete statistics; bit 3 = Save/Print; bit 4 = Process start; bit 5 = Process stop
HR501doc 40502Command with parameterSet a bit to apply the matching parameter register below.BITMAP16Read/writebit 0 = Set tare value; bit 1 = Set LO threshold; bit 2 = Set lot number; bit 3 = Set outputs state; bit 4 = Select operator; bit 5 = Select product; bit 6 = Select packaging; bit 7 = Set MIN threshold; bit 8 = Select customer; bit 9 = Select source warehouse; bit 10 = Select target warehouse; bit 11 = Select dosing process; bit 12 = Set MAX threshold
HR502doc 40503PlatformPlatform number (1, or 2 on PUE HX7) for parameter commands.UINT16Read/write
HR503doc 40504TareTare to set, adjustment unit.FLOAT32ABCDRead/write
HR505doc 40506LO thresholdLO threshold to set, adjustment unit.FLOAT32ABCDRead/write
HR507doc 40508Output stateOutput mask to apply, bit 0 = output 1.UINT16Read/write
HR508doc 40509MinMIN threshold to set.FLOAT32ABCDRead/write
HR510doc 40511MaxMAX threshold to set.FLOAT32ABCDRead/write
HR516doc 40517Lot numberNumeric lot number to set.INT32ABCDRead/write
HR518doc 40519OperatorOperator code to select.UINT16Read/write
HR519doc 40520ProductProduct code to select.UINT16Read/write
HR520doc 40521CustomerCustomer code to select.UINT16Read/write
HR521doc 40522PackagingPackaging code to select.UINT16Read/write
HR524doc 40525FormulationFormulation code to select.UINT16Read/write
HR525doc 40526Dosing processDosing process code to select.UINT16Read/write

Notes on this map

Check it against the live device. A register map tells you what the controller can expose; the unit in front of you may differ by firmware or options. Read a few registers and compare them with the unit’s own display; for a 32-bit value the float and byte-order decoder shows all four word orders at once, so a wrong word order or scale is obvious. New to this? Read what a register map is and why addresses are off by one.

Questions about the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators

What Modbus registers does the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators have?

This map lists 36 registers: 21 input registers, 15 holding registers. 15 of them can be written. Registers vary with firmware, options and configuration, so confirm against the device.

Are the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators Modbus addresses 0-based or 1-based?

The manufacturer's document numbers registers Modicon 3xxxx/4xxxx (drop the prefix, subtract 1). Every address on this page is the 0-based wire address you put in the request; the documented number is shown under it where it differs. Each table gives both a 0-based Offset and a Modbus address (30001/40001 style); the offset is what goes on the wire, e.g. platform 1 mass is offset 0 = 30001 and Command is offset 500 = 40501. The 'documented' field holds the Modbus address column.

What byte order does the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators use for 32-bit values?

Multi-register values in this map use ABCD (high word first (big-endian)). Set your client to this order for every 32/64-bit value in the table.

What are the minimum Modbus registers to integrate the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators?

Start with IR 0 Platform 1 mass (Platform 1 mass); IR 5 Platform 1 status (Platform 1 status (stable, zero, errors)); IR 4 Platform 1 unit (Displayed unit); IR 2 Platform 1 tare (Platform 1 tare); HR 500 Command (Zero / tare / print command bits).

How do I enable Modbus control of the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators?

Turn Modbus on in the indicator menu under SETUP / Peripherals / Modbus; settings are described in the indicator software manual.

Where is the official Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators Modbus documentation?

The source is Software manual MODBUS - Communication protocol: PUE C32, PUE HX7, PUE HX5.EX Weighing Terminal (ITKP-56-01-05-24-EN), revision 2024-05, published free by Radwag at https://radwag.com/pdf2/en/MODBUS-ITKP-56-01-05-24-EN.pdf.

More Radwag Modbus register maps

Other valves, actuators & sensors register maps

All valves, actuators & sensors Modbus register maps →

Source: Software manual MODBUS - Communication protocol: PUE C32, PUE HX7, PUE HX5.EX Weighing Terminal · ITKP-56-01-05-24-EN · rev. 2024-05. Register numbers, types, scaling and units are reproduced as published; descriptions are our own.