- Home
- Modbus register maps
- Radwag
- PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators
Sensor (weighing indicator) · Modbus register map
Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators Modbus register map
Updated 17 November 2026
Radwag's PUE C32, HX7 and HX5.EX indicators act as Modbus slaves over RS-232, RS-485 (not on the C32) or Ethernet. Input registers give mass, tare, unit, status and thresholds per platform; holding registers take zero/tare/print commands and parameter writes such as a tare value, thresholds, outputs and product codes.
This page lists 36 Modbus registers for the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators — 15 writable — with the table, 0-based wire address, data type, word order, scale, units and access of each, re-typed from Radwag’s own free documentation (linked below) and checked against it.
- Manufacturer
- Radwag
- Model
- PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators
- Equipment type
- Sensor (weighing indicator)
- Registers
- 36 (15 writable)
- Transport
- Modbus RTU, Modbus TCP
- Default unit ID
- Not stated
- Documented addressing
- Modicon 3xxxx/4xxxx (drop the prefix, subtract 1)
- 32-bit word order
- ABCD high word first (big-endian)
- Also known as
- PUE C32, PUE HX7, PUE HX5.EX
Integration notes: Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators
Read input registers (FC04) from offset 0: platform 1 mass is a big-endian float at offset 0 and its status bits at offset 5. To zero or tare, write the matching bit to the Command register at offset 500, then clear it before the next use.
Minimum registers to map
Addresses are 0-based wire addresses. Enough to monitor and run the unit; map more as the job needs.
| Register | Name | Type | Why | Access |
|---|---|---|---|---|
| IR 0 | Platform 1 mass | FLOAT32 ABCD | Platform 1 mass | Read |
| IR 5 | Platform 1 status | BITMAP16 | Platform 1 status (stable, zero, errors) | Read |
| IR 4 | Platform 1 unit | BITMAP16 | Displayed unit | Read |
| IR 2 | Platform 1 tare | FLOAT32 ABCD | Platform 1 tare | Read |
| HR 500 | Command | BITMAP16 | Zero / tare / print command bits | Read/write |
Enable Modbus control & writes
Turn Modbus on in the indicator menu under SETUP / Peripherals / Modbus; settings are described in the indicator software manual.
Byte order & addressing
- Commands act on the rising edge of a bit: write 0 before writing the same bit again.
- Parameter commands: write the value register(s) and the 'Command with parameter' bit, using FC16.
Network setup
- Modbus communication is configured under SETUP / Peripherals / Modbus; the indicator software manual covers the individual settings.
- RS-485 is not available on the PUE C32.
Watch out for
- Platform 2 registers only answer on the PUE HX7.
- The manual's write examples mix two numbering styles (40000 + offset and Modicon); go by the offset column.
Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators Modbus registers
Address = 0-based wire address (the documented number is shown under it when it differs) · Order: ABCD = high word first, CDAB = low word first · Scale: engineering value = raw × scale.
| Table | Address | Name | Description | Type | Order | Scale | Units | Access | Values / bits |
|---|---|---|---|---|---|---|---|---|---|
| IR | 0doc 30001 | Platform 1 mass | Mass on platform 1 in the currently displayed unit. | FLOAT32 | ABCD | Read | |||
| IR | 2doc 30003 | Platform 1 tare | Tare of platform 1 in the adjustment (calibration) unit. | FLOAT32 | ABCD | Read | |||
| IR | 4doc 30005 | Platform 1 unit | Bit set shows the unit now displayed. | BITMAP16 | Read | bit 0 = gram [g]; bit 1 = kilogram [kg]; bit 2 = carat [ct]; bit 3 = pound [lb]; bit 4 = ounce [oz]; bit 5 = Newton [N] | |||
| IR | 5doc 30006 | Platform 1 status | BITMAP16 | Read | bit 0 = Measurement correct (no error); bit 1 = Measurement stable; bit 2 = Indicating zero; bit 3 = Tared; bit 4 = In weighing range II; bit 5 = In weighing range III; bit 6 = NULL error; bit 7 = LH error; bit 8 = FULL error | ||||
| IR | 6doc 30007 | Platform 1 LO threshold | In the adjustment unit. | FLOAT32 | ABCD | Read | |||
| IR | 8doc 30009 | Platform 2 mass | PUE HX7 only. | FLOAT32 | ABCD | Read | |||
| IR | 10doc 30011 | Platform 2 tare | PUE HX7 only; adjustment unit. | FLOAT32 | ABCD | Read | |||
| IR | 12doc 30013 | Platform 2 unit | PUE HX7 only. | BITMAP16 | Read | bit 0 = gram [g]; bit 1 = kilogram [kg]; bit 2 = carat [ct]; bit 3 = pound [lb]; bit 4 = ounce [oz]; bit 5 = Newton [N] | |||
| IR | 13doc 30014 | Platform 2 status | PUE HX7 only. | BITMAP16 | Read | bit 0 = Measurement correct (no error); bit 1 = Measurement stable; bit 2 = Indicating zero; bit 3 = Tared; bit 4 = In weighing range II; bit 5 = In weighing range III; bit 6 = NULL error; bit 7 = LH error; bit 8 = FULL error | |||
| IR | 14doc 30015 | Platform 2 LO threshold | PUE HX7 only. | FLOAT32 | ABCD | Read | |||
| IR | 32doc 30033 | Process status | Dosing / formulation process state. | UINT16 | Read | 0 = Process disabled; 1 = Process stopped; 2 = Process aborted; 3 = Process finished | |||
| IR | 33doc 30034 | Input status | Lowest 4 bits are the indicator inputs 1-4. | BITMAP16 | Read | bit 0 = Input 1; bit 1 = Input 2; bit 2 = Input 3; bit 3 = Input 4 | |||
| IR | 34doc 30035 | MIN threshold | In the current unit. | FLOAT32 | ABCD | Read | |||
| IR | 36doc 30037 | MAX threshold | In the current unit. | FLOAT32 | ABCD | Read | |||
| IR | 42doc 30043 | Lot number | Numeric lot number (non-digits are dropped). | INT32 | ABCD | Read | |||
| IR | 44doc 30045 | Operator | Code of the logged-in operator. | UINT16 | Read | ||||
| IR | 45doc 30046 | Product | Code of the selected product. | UINT16 | Read | ||||
| IR | 46doc 30047 | Customer | Code of the selected customer. | UINT16 | Read | ||||
| IR | 47doc 30048 | Packaging | Code of the selected packaging. | UINT16 | Read | ||||
| IR | 50doc 30051 | Formulation | Code of the selected formulation. | UINT16 | Read | ||||
| IR | 51doc 30052 | Dosing process | Code of the selected dosing process. | UINT16 | Read | ||||
| HR | 500doc 40501 | Command | Set a bit to run that action once; clear it before repeating. | BITMAP16 | Read/write | bit 0 = Zero the platform; bit 1 = Tare the platform; bit 2 = Delete statistics; bit 3 = Save/Print; bit 4 = Process start; bit 5 = Process stop | |||
| HR | 501doc 40502 | Command with parameter | Set a bit to apply the matching parameter register below. | BITMAP16 | Read/write | bit 0 = Set tare value; bit 1 = Set LO threshold; bit 2 = Set lot number; bit 3 = Set outputs state; bit 4 = Select operator; bit 5 = Select product; bit 6 = Select packaging; bit 7 = Set MIN threshold; bit 8 = Select customer; bit 9 = Select source warehouse; bit 10 = Select target warehouse; bit 11 = Select dosing process; bit 12 = Set MAX threshold | |||
| HR | 502doc 40503 | Platform | Platform number (1, or 2 on PUE HX7) for parameter commands. | UINT16 | Read/write | ||||
| HR | 503doc 40504 | Tare | Tare to set, adjustment unit. | FLOAT32 | ABCD | Read/write | |||
| HR | 505doc 40506 | LO threshold | LO threshold to set, adjustment unit. | FLOAT32 | ABCD | Read/write | |||
| HR | 507doc 40508 | Output state | Output mask to apply, bit 0 = output 1. | UINT16 | Read/write | ||||
| HR | 508doc 40509 | Min | MIN threshold to set. | FLOAT32 | ABCD | Read/write | |||
| HR | 510doc 40511 | Max | MAX threshold to set. | FLOAT32 | ABCD | Read/write | |||
| HR | 516doc 40517 | Lot number | Numeric lot number to set. | INT32 | ABCD | Read/write | |||
| HR | 518doc 40519 | Operator | Operator code to select. | UINT16 | Read/write | ||||
| HR | 519doc 40520 | Product | Product code to select. | UINT16 | Read/write | ||||
| HR | 520doc 40521 | Customer | Customer code to select. | UINT16 | Read/write | ||||
| HR | 521doc 40522 | Packaging | Packaging code to select. | UINT16 | Read/write | ||||
| HR | 524doc 40525 | Formulation | Formulation code to select. | UINT16 | Read/write | ||||
| HR | 525doc 40526 | Dosing process | Dosing process code to select. | UINT16 | Read/write |
Notes on this map
- Floats and the lot number are two registers with the first register holding the high bytes (ABCD); the manual's worked example 0x3E68 / 0x72B0 = 0.227 confirms it.
- Mass, MIN and MAX are in the currently displayed unit; tare and LO threshold use the adjustment (calibration) unit.
- The worked write examples on pp. 9-10 label registers as 40000 + offset for the tare value (40503/40504 = offsets 503-504) and the output mask (40507 = offset 507), which agrees with the offset column, but label 'command with parameter' as 40502, which is the table's Modicon number for offset 501. This map follows the table's offset column for every wire address.
- The manual gives no default serial settings or unit ID; they are set in the indicator menu.
- The table types the lot number as 'long' (signed 32-bit); it is mapped as INT32, high word first as the manual states.
Check it against the live device. A register map tells you what the controller can expose; the unit in front of you may differ by firmware or options. Read a few registers and compare them with the unit’s own display; for a 32-bit value the float and byte-order decoder shows all four word orders at once, so a wrong word order or scale is obvious. New to this? Read what a register map is and why addresses are off by one.
Questions about the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators
What Modbus registers does the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators have?
This map lists 36 registers: 21 input registers, 15 holding registers. 15 of them can be written. Registers vary with firmware, options and configuration, so confirm against the device.
Are the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators Modbus addresses 0-based or 1-based?
The manufacturer's document numbers registers Modicon 3xxxx/4xxxx (drop the prefix, subtract 1). Every address on this page is the 0-based wire address you put in the request; the documented number is shown under it where it differs. Each table gives both a 0-based Offset and a Modbus address (30001/40001 style); the offset is what goes on the wire, e.g. platform 1 mass is offset 0 = 30001 and Command is offset 500 = 40501. The 'documented' field holds the Modbus address column.
What byte order does the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators use for 32-bit values?
Multi-register values in this map use ABCD (high word first (big-endian)). Set your client to this order for every 32/64-bit value in the table.
What are the minimum Modbus registers to integrate the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators?
Start with IR 0 Platform 1 mass (Platform 1 mass); IR 5 Platform 1 status (Platform 1 status (stable, zero, errors)); IR 4 Platform 1 unit (Displayed unit); IR 2 Platform 1 tare (Platform 1 tare); HR 500 Command (Zero / tare / print command bits).
How do I enable Modbus control of the Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators?
Turn Modbus on in the indicator menu under SETUP / Peripherals / Modbus; settings are described in the indicator software manual.
Where is the official Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators Modbus documentation?
The source is Software manual MODBUS - Communication protocol: PUE C32, PUE HX7, PUE HX5.EX Weighing Terminal (ITKP-56-01-05-24-EN), revision 2024-05, published free by Radwag at https://radwag.com/pdf2/en/MODBUS-ITKP-56-01-05-24-EN.pdf.
More Radwag Modbus register maps
Other valves, actuators & sensors register maps
- Sensor (chart recorder)
Shinko Technos HR-700 hybrid recorder (Modbus RTU, firmware 4.00+)
Shinko HR-700 hybrid recorders (multipoint and pen types), body version 4.00 or later, Modbus map version 01; manual No. HR7CE4.
Also known as: HR-700 - Actuator / valve (comfort module air damper)
Swegon REACT Parasol Zenith with Belimo actuator (Modbus RTU)
Swegon REACT Parasol Zenith comfort modules fitted with the Belimo Modbus actuator; Swegon Modbus settings sheet dated 2022-08-10.
Also known as: Parasol Zenith REACT, REACT damper actuator - Sensor (CO2 / pressure)
Teracom TSM400-4-CP Modbus CO2 and barometric pressure sensor
TSM400-4-CP per the Teracom TSM400-4-CP user manual TSM400-4-CP-R1.9.
Also known as: TSM400-4-CP - Sensor (dew point transmitter)
Vaisala DMT143 miniature dew point transmitter (Modbus RTU)
DMT143 dew point transmitters ordered with the Modbus RTU option (user guide M211435EN-L).
Also known as: DMT143, DRYCAP DMT143 - Sensor (humidity and temperature module)
Vaisala HMM170 humidity module (Modbus RTU)
HMM170 OEM humidity and temperature module with the Modbus RTU interface (user guide M212259EN-D).
Also known as: HMM170, HUMICAP HMM170 - Sensor (transformer oil monitor)
Vaisala MHT410 moisture, hydrogen and temperature transmitter for transformer oil (Modbus RTU)
MHT410 transformer oil monitor, Modbus RTU on the RS-485 screw terminals (user guide M211737EN-J).
Also known as: MHT410
All valves, actuators & sensors Modbus register maps →
Source: Software manual MODBUS - Communication protocol: PUE C32, PUE HX7, PUE HX5.EX Weighing Terminal · ITKP-56-01-05-24-EN · rev. 2024-05. Register numbers, types, scaling and units are reproduced as published; descriptions are our own.