Sensor (weighing transmitter / mass converter) · Modbus register map

Radwag MW-01-A / MW-01-A7 mass converter Modbus register map

Updated 17 November 2026

Short answer

The MW-01-A mass converter serves Modbus RTU on RS-485 and Modbus TCP (and RTU over TCP) on Ethernet. Input registers return mass, tare, unit, status, thresholds and dosing/adjustment state; holding registers take zero, tare, dosing and calibration commands plus threshold writes. Floats here are low word first, unlike Radwag's PUE indicators.

This page lists 20 Modbus registers for the Radwag MW-01-A / MW-01-A7 mass converter — 9 writable — with the table, 0-based wire address, data type, word order, scale, units and access of each, re-typed from Radwag’s own free documentation (linked below) and checked against it.

Manufacturer
Radwag
Model
MW-01-A / MW-01-A7 mass converter
Equipment type
Sensor (weighing transmitter / mass converter)
Registers
20 (9 writable)
Transport
Modbus RTU, Modbus TCP
Default unit ID
Not stated
Documented addressing
Modicon 3xxxx/4xxxx (drop the prefix, subtract 1)
32-bit word order
CDAB low word first (word-swapped)
Also known as
MW-01-A, MW-01-A7
Source document
Software manual MODBUS - Communication Protocol of MW-01-A Mass Converter · ITKP-41-01-11-21-EN · rev. 2021-11
Download CSV all columns JSON Manufacturer PDF ↗

Integration notes: Radwag MW-01-A / MW-01-A7 mass converter

Integration in brief

Read input registers (FC04) from offset 0: mass is a low-word-first float at offset 0 and the status bits at offset 5. Write bit 0 or 1 of the Command register (offset 256) to zero or tare, clearing it before repeating.

Minimum registers to map

Addresses are 0-based wire addresses. Enough to monitor and run the unit; map more as the job needs.

RegisterNameTypeWhyAccess
IR 0MassFLOAT32 CDABMassRead
IR 5StatusBITMAP16Status (stable, zero, errors)Read
IR 4UnitBITMAP16Displayed unitRead
IR 32Dosing statusUINT16Dosing statusRead
HR 256CommandBITMAP16Zero / tare / dosing commandsRead/write

Enable Modbus control & writes

Configure communication with the MwManager PC software under Parameters / Set Communication.

Byte order & addressing

Network setup

Watch out for

Radwag MW-01-A / MW-01-A7 mass converter Modbus registers

Address = 0-based wire address (the documented number is shown under it when it differs) · Order: ABCD = high word first, CDAB = low word first · Scale: engineering value = raw × scale.

20 of 20 shown
Radwag MW-01-A / MW-01-A7 mass converter Modbus register map
TableAddressNameDescriptionTypeOrderScaleUnitsAccessValues / bits
IR0doc 30001MassMass in the currently displayed unit.FLOAT32CDABRead
IR2doc 30003TareTare in the adjustment (calibration) unit.FLOAT32CDABRead
IR4doc 30005UnitBit set shows the unit now displayed.BITMAP16Readbit 0 = gram [g]; bit 1 = kilogram [kg]; bit 2 = carat [ct]; bit 3 = pound [lb]; bit 4 = ounce [oz]; bit 5 = Newton [N]
IR5doc 30006StatusBITMAP16Readbit 0 = Measurement correct (no error); bit 1 = Measurement stable; bit 2 = Indicating zero; bit 3 = Tared; bit 4 = In weighing range II; bit 5 = In weighing range III; bit 6 = NULL error; bit 7 = LH error; bit 8 = FULL error
IR6doc 30007LO thresholdIn the adjustment unit.FLOAT32CDABRead
IR32doc 30033Dosing statusUINT16Read0 = Process disabled; 1 = Dosing in progress; 2 = Process stopped; 3 = Process finished
IR33doc 30034Input statusBitmask of converter inputs; the 8 low bits are the inputs, bit 0 = input 1.UINT16Read
IR34doc 30035MinMIN threshold, adjustment unit.FLOAT32CDABRead
IR36doc 30037MaxMAX threshold, adjustment unit.FLOAT32CDABRead
IR38doc 30039Fast dosing thresholdAdjustment unit.FLOAT32CDABRead
IR50doc 30051Adjustment statusState of the adjustment (calibration) process.UINT16Read0 = Process finished correctly; 1 = Start mass / adjustment factor determination in progress; 2 = Range exceeded; 3 = Time exceeded; 4 = Process aborted; 5 = Awaiting data
HR256doc 40257CommandSet a bit to run that action once; clear it before repeating.BITMAP16Read/writebit 0 = Zero the platform; bit 1 = Tare the platform; bit 5 = Process start; bit 6 = Process stop; bit 8 = Start mass determination; bit 9 = Adjustment factor determination; bit 10 = Record adjustment parameters
HR257doc 40258Command with parameterSet a bit to apply the matching parameter register below.BITMAP16Read/writebit 0 = Set tare value; bit 1 = Set LO threshold; bit 2 = Set outputs state; bit 3 = Set MIN threshold; bit 4 = Set MAX threshold; bit 5 = Set fast dosing threshold; bit 6 = Set precise dosing threshold; bit 9 = Set adjustment weight mass
HR258doc 40259PlatformPlatform number for parameter commands (always 1).UINT16Read/write
HR259doc 40260TareTare to set, adjustment unit.FLOAT32CDABRead/write
HR261doc 40262LO thresholdLO threshold to set, adjustment unit.FLOAT32CDABRead/write
HR263doc 40264Output stateOutput mask to apply, bit 0 = output 1.UINT16Read/write
HR264doc 40265MinMIN threshold to set.FLOAT32CDABRead/write
HR266doc 40267MaxMAX threshold to set.FLOAT32CDABRead/write
HR268doc 40269Fast dosing thresholdFast dosing threshold to set.FLOAT32CDABRead/write

Notes on this map

Check it against the live device. A register map tells you what the controller can expose; the unit in front of you may differ by firmware or options. Read a few registers and compare them with the unit’s own display; for a 32-bit value the float and byte-order decoder shows all four word orders at once, so a wrong word order or scale is obvious. New to this? Read what a register map is and why addresses are off by one.

Questions about the Radwag MW-01-A / MW-01-A7 mass converter

What Modbus registers does the Radwag MW-01-A / MW-01-A7 mass converter have?

This map lists 20 registers: 11 input registers, 9 holding registers. 9 of them can be written. Registers vary with firmware, options and configuration, so confirm against the device.

Are the Radwag MW-01-A / MW-01-A7 mass converter Modbus addresses 0-based or 1-based?

The manufacturer's document numbers registers Modicon 3xxxx/4xxxx (drop the prefix, subtract 1). Every address on this page is the 0-based wire address you put in the request; the documented number is shown under it where it differs. The tables list a 0-based Offset next to a 30001/40001-style Modbus address; the offset goes on the wire (mass = offset 0 = 30001, Command = offset 256 = 40257).

What byte order does the Radwag MW-01-A / MW-01-A7 mass converter use for 32-bit values?

Multi-register values in this map use CDAB (low word first (word-swapped)). Set your client to this order for every 32/64-bit value in the table.

What are the minimum Modbus registers to integrate the Radwag MW-01-A / MW-01-A7 mass converter?

Start with IR 0 Mass (Mass); IR 5 Status (Status (stable, zero, errors)); IR 4 Unit (Displayed unit); IR 32 Dosing status (Dosing status); HR 256 Command (Zero / tare / dosing commands).

How do I enable Modbus control of the Radwag MW-01-A / MW-01-A7 mass converter?

Configure communication with the MwManager PC software under Parameters / Set Communication.

Where is the official Radwag MW-01-A / MW-01-A7 mass converter Modbus documentation?

The source is Software manual MODBUS - Communication Protocol of MW-01-A Mass Converter (ITKP-41-01-11-21-EN), revision 2021-11, published free by Radwag at https://radwag.com/document/en/user-manual-modbus-communication-protocol-of-mw-01-a-mw-01-a7-mass-converter__MODBUS-MW-01-A-ITKP-41-01-11-21-EN.pdf.

More Radwag Modbus register maps

Other valves, actuators & sensors register maps

All valves, actuators & sensors Modbus register maps →

Source: Software manual MODBUS - Communication Protocol of MW-01-A Mass Converter · ITKP-41-01-11-21-EN · rev. 2021-11. Register numbers, types, scaling and units are reproduced as published; descriptions are our own.