- digiLine O-DO S10 optical oxygen sensor (Modbus RTU)
Sensor (dissolved oxygen) · Modbus register map
JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU) Modbus register map
Updated 19 November 2026
The JUMO digiLine O-DO S10 is an optical dissolved oxygen sensor with digiLine electronics that can be read directly over Modbus RTU or through a JUMO digiLine master. This map covers the oxygen and temperature readings, status word, compensation values (flash and RAM), unit selection, analog output, serial and float-format settings and identification.
This page lists 29 Modbus registers for the JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU) — 19 writable — with the table, 0-based wire address, data type, word order, scale, units and access of each, re-typed from JUMO’s own free documentation (linked below) and checked against it.
- Manufacturer
- JUMO
- Model
- digiLine O-DO S10 optical oxygen sensor (Modbus RTU)
- Equipment type
- Sensor (dissolved oxygen)
- Registers
- 29 (19 writable)
- Transport
- Modbus RTU
- Default unit ID
- Not stated
- Documented addressing
- 0-based (the number printed is the address on the wire)
- 32-bit word order
- CDAB low word first (word-swapped)
- Also known as
- JUMO digiLine O-DO S10, digiLine O-DO
Integration notes: JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU)
Read 9728-9729 (oxygen) and 9760-9761 (temperature) as floats in the format set by register 775, plus the status word at 9792.
Minimum registers to map
Addresses are 0-based wire addresses. Enough to monitor and run the unit; map more as the job needs.
| Register | Name | Type | Why | Access |
|---|---|---|---|---|
| HR 9728 | Measured oxygen value | FLOAT32 CDAB | Oxygen value | Read |
| HR 9760 | Measured temperature | FLOAT32 CDAB | Temperature | Read |
| HR 9792 | Status word | BITMAP16 | Status / warnings | Read |
| HR 775 | Float value format | UINT16 | Float format in use | Read/write |
| HR 9520 | Oxygen measurement unit | UINT16 | Oxygen unit | Read/write |
Enable Modbus control & writes
Modbus RTU is available directly on the sensor RS-485 line; set address and format with JUMO DSM software or registers 768-775.
Byte order & addressing
- Float byte order follows register 775; standard Modbus format is low word first (CDAB).
- Function 03 and 04 both read; writes use 06/16.
Network setup
- Baud 9600/19200/38400 and data format are set in registers 769-770.
Watch out for
- Treat values near 1E37-9E37 as error codes, not readings.
- The sensor origin text (638) is managed by digiLine masters and must not be written.
JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU) Modbus registers
Address = 0-based wire address (the documented number is shown under it when it differs) · Order: ABCD = high word first, CDAB = low word first · Scale: engineering value = raw × scale.
| Table | Address | Name | Description | Type | Order | Scale | Units | Access | Values / bits |
|---|---|---|---|---|---|---|---|---|---|
| HR | 0doc 0 (0x0000) | Device name | ASCII, zero-terminated. | STRING[10] | Read | ||||
| HR | 5doc 5 (0x0005) | Software version | ASCII, zero-terminated. | STRING[12] | Read | ||||
| HR | 256doc 256 (0x0100) | Serial number | ASCII, zero-terminated. | STRING[20] | Read | ||||
| HR | 266doc 266 (0x010A) | Sensor type | UINT16 | Read | 1 = digiLine pH/ORP/T; 2 = digiLine CR/Ci; 4 = digiLine O-DO S10 | ||||
| HR | 267doc 267 (0x010B) | Oxygen sensor sub-type | UINT16 | Read | 0 = digiLine O-DO S10 | ||||
| HR | 768doc 768 (0x0300) | Device address | UINT16 | Read/write | |||||
| HR | 769doc 769 (0x0301) | Baud rate | UINT16 | Read/write | 0 = 9600; 1 = 19200; 2 = 38400 | ||||
| HR | 770doc 770 (0x0302) | Data format | UINT16 | Read/write | 0 = 8-1 no parity; 1 = 8-1 odd parity; 2 = 8-1 even parity; 3 = 8-2 no parity | ||||
| HR | 773doc 773 (0x0305) | Minimum response time | 0-500 ms delay before the sensor answers. | UINT16 | ms | Read/write | |||
| HR | 775doc 775 (0x0307) | Float value format | Word/byte order of all floats. | UINT16 | Read/write | 0 = Standard Modbus format; 1 = IEEE 754 Little Endian; 2 = IEEE 754 Big Endian | |||
| HR | 1026doc 1026 (0x0402) | Sensor replacement counter | UINT16 | Read/write | |||||
| HR | 1289doc 1289 (0x0509) | Lowest temperature value | Drag indicator. | FLOAT32 | CDAB | °C | Read | ||
| HR | 1291doc 1291 (0x050B) | Highest temperature value | Drag indicator. | FLOAT32 | CDAB | °C | Read | ||
| HR | 8769doc 8769 (0x2241) | Analog output function | UINT16 | Read/write | 0 = Inactive; 1 = Fixed value; 2 = Standard signal 4-20 mA | ||||
| HR | 8770doc 8770 (0x2242) | Analog output scaling start | Measured value at 4 mA. | FLOAT32 | CDAB | Read/write | |||
| HR | 8772doc 8772 (0x2244) | Analog output scaling end | Measured value at 20 mA. | FLOAT32 | CDAB | Read/write | |||
| HR | 8776doc 8776 (0x2248) | Analog output fixed value | Used when function = 1. | FLOAT32 | CDAB | mA | Read/write | ||
| HR | 8780doc 8780 (0x224C) | Analog output warning value | 3.5-21.5 mA. | FLOAT32 | CDAB | mA | Read/write | ||
| HR | 8782doc 8782 (0x224E) | Analog output error value | 3.5-21.5 mA. | FLOAT32 | CDAB | mA | Read/write | ||
| HR | 9520doc 9520 (0x2530) | Oxygen measurement unit | UINT16 | Read/write | 0 = % vol; 1 = % SAT; 3 = ppm; 4 = mbar (partial pressure) | ||||
| HR | 9600doc 9600 (0x2580) | Fixed compensation salt content | Stored in flash; do not write cyclically. | FLOAT32 | CDAB | mS/cm | Read/write | ||
| HR | 9602doc 9602 (0x2582) | Fixed ambient pressure | Stored in flash; do not write cyclically. | FLOAT32 | CDAB | hPa | Read/write | ||
| HR | 9647doc 9647 (0x25AF) | Compensation control byte | Low byte selects compensation sources. | BITMAP16 | Read/write | bit 0 = Use RAM compensation temperature (else internal sensor); bit 1 = Use RAM salt content (else flash value); bit 2 = Use RAM ambient pressure (else flash value) | |||
| HR | 9648doc 9648 (0x25B0) | RAM compensation temperature | May be written cyclically. | FLOAT32 | CDAB | °C | Read/write | ||
| HR | 9650doc 9650 (0x25B2) | RAM compensation salt content | May be written cyclically. | FLOAT32 | CDAB | mS/cm | Read/write | ||
| HR | 9652doc 9652 (0x25B4) | RAM ambient pressure | May be written cyclically. | FLOAT32 | CDAB | hPa | Read/write | ||
| HR | 9728doc 9728 (0x2600) | Measured oxygen value | Unit per register 9520; error codes 1.0E37-9.0E37. | FLOAT32 | CDAB | Read | |||
| HR | 9760doc 9760 (0x2620) | Measured temperature | FLOAT32 | CDAB | °C | Read | |||
| HR | 9792doc 9792 (0x2640) | Status word | BITMAP16 | Read | bit 0 = Invalid oxygen value (warning); bit 1 = Invalid temperature value (warning); bit 2 = Sensor cap missing (alarm); bit 8 = Oxygen outside measuring range (warning); bit 9 = Temperature outside measuring range (warning); bit 10 = Cap replacement recommended (warning) |
Notes on this map
- Copy hosted by the Australian distributor Onetemp; JUMO does not offer a public direct link to this document. The file is the unaltered JUMO interface description.
- Float order is selectable in register 775. In "Standard Modbus format" (0) the low word comes first (CDAB): the guide says bytes 1-2 and 3-4 are swapped and its temperature example returns 00 00 41 A2 = 20.25 °C. The guide does not state which setting is the factory default, so read 775 first.
- Invalid float readings are replaced by error codes: 1.0E37 underrange, 2.0E37 overrange, 3.0E37 invalid input, 4.0E37 division by zero, 5.0E37 math error, 6.0E37 invalid compensation temperature, 7.0E37 probe short circuit, 8.0E37 probe break, 9.0E37 timeout.
- Write cyclic compensation values only to the RAM registers (9648-9653); the flash values (9600-9603) must not be written cyclically.
- Left out: operating hours and other Uint32 counters (no stated word order), date/time and text fields, the analog output warning/error enable flags (char booleans) and the out-of-range temperature output value.
- Default serial settings are not stated in the document.
Check it against the live device. A register map tells you what the controller can expose; the unit in front of you may differ by firmware or options. Read a few registers and compare them with the unit’s own display; for a 32-bit value the float and byte-order decoder shows all four word orders at once, so a wrong word order or scale is obvious. New to this? Read what a register map is and why addresses are off by one.
Questions about the JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU)
What Modbus registers does the JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU) have?
This map lists 29 registers: 29 holding registers. 19 of them can be written. Registers vary with firmware, options and configuration, so confirm against the device.
Are the JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU) Modbus addresses 0-based or 1-based?
The manufacturer's document numbers registers 0-based (the number printed is the address on the wire). Every address on this page is the 0-based wire address you put in the request; the documented number is shown under it where it differs. Tables give the Modbus PDU address in decimal and hex; it is sent unchanged (the example reads temperature at 0x2620).
What byte order does the JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU) use for 32-bit values?
Multi-register values in this map use CDAB (low word first (word-swapped)). Set your client to this order for every 32/64-bit value in the table.
What are the minimum Modbus registers to integrate the JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU)?
Start with HR 9728 Measured oxygen value (Oxygen value); HR 9760 Measured temperature (Temperature); HR 9792 Status word (Status / warnings); HR 775 Float value format (Float format in use); HR 9520 Oxygen measurement unit (Oxygen unit).
How do I enable Modbus control of the JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU)?
Modbus RTU is available directly on the sensor RS-485 line; set address and format with JUMO DSM software or registers 768-775.
Where is the official JUMO digiLine O-DO S10 optical oxygen sensor (Modbus RTU) Modbus documentation?
The source is JUMO digiLine O-DO S10 - Oxygen sensor with digital interface - Modbus interface description (20261400T92Z001K000 V2.00/EN/00753777), revision 2021-08-20. JUMO doesn't publish it on its own website, so this map was re-typed from a free copy at https://www.onetemp.com.au/wp-content/uploads/2024/02/digiLine-O-DO-S10-Digital-DO-Sensor-Modbus-Interface-Description.pdf. It may be an older revision, so check it against the live unit.
More JUMO Modbus register maps
- Sensor (turbidity)
JUMO ecoLine NTU digital turbidity sensor (Modbus RTU)
JUMO ecoLine NTU type 202670 with RS-485 Modbus RTU (interface description V1.01).
Also known as: 202670, JUMO ecoLine NTU, ecoLine NTU type 202670 - Sensor (dissolved oxygen)
JUMO ecoLine O-DO digital optical dissolved oxygen sensor (Modbus RTU)
JUMO ecoLine O-DO type 202613 with RS-485 Modbus RTU (interface description V1.01).
Also known as: 202613, JUMO ecoLine O-DO, ecoLine O-DO type 202613 - Sensor (disinfectant / chlorine)
JUMO tecLine digital disinfection sensors (Cl2, TC, ClO2/O3, H2O2/PAA, Br, Cl2 OM) (Modbus RTU)
JUMO tecLine 202630, 202631, 202634, 202636, 202637 and 202681 sensors with RS-485 Modbus RTU (interface description V2.00, 2022-03-24).
Also known as: 202630, 202631, 202634, 202636, 202637, 202681, tecLine Cl2, tecLine TC, tecLine ClO2, tecLine O3, tecLine H2O2, tecLine PAA, tecLine Br, tecLine Cl2 OM
Other valves, actuators & sensors register maps
- Sensor (load-cell weight indicator)
Laumas WLIGHT weight indicator / transmitter (Modbus RTU)
Laumas WLIGHT, installation and user manual version 1.03.
Also known as: WLIGHT, W LIGHT - Sensor (CO2, humidity, temperature module)
Sensirion SCD30 CO2, humidity and temperature sensor module
Sensirion SCD30 sensor module with the SEL pin pulled high at power-up (Modbus mode), per interface description version 1.0.
Also known as: SCD30 - Sensor (flow transmitter)
Siemens SITRANS F C MASS 6000 with USM II Modbus RTU module
SITRANS F C MASS 6000 Coriolis flow transmitters fitted with the USM II Modbus RTU RS-485 add-on module (operating instructions 03/2021).
Also known as: MASS 6000, SITRANS FC MASS 6000 - Sensor (relay output module)
Teracom TDO340 4-channel Modbus RTU relay output module
TDO340 per the Teracom TDO340 user manual Revision R1.2, firmware V1.03 or later.
Also known as: TDO340 - Sensor (AC/DC voltage transducer)
Teracom TSV300 AC/DC voltage transducer with Modbus RTU
TSV300 per the Teracom TSV300 user manual Revision R1.4, firmware V1.06 or later.
Also known as: TSV300 - Sensor (CO2 probe)
Vaisala GMP252 CARBOCAP carbon dioxide probe (Modbus RTU)
GMP252 probe in Modbus RTU mode, user guide M211897EN-G.
Also known as: GMP252, CARBOCAP GMP252, Vaisala GMP252
All valves, actuators & sensors Modbus register maps →
Source: JUMO digiLine O-DO S10 - Oxygen sensor with digital interface - Modbus interface description · 20261400T92Z001K000 V2.00/EN/00753777 · rev. 2021-08-20 · unofficial copy. Register numbers, types, scaling and units are reproduced as published; descriptions are our own.