Sensor (infrared gas analyzer) · Modbus register map
Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU) Modbus register map
Updated 17 November 2026
The Teledyne 7500 and 7600 are multi-component infrared gas analyzers with a Modbus RTU port on RS-232 (RS-485 through an external converter). Each channel reports an integer concentration plus separate decimal-point and unit registers. This map covers all channel readings, range selection and the key/display command registers.
This page lists 43 Modbus registers for the Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU) — 7 writable — with the table, 0-based wire address, data type, word order, scale, units and access of each, re-typed from Teledyne Analytical Instruments’s own free documentation (linked below) and checked against it.
- Manufacturer
- Teledyne Analytical Instruments
- Model
- 7500 / 7600 infrared gas analyzer (Modbus RTU)
- Equipment type
- Sensor (infrared gas analyzer)
- Registers
- 43 (7 writable)
- Transport
- Modbus RTU
- Default unit ID
- Not stated
- Default serial
- 9600 baud, 8N1
- Documented addressing
- Modicon 3xxxx/4xxxx (drop the prefix, subtract 1)
- Also known as
- Teledyne 7500, Teledyne 7600, Model 7500, Model 7600
Integration notes: Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU)
Read input registers 0-23 (7500) or 0-35 (7600) in one function-04 request; for each channel combine concentration, decimal position and unit.
Minimum registers to map
Addresses are 0-based wire addresses. Enough to monitor and run the unit; map more as the job needs.
| Register | Name | Type | Why | Access |
|---|---|---|---|---|
| IR 0 | CH1 concentration | INT16 | CH1 concentration | Read |
| IR 1 | CH1 decimal point position | UINT16 | CH1 decimal position | Read |
| IR 2 | CH1 measurement unit | UINT16 | CH1 unit | Read |
| IR 3 | CH2 concentration | INT16 | CH2 concentration | Read |
| IR 4 | CH2 decimal point position | UINT16 | CH2 decimal position | Read |
Enable Modbus control & writes
Set the analyzer's station number in its communication settings menu; all other serial settings are fixed.
Byte order & addressing
- Concentration needs the separate decimal register to scale.
- Port is RS-232; use a converter for multi-drop RS-485.
Network setup
- 9600 8N1 RTU; give each analyzer a unique station number.
Watch out for
- Writing 64 (ZERO) or 128 (SPAN) to the keying register starts a calibration action.
Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU) Modbus registers
Address = 0-based wire address (the documented number is shown under it when it differs) · Order: ABCD = high word first, CDAB = low word first · Scale: engineering value = raw × scale.
| Table | Address | Name | Description | Type | Order | Scale | Units | Access | Values / bits |
|---|---|---|---|---|---|---|---|---|---|
| IR | 0doc 30001 | CH1 concentration | Reading without decimal point (-9999 to 9999); apply the CH1 decimal register. | INT16 | Read | ||||
| IR | 1doc 30002 | CH1 decimal point position | Decimal places of the CH1 reading. | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 2doc 30003 | CH1 measurement unit | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | ||||
| IR | 3doc 30004 | CH2 concentration | Reading without decimal point (-9999 to 9999); apply the CH2 decimal register. | INT16 | Read | ||||
| IR | 4doc 30005 | CH2 decimal point position | Decimal places of the CH2 reading. | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 5doc 30006 | CH2 measurement unit | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | ||||
| IR | 6doc 30007 | CH3 concentration | Reading without decimal point (-9999 to 9999); apply the CH3 decimal register. | INT16 | Read | ||||
| IR | 7doc 30008 | CH3 decimal point position | Decimal places of the CH3 reading. | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 8doc 30009 | CH3 measurement unit | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | ||||
| IR | 9doc 30010 | CH4 concentration | Reading without decimal point (-9999 to 9999); apply the CH4 decimal register. | INT16 | Read | ||||
| IR | 10doc 30011 | CH4 decimal point position | Decimal places of the CH4 reading. | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 11doc 30012 | CH4 measurement unit | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | ||||
| IR | 12doc 30013 | CH5 concentration | Reading without decimal point (-9999 to 9999); apply the CH5 decimal register. | INT16 | Read | ||||
| IR | 13doc 30014 | CH5 decimal point position | Decimal places of the CH5 reading. | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 14doc 30015 | CH5 measurement unit | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | ||||
| IR | 15doc 30016 | CH6 concentration | Reading without decimal point (-9999 to 9999); apply the CH6 decimal register. | INT16 | Read | ||||
| IR | 16doc 30017 | CH6 decimal point position | Decimal places of the CH6 reading. | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 17doc 30018 | CH6 measurement unit | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | ||||
| IR | 18doc 30019 | CH7 concentration | Reading without decimal point (-9999 to 9999); apply the CH7 decimal register. | INT16 | Read | ||||
| IR | 19doc 30020 | CH7 decimal point position | Decimal places of the CH7 reading. | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 20doc 30021 | CH7 measurement unit | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | ||||
| IR | 21doc 30022 | CH8 concentration | Reading without decimal point (-9999 to 9999); apply the CH8 decimal register. | INT16 | Read | ||||
| IR | 22doc 30023 | CH8 decimal point position | Decimal places of the CH8 reading. | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 23doc 30024 | CH8 measurement unit | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | ||||
| IR | 24doc 30025 | CH9 concentration | Reading without decimal point (-9999 to 9999); apply the CH9 decimal register (7600 only). | INT16 | Read | ||||
| IR | 25doc 30026 | CH9 decimal point position | Decimal places of the CH9 reading (7600 only). | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 26doc 30027 | CH9 measurement unit | 7600 only. | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | |||
| IR | 27doc 30028 | CH10 concentration | Reading without decimal point (-9999 to 9999); apply the CH10 decimal register (7600 only). | INT16 | Read | ||||
| IR | 28doc 30029 | CH10 decimal point position | Decimal places of the CH10 reading (7600 only). | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 29doc 30030 | CH10 measurement unit | 7600 only. | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | |||
| IR | 30doc 30031 | CH11 concentration | Reading without decimal point (-9999 to 9999); apply the CH11 decimal register (7600 only). | INT16 | Read | ||||
| IR | 31doc 30032 | CH11 decimal point position | Decimal places of the CH11 reading (7600 only). | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 32doc 30033 | CH11 measurement unit | 7600 only. | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | |||
| IR | 33doc 30034 | CH12 concentration | Reading without decimal point (-9999 to 9999); apply the CH12 decimal register (7600 only). | INT16 | Read | ||||
| IR | 34doc 30035 | CH12 decimal point position | Decimal places of the CH12 reading (7600 only). | UINT16 | Read | 0 = No decimals; 1 = Divide by 10; 2 = Divide by 100; 3 = Divide by 1000 | |||
| IR | 35doc 30036 | CH12 measurement unit | 7600 only. | UINT16 | Read | 0 = vol%; 1 = ppm; 2 = mg/m3; 3 = g/m3 | |||
| HR | 105doc 40106 | CH1 range change setting | Ignored while remote range is on. | UINT16 | Read/write | 0 = Range 1; 1 = Range 2 | |||
| HR | 106doc 40107 | CH2 range change setting | Ignored while remote range is on. | UINT16 | Read/write | 0 = Range 1; 1 = Range 2 | |||
| HR | 107doc 40108 | CH3 range change setting | Ignored while remote range is on. | UINT16 | Read/write | 0 = Range 1; 1 = Range 2 | |||
| HR | 108doc 40109 | CH4 range change setting | Ignored while remote range is on. | UINT16 | Read/write | 0 = Range 1; 1 = Range 2 | |||
| HR | 109doc 40110 | CH5 range change setting | Ignored while remote range is on. 7600 only. | UINT16 | Read/write | 0 = Range 1; 1 = Range 2 | |||
| HR | 2000doc 42001 | Keying command | Simulates a front-panel key press. | UINT16 | Write | 1 = MODE; 2 = Side; 4 = Up; 8 = Down; 16 = ESC; 32 = ENT; 64 = ZERO; 128 = SPAN | |||
| HR | 2001doc 42002 | Display change | UINT16 | Write | 1 = Return to measurement display |
Notes on this map
- Concentrations are sent without a decimal point. Divide by 10^n where n is the channel's decimal-point register (0-3); the unit register gives vol%, ppm, mg/m3 or g/m3. Example in the manual: 1270 with decimal 2 and unit 0 = 12.70 vol%. Because the decimal place is per channel and range, the concentration registers are listed unscaled.
- Relevant model per the address map: R = 7500, K = 7600. Channel registers CH1-CH8 are marked for both models; CH9-CH12 and the CH5 range-change setting are 7600 only.
- Communication is fixed at 9600 bps, 8 data bits, no parity, 1 stop bit, RTU, over RS-232; up to 31 analyzers can share an RS-485 line through a converter.
- Readings are transmitted even when the display shows over-range dashes.
- Left out: the alarm/calibration status block from 30037 upward, whose rows are out of line in the source table, and the user-setting block 40001-40105 (calibration, alarm and timer settings).
Check it against the live device. A register map tells you what the controller can expose; the unit in front of you may differ by firmware or options. Read a few registers and compare them with the unit’s own display; for a 32-bit value the float and byte-order decoder shows all four word orders at once, so a wrong word order or scale is obvious. New to this? Read what a register map is and why addresses are off by one.
Questions about the Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU)
What Modbus registers does the Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU) have?
This map lists 43 registers: 36 input registers, 7 holding registers. 7 of them can be written. Registers vary with firmware, options and configuration, so confirm against the device.
Are the Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU) Modbus addresses 0-based or 1-based?
The manufacturer's document numbers registers Modicon 3xxxx/4xxxx (drop the prefix, subtract 1). Every address on this page is the 0-based wire address you put in the request; the documented number is shown under it where it differs. The manual prints a register number (30001, 40001, 42001) and a relative address (0000H, 07D0H) and says the relative address sent in the message is the lower four digits of the register number minus 1.
What are the minimum Modbus registers to integrate the Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU)?
Start with IR 0 CH1 concentration (CH1 concentration); IR 1 CH1 decimal point position (CH1 decimal position); IR 2 CH1 measurement unit (CH1 unit); IR 3 CH2 concentration (CH2 concentration); IR 4 CH2 decimal point position (CH2 decimal position).
How do I enable Modbus control of the Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU)?
Set the analyzer's station number in its communication settings menu; all other serial settings are fixed.
Where is the official Teledyne Analytical Instruments 7500 / 7600 infrared gas analyzer (Modbus RTU) Modbus documentation?
The source is Instruction Manual - Infrared Gas Analyzer Communication Functions (Modbus), Type 7500 / 7600, published free by Teledyne Analytical Instruments at https://teledyne-ai.com/en-us/Products_/Documents/Manuals/man_75007600comm.pdf.
More Teledyne Analytical Instruments Modbus register maps
Other gas detectors register maps
- Sensor (fixed gas detector)
Teledyne Detcon Model TP-700 MOS (metal oxide) H2S sensor (RS-485 Modbus)
Teledyne Detcon TP-700 smart sensors with the 4-20 mA and RS-485 Modbus outputs, instruction manual Document #3150, Rev. 4.6.
Also known as: TP-700, Detcon TP-700, Teledyne Detcon TP-700 - Sensor (CO2 / gas)
CO2Meter CM500 CO2 sensor / controller (RS-485 Modbus RTU)
CO2Meter.com CM500 with the RS-485 Modbus RTU interface, per the vendor's 'RS485 Communications Modbus Protocol' sheet (no document number).
Also known as: CM500, CO2Meter CM500 - Sensor (environmental / leak monitor)
RLE Technologies Falcon F200 environment and leak monitoring appliance (Modbus TCP)
RLE Falcon F200 with firmware 4100.1 or later; register guide v06.21.
Also known as: F200, Falcon F200 - Sensor (fixed gas detector)
Teledyne Detcon Model IR-700 infrared combustible / CO2 gas sensor (RS-485 Modbus)
Teledyne Detcon IR-700 smart sensors with the 4-20 mA and RS-485 Modbus outputs, instruction manual Document #3169, Rev. 4.9.
Also known as: IR-700, Detcon IR-700, Teledyne Detcon IR-700 - Sensor (refrigerant leak detector)
Amphenol Telaire T3400 series refrigerant (A2L) leak sensor
Amphenol Telaire T3400 series refrigerant sensors (R454B / R32) with RS-485 Modbus RTU, user instructions AS-T63381B (04/2025, marked provisional).
Also known as: Telaire T3400, T3400 R454B, T3400 R32 - Sensor (fixed gas detector)
New Cosmos Electric PS-7-M series extractive pump gas detector (Modbus RTU / TCP)
New Cosmos PS-7-M series extractive pump gas detectors (main unit with base unit) over RS-485 Modbus RTU or Ethernet Modbus TCP, communication specifications guide GAE-159-01.
Also known as: PS-7-M, PS-7
All gas detectors Modbus register maps →
Source: Instruction Manual - Infrared Gas Analyzer Communication Functions (Modbus), Type 7500 / 7600. Register numbers, types, scaling and units are reproduced as published; descriptions are our own.