Sensor (water analysis transmitter) · Modbus register map

Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP) Modbus register map

Updated 17 November 2026

Short answer

The Liquiline CM44x is Endress+Hauser's multichannel transmitter for Memosens pH, ORP, conductivity, oxygen, turbidity and similar sensors. With the Modbus option it exposes 16 configurable analog and 8 digital device variables, 4 analog and 8 digital write-back variables, diagnostics and per-sensor information blocks. This map covers those common blocks, with the sensor block shown for sensor 1.

This page lists 221 Modbus registers for the Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP) — 28 writable — with the table, 0-based wire address, data type, word order, scale, units and access of each, re-typed from Endress+Hauser’s own free documentation (linked below) and checked against it.

Manufacturer
Endress+Hauser
Model
Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP)
Equipment type
Sensor (water analysis transmitter)
Registers
221 (28 writable)
Transport
Modbus RTU, Modbus ASCII, Modbus TCP
Default unit ID
247
Default serial
19200 baud, 8E1
TCP port
502
Documented addressing
0-based (the number printed is the address on the wire)
32-bit word order
CDAB low word first (word-swapped)
Firmware
Software 1.11.00 and later
Also known as
CM442, CM444, CM448, CM44P, CM44R, Liquiline CM442, Liquiline CM444, Liquiline CM448
Source document
Liquiline CM44x / Liquistation CSFxx / Liquiline System CA80xx - Data transmission via Modbus (Special Documentation) · SD01189C/07/EN/07.22 · rev. 2022-07
Download CSV all columns JSON Manufacturer PDF ↗

Integration notes: Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP)

Integration in brief

Assign the measured values you need to AI 1-16 in the transmitter menu, then read registers 0-63 (value, status, unit for each AI) with function 03 or 04 and decode floats as CDAB. Check each status byte is 0 before using the value.

Minimum registers to map

Addresses are 0-based wire addresses. Enough to monitor and run the unit; map more as the job needs.

RegisterNameTypeWhyAccess
HR 0MODBUS_AI_1_VALUEFLOAT32 CDABAI 1 value (e.g. main measured value)Read
HR 2MODBUS_AI_1_STATUSUINT16AI 1 statusRead
HR 3MODBUS_AI_1_UNITUINT16AI 1 unit codeRead
HR 4MODBUS_AI_2_VALUEFLOAT32 CDABAI 2 value (e.g. temperature)Read
HR 6MODBUS_AI_2_STATUSUINT16AI 2 statusRead
HR 112MODBUS_DI_1_VALUEUINT16DI 1 value (e.g. limit contact)Read

Enable Modbus control & writes

Menu Setup/General settings/Extended setup/Modbus/Settings: bus address (default 247, or set by the module's address switch), RTU/ASCII, baud (default 19200), parity (default even, 1 stop bit), byte order and watchdog. For TCP set the IP under .../Ethernet/Settings (DHCP by default) and use port 502. Assign sources to the device variables under Setup/Outputs.

Communication loss & watchdog

Modbus watchdog default 5 s (0-999 s, 0 = off), set in the Modbus settings menu. If no valid request arrives in that time the device raises diagnostic S969 and lights the red BF LED until the next valid request.

Byte order & addressing

Network setup

Watch out for

Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP) Modbus registers

Address = 0-based wire address (the documented number is shown under it when it differs) · Order: ABCD = high word first, CDAB = low word first · Scale: engineering value = raw × scale.

221 of 221 shown
Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP) Modbus register map
TableAddressNameDescriptionTypeOrderScaleUnitsAccessValues / bits
IR522MODBUS_FRAMING_ERRORSUINT32ABCDRead
IR524MODBUS_CRC_ERRORSUINT32ABCDRead
IR526MODBUS_PARITY_ERRORSUINT32ABCDRead
IR528MODBUS_01_EXCEPTIONSUINT32ABCDRead
IR530MODBUS_02_EXCEPTIONSUINT32ABCDRead
IR532MODBUS_03_EXCEPTIONSUINT32ABCDRead
IR534MODBUS_04_EXCEPTIONSUINT32ABCDRead
IR536ETHERNET_RECEIVED_FRAMESUINT32ABCDRead
IR538ETHERNET_SEND_FRAMESUINT32ABCDRead
IR540ETHERNET_VALID_FRAMESUINT32ABCDRead
IR542ETHERNET_INVALID_FRAMESUINT32ABCDRead
IR544ETHERNET_CRC_ERRORSUINT32ABCDRead
IR546DEVICE_SW_VERSION16 characters.STRING[16]Read
IR554DEVICE_BOOTLOADER_VERSION16 characters.STRING[16]Read
IR562DEVICE_SERIAL32 characters.STRING[32]Read
IR578DISPLAY_SER_NUM32 characters.STRING[32]Read
IR626BASE_MODUL_SER_NUM32 characters.STRING[32]Read
IR642EXTENSION_MODUL_1_SER_NUM32 characters.STRING[32]Read
IR658EXTENSION_MODUL_2_SER_NUM32 characters.STRING[32]Read
IR674EXTENSION_MODUL_3_SER_NUM32 characters.STRING[32]Read
IR738EXTENSION_MODUL_7_SER_NUM32 characters.STRING[32]Read
IR754EXTENSION_MODUL_8_SER_NUM32 characters.STRING[32]Read
IR770SENSOR_1_SER_NUM32 characters.STRING[32]Read
IR786SENSOR_2_SER_NUM32 characters.STRING[32]Read
IR802SENSOR_3_SER_NUM32 characters.STRING[32]Read
IR818SENSOR_4_SER_NUM32 characters.STRING[32]Read
IR834SENSOR_5_SER_NUM32 characters.STRING[32]Read
IR850SENSOR_6_SER_NUM32 characters.STRING[32]Read
IR866SENSOR_7_SER_NUM32 characters.STRING[32]Read
IR882SENSOR_8_SER_NUM32 characters.STRING[32]Read
IR898FMSY1_MODUL_SER_NUM32 characters.STRING[32]Read
IR941RS485_TERMINATION_STATE8-bit value in one register.UINT16Read0 = Off; 1 = On
IR942ETHERNET_MAC6 characters.STRING[6]Read
IR945CLEANINGACTIONSTATE_18-bit value in one register.UINT16Read0 = Off; 1 = Waiting; 2 = Waiting; 3 = Pre-rinse; 4 = Cleaning; 5 = Postrinsing; 6 = Failsafe; 7 = Manual mode; 8 = Waiting for execution
IR946CLEANINGACTIONSTATE_28-bit value in one register.UINT16Read0 = Off; 1 = Waiting; 2 = Waiting; 3 = Pre-rinse; 4 = Cleaning; 5 = Postrinsing; 6 = Failsafe; 7 = Manual mode; 8 = Waiting for execution
IR947CLEANINGACTIONSTATE_38-bit value in one register.UINT16Read0 = Off; 1 = Waiting; 2 = Waiting; 3 = Pre-rinse; 4 = Cleaning; 5 = Postrinsing; 6 = Failsafe; 7 = Manual mode; 8 = Waiting for execution
IR948CLEANINGACTIONSTATE_48-bit value in one register.UINT16Read0 = Off; 1 = Waiting; 2 = Waiting; 3 = Pre-rinse; 4 = Cleaning; 5 = Postrinsing; 6 = Failsafe; 7 = Manual mode; 8 = Waiting for execution
IR953HEARTBEAT_DEVICE_STATUSFLOAT32CDABRead
IR955HEARTBEAT_DEVICE_HEALTH_CONDITIONFLOAT32CDAB%Read
IR957MAINTENANCE_TIMER_VALUEFLOAT32CDABRead
IR959HEARTBEAT_INDEX_DEVICEFLOAT32CDABRead
IR965KEY_PERF_INDICATOR_DEVICE_AVAILABILITYFLOAT32CDAB%Read
IR975KEY_PERF_INDICATOR_DEVICE_NUMBER_OF_FAILURESUINT16Read
IR1019doc X019 (sensor 1)SENSOR_SERIAL_CH0Sensor 1 block. 16 characters.STRING[16]Read
IR1027doc X027 (sensor 1)SENSOR_HW_VERSION_CH0Sensor 1 block. 16 characters.STRING[16]Read
IR1035doc X035 (sensor 1)SENSOR_SW_VERSION_CH0Sensor 1 block. 16 characters.STRING[16]Read
IR1043doc X043 (sensor 1)SENSOR_MANUFACTURER_CH0Sensor 1 block. 32 characters.STRING[32]Read
IR1059doc X059 (sensor 1)SENSOR_ORDERCODE_CH0Sensor 1 block. 32 characters.STRING[32]Read
IR1083doc X083 (sensor 1)SENSOR_EXTREME_SPECIFIC_MIN_CH0Sensor 1 block.FLOAT32CDABRead
IR1085doc X085 (sensor 1)SENSOR_EXTREME_TEMP_MIN_CH0Sensor 1 block.FLOAT32CDAB°CRead
IR1087doc X087 (sensor 1)SENSOR_EXTREME_SPECIFIC_MAX_CH0Sensor 1 block.FLOAT32CDABRead
IR1089doc X089 (sensor 1)SENSOR_EXTREME_TEMP_MAX_CH0Sensor 1 block.FLOAT32CDAB°CRead
IR1091doc X091 (sensor 1)SENSOR_CURRENT_DIAG_SOURCETYPE_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = General; 1 = Temperature Input; 2 = Binary Input; 3 = Binary Output; 4 = Current Input; 5 = Current Output; 6 = Relay; 7 = Memosens Sensor; 8 = PID Controller; 9 = Cleaning; 10 = Limit switch; 11 = Binary formula result; 12 = Binary Input Module; 13 = Binary Output Module; 14 = Sample point
IR1092doc X092 (sensor 1)SENSOR_CURRENT_DIAG_MODUL_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = Slot 1; 1 = Slot 1; 2 = Slot 2; 3 = Slot 3; 4 = Slot 4; 5 = Slot 5; 6 = Slot 6; 7 = Slot 7; 8 = Sampler; 9 = Software; 10 = Analyzer
IR1093doc X093 (sensor 1)SENSOR_CURRENT_DIAG_PORT_CH0Sensor 1 block. 8-bit value in one register.UINT16Read
IR1094doc X094 (sensor 1)SENSOR_CURRENT_DIAG_NE107_STATUS_CH0Sensor 1 block. NAMUR NE 107 category bits.BITMAP16Readbit 0 = Failure (F); bit 1 = Function check (C); bit 2 = Maintenance required (M); bit 3 = Out of specification (S)
IR1095doc X095 (sensor 1)SENSOR_CURRENT_DIAG_CODE_CH0Sensor 1 block.UINT16Read
IR1096doc X096 (sensor 1)SENSOR_CURRENT_DIAG_TIMESTAMP_CH0Sensor 1 block.UINT32ABCDRead
IR1098doc X098 (sensor 1)SENSOR_NUMBER_ADDITIONAL_DIAG_CH0Sensor 1 block. 8-bit value in one register.UINT16Read
IR1099doc X099 (sensor 1)SENSOR_PAST_DIAG_SOURCETYPE_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = General; 1 = Temperature Input; 2 = Binary Input; 3 = Binary Output; 4 = Current Input; 5 = Current Output; 6 = Relay; 7 = Memosens Sensor; 8 = PID Controller; 9 = Cleaning; 10 = Limit switch; 11 = Binary formula result; 12 = Binary Input Module; 13 = Binary Output Module; 14 = Sample preparation
IR1100doc X100 (sensor 1)SENSOR_PAST_DIAG_MODUL_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = Slot 1; 1 = Slot 1; 2 = Slot 2; 3 = Slot 3; 4 = Slot 4; 5 = Slot 5; 6 = Slot 6; 7 = Slot 7; 8 = Sampler; 9 = Software; 10 = Analyzer
IR1101doc X101 (sensor 1)SENSOR_PAST_DIAG_PORT_CH0Sensor 1 block. 8-bit value in one register.UINT16Read
IR1102doc X102 (sensor 1)SENSOR_PAST_DIAG_NE107_STATUS_CH0Sensor 1 block. NAMUR NE 107 category bits.BITMAP16Readbit 0 = Failure (F); bit 1 = Function check (C); bit 2 = Maintenance required (M); bit 3 = Out of specification (S)
IR1103doc X103 (sensor 1)SENSOR_PAST_DIAG_CODE_CH0Sensor 1 block.UINT16Read
IR1104doc X104 (sensor 1)SENSOR_PAST_DIAG_TIMESTAMP_CH0Sensor 1 block.UINT32ABCDRead
IR1106doc X106 (sensor 1)OP_TIME_TOTAL_CH0Sensor 1 block.FLOAT32CDABsRead
IR1108doc X108 (sensor 1)OP_TIME_ABOVE_SPEC_TEMP_CH0Sensor 1 block.FLOAT32CDABsRead
IR1110doc X110 (sensor 1)OP_TIME_BELOW_SPEC_TEMP_CH0Sensor 1 block.FLOAT32CDABsRead
IR1112doc X112 (sensor 1)OP_TIME_LAMP_LIFE_CH0Sensor 1 block.FLOAT32CDABRead
IR1114doc X114 (sensor 1)OP_TIME_STERILISATIONS_CH0Sensor 1 block.UINT16Read
IR1115doc X115 (sensor 1)OP_TIME_CIP_CYCLES_CH0Sensor 1 block.UINT16Read
IR1116doc X116 (sensor 1)OP_TIME_CAP_CALIBRATIONS_CH0Sensor 1 block.UINT16Read
IR1117doc X117 (sensor 1)OP_TIME_CAP_CALIB_TIMER_CH0Sensor 1 block.FLOAT32CDABsRead
IR1119doc X119 (sensor 1)OP_TIME_CAP_STERILISATIONS_CH0Sensor 1 block.UINT16Read
IR1120doc X120 (sensor 1)OP_TIME_CAP_FILTER_CHANGED_CH0Sensor 1 block.FLOAT32CDABsRead
IR1122doc X122 (sensor 1)CAL_COUNT_0_CH0Sensor 1 block.UINT16Read
IR1123doc X123 (sensor 1)CAL_COUNT_1_CH0Sensor 1 block.UINT16Read
IR1124doc X124 (sensor 1)CAL_COUNT_2_CH0Sensor 1 block.UINT16Read
IR1125doc X125 (sensor 1)CAL_COUNT_3_CH0Sensor 1 block.UINT16Read
IR1142doc X142 (sensor 1)CAL_METHOD_0_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = None; 1 = Numeric input; 2 = 1-point calibration; 3 = 2-point calibration; 4 = Multipoint Calibration; 5 = Table; 6 = Grab sample; 7 = Zero point; 8 = Air 100%% rh; 9 = H2O air-saturated; 10 = Air variable; 11 = Zero point sample; 12 = Slope sample; 13 = Offset; 14 = Slope; 15 = with temp. comp; 16 = without temp. comp; 17 = Standard addition; 18 = Numeric input; 19 = Numeric input zero point; 20 = Numeric input; 21 = Factory calib.; 22 = Test gas calibration; 23 = Value acquisition; 24 = 3-point calibration; 25 = Ext. Reference; 26 = Int. Reference; 27 = in water; 28 = in air
IR1143doc X143 (sensor 1)CAL_METHOD_1_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = None; 1 = Numeric input; 2 = 1-point calibration; 3 = 2-point calibration; 4 = Multipoint Calibration; 5 = Table; 6 = Grab sample; 7 = Zero point; 8 = Air 100%% rh; 9 = H2O air-saturated; 10 = Air variable; 11 = Zero point sample; 12 = Slope sample; 13 = Offset; 14 = Slope; 15 = with temp. comp; 16 = without temp. comp; 17 = Standard addition; 18 = Numeric input; 19 = Numeric input zero point; 20 = Numeric input; 21 = Factory calib.; 22 = Test gas calibration; 23 = Value acquisition; 24 = 3-point calibration; 25 = Ext. Reference; 26 = Int. Reference; 27 = in water; 28 = in air
IR1144doc X144 (sensor 1)CAL_METHOD_2_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = None; 1 = Numeric input; 2 = 1-point calibration; 3 = 2-point calibration; 4 = Multipoint Calibration; 5 = Table; 6 = Grab sample; 7 = Zero point; 8 = Air 100%% rh; 9 = H2O air-saturated; 10 = Air variable; 11 = Zero point sample; 12 = Slope sample; 13 = Offset; 14 = Slope; 15 = with temp. comp; 16 = without temp. comp; 17 = Standard addition; 18 = Numeric input; 19 = Numeric input zero point; 20 = Numeric input; 21 = Factory calib.; 22 = Test gas calibration; 23 = Value acquisition; 24 = 3-point calibration; 25 = Ext. Reference; 26 = Int. Reference; 27 = in water; 28 = in air
IR1145doc X145 (sensor 1)CAL_METHOD_3_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = None; 1 = Numeric input; 2 = 1-point calibration; 3 = 2-point calibration; 4 = Multipoint Calibration; 5 = Table; 6 = Grab sample; 7 = Zero point; 8 = Air 100%% rh; 9 = H2O air-saturated; 10 = Air variable; 11 = Zero point sample; 12 = Slope sample; 13 = Offset; 14 = Slope; 15 = with temp. comp; 16 = without temp. comp; 17 = Standard addition; 18 = Numeric input; 19 = Numeric input zero point; 20 = Numeric input; 21 = Factory calib.; 22 = Test gas calibration; 23 = Value acquisition; 24 = 3-point calibration; 25 = Ext. Reference; 26 = Int. Reference; 27 = in water; 28 = in air
IR1146doc X146 (sensor 1)CAL_COUNT_ZERO_CH0Sensor 1 block.UINT16Read
IR1151doc X151 (sensor 1)CAL_METHOD_ZERO_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = None; 1 = Numeric input; 2 = 1-point calibration; 3 = 2-point calibration; 4 = Multipoint Calibration; 5 = Table; 6 = Grab sample; 7 = Zero point; 8 = Air 100%% rh; 9 = H2O air-saturated; 10 = Air variable; 11 = Zero point sample; 12 = Slope sample; 13 = Offset; 14 = Slope; 15 = with temp. comp; 16 = without temp. comp; 17 = Standard addition; 18 = Numeric input; 19 = Numeric input zero point; 20 = Numeric input 2-point; 21 = Factory calib.; 22 = Test gas calibration; 23 = Value acquisition; 24 = 3-point calibration; 25 = Ext. Reference; 26 = Int. Reference; 27 = in water; 28 = in air
IR1152doc X152 (sensor 1)CAL_COUNT_TEMP_CH0Sensor 1 block.UINT16Read
IR1157doc X157 (sensor 1)CAL_METHOD_TEMP_CH0Sensor 1 block. 8-bit value in one register.UINT16Read0 = None; 1 = Numeric input; 2 = 1-point calibration; 3 = 2-point calibration; 4 = Multipoint Calibration; 5 = Table; 6 = Grab sample; 7 = Zero point; 8 = Air 100%% rh; 9 = H2O air-saturated; 10 = Air variable; 11 = Zero point sample; 12 = Slope sample; 13 = Offset; 14 = Slope; 15 = with temp. comp; 16 = without temp. comp; 17 = Standard addition; 18 = Numeric input; 19 = Numeric input zero point; 20 = Numeric input; 21 = Factory calib.; 22 = Test gas calibration; 23 = Value acquisition; 24 = 3-point calibration; 25 = Ext. Reference; 26 = Int. Reference; 27 = in water; 28 = in air
IR1158doc X158 (sensor 1)HEARTBEAT_SENSOR_STATUS_CH0Sensor 1 block.FLOAT32CDAB%Read
IR1160doc X160 (sensor 1)HEARTBEAT_SENSOR_HEALTH_CONDITION_CH0Sensor 1 block.FLOAT32CDAB%Read
IR1162doc X162 (sensor 1)MAINTENANCE_TIMER_VALUE_CH0Sensor 1 block.FLOAT32CDAB%Read
IR1164doc X164 (sensor 1)HEARTBEAT_INDEX_SENSOR_CH0Sensor 1 block.FLOAT32CDAB%Read
IR1166doc X166 (sensor 1)KEY_PERF_INDICATOR_SENSOR_AVAILABILITY_CH1Sensor 1 block.FLOAT32CDAB%Read
IR1176doc X176 (sensor 1)KEY_PERF_INDICATOR_SENSOR_NUMBER_OF_FAILURES_CH1Sensor 1 block.UINT16Read
IR1189doc X189 (sensor 1)KEY_PERF_INDICATOR_SENSOR_NUMBER_OF_CALIBRATIONS_CH1Sensor 1 block.UINT16Read
IR9000BUS_ADDRESS8-bit value in one register.UINT16Read
IR9001DEVICE_TAG32 characters.STRING[32]Read
IR9017DEVICE_ID8-bit value in one register.UINT16Read0 = Liquiline CM442; 1 = Liquiline CM448; 2 = Liquiport CSP44; 3 = Liquistation CSF48; 4 = Liquistation CSF22; 5 = Liquistation CSF33; 6 = Liquiline System CA80; 7 = Liquistation CSF34; 8 = Liquistation CSF39; 9 = Liquiline CM44P
HR0MODBUS_AI_1_VALUEFLOAT32CDABRead
HR2MODBUS_AI_1_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR3MODBUS_AI_1_UNITUnit code (table 7.1 of the guide).UINT16Read
HR4MODBUS_AI_2_VALUEFLOAT32CDABRead
HR6MODBUS_AI_2_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR7MODBUS_AI_2_UNITUnit code (table 7.1 of the guide).UINT16Read
HR8MODBUS_AI_3_VALUEFLOAT32CDABRead
HR10MODBUS_AI_3_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR11MODBUS_AI_3_UNITUnit code (table 7.1 of the guide).UINT16Read
HR12MODBUS_AI_4_VALUEFLOAT32CDABRead
HR14MODBUS_AI_4_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR15MODBUS_AI_4_UNITUnit code (table 7.1 of the guide).UINT16Read
HR16MODBUS_AI_5_VALUEFLOAT32CDABRead
HR18MODBUS_AI_5_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR19MODBUS_AI_5_UNITUnit code (table 7.1 of the guide).UINT16Read
HR20MODBUS_AI_6_VALUEFLOAT32CDABRead
HR22MODBUS_AI_6_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR23MODBUS_AI_6_UNITUnit code (table 7.1 of the guide).UINT16Read
HR24MODBUS_AI_7_VALUEFLOAT32CDABRead
HR26MODBUS_AI_7_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR27MODBUS_AI_7_UNITUnit code (table 7.1 of the guide).UINT16Read
HR28MODBUS_AI_8_VALUEFLOAT32CDABRead
HR30MODBUS_AI_8_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR31MODBUS_AI_8_UNITUnit code (table 7.1 of the guide).UINT16Read
HR32MODBUS_AI_9_VALUEFLOAT32CDABRead
HR34MODBUS_AI_9_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR35MODBUS_AI_9_UNITUnit code (table 7.1 of the guide).UINT16Read
HR36MODBUS_AI_10_VALUEFLOAT32CDABRead
HR38MODBUS_AI_10_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR39MODBUS_AI_10_UNITUnit code (table 7.1 of the guide).UINT16Read
HR40MODBUS_AI_11_VALUEFLOAT32CDABRead
HR42MODBUS_AI_11_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR43MODBUS_AI_11_UNITUnit code (table 7.1 of the guide).UINT16Read
HR44MODBUS_AI_12_VALUEFLOAT32CDABRead
HR46MODBUS_AI_12_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR47MODBUS_AI_12_UNITUnit code (table 7.1 of the guide).UINT16Read
HR48MODBUS_AI_13_VALUEFLOAT32CDABRead
HR50MODBUS_AI_13_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR51MODBUS_AI_13_UNITUnit code (table 7.1 of the guide).UINT16Read
HR52MODBUS_AI_14_VALUEFLOAT32CDABRead
HR54MODBUS_AI_14_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR55MODBUS_AI_14_UNITUnit code (table 7.1 of the guide).UINT16Read
HR56MODBUS_AI_15_VALUEFLOAT32CDABRead
HR58MODBUS_AI_15_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR59MODBUS_AI_15_UNITUnit code (table 7.1 of the guide).UINT16Read
HR60MODBUS_AI_16_VALUEFLOAT32CDABRead
HR62MODBUS_AI_16_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR63MODBUS_AI_16_UNITUnit code (table 7.1 of the guide).UINT16Read
HR64CURRENTOUT_01_PV_VALUEFLOAT32CDABRead
HR66CURRENTOUT_01_PV_STATUS8-bit value in one register.UINT16Read
HR67CURRENTOUT_01_PV_UNITUnit code (table 7.1 of the guide).UINT16Read
HR68CURRENTOUT_02_PV_VALUEFLOAT32CDABRead
HR70CURRENTOUT_02_PV_STATUS8-bit value in one register.UINT16Read
HR71CURRENTOUT_02_PV_UNITUnit code (table 7.1 of the guide).UINT16Read
HR72CURRENTOUT_03_PV_VALUEFLOAT32CDABRead
HR74CURRENTOUT_03_PV_STATUS8-bit value in one register.UINT16Read
HR75CURRENTOUT_03_PV_UNITUnit code (table 7.1 of the guide).UINT16Read
HR76CURRENTOUT_04_PV_VALUEFLOAT32CDABRead
HR78CURRENTOUT_04_PV_STATUS8-bit value in one register.UINT16Read
HR79CURRENTOUT_04_PV_UNITUnit code (table 7.1 of the guide).UINT16Read
HR80CURRENTOUT_05_PV_VALUEFLOAT32CDABRead
HR82CURRENTOUT_05_PV_STATUS8-bit value in one register.UINT16Read
HR83CURRENTOUT_05_PV_UNITUnit code (table 7.1 of the guide).UINT16Read
HR84CURRENTOUT_06_PV_VALUEFLOAT32CDABRead
HR86CURRENTOUT_06_PV_STATUS8-bit value in one register.UINT16Read
HR87CURRENTOUT_06_PV_UNITUnit code (table 7.1 of the guide).UINT16Read
HR88CURRENTOUT_07_PV_VALUEFLOAT32CDABRead
HR90CURRENTOUT_07_PV_STATUS8-bit value in one register.UINT16Read
HR91CURRENTOUT_07_PV_UNITUnit code (table 7.1 of the guide).UINT16Read
HR92CURRENTOUT_08_PV_VALUEFLOAT32CDABRead
HR94CURRENTOUT_08_PV_STATUS8-bit value in one register.UINT16Read
HR95CURRENTOUT_08_PV_UNITUnit code (table 7.1 of the guide).UINT16Read
HR96CURRENTOUT_01_MA_VALUECurrent output signal; the table prints the unit as ampere.FLOAT32CDABRead
HR98CURRENTOUT_02_MA_VALUECurrent output signal; the table prints the unit as ampere.FLOAT32CDABRead
HR100CURRENTOUT_03_MA_VALUECurrent output signal; the table prints the unit as ampere.FLOAT32CDABRead
HR102CURRENTOUT_04_MA_VALUECurrent output signal; the table prints the unit as ampere.FLOAT32CDABRead
HR104CURRENTOUT_05_MA_VALUECurrent output signal; the table prints the unit as ampere.FLOAT32CDABRead
HR106CURRENTOUT_06_MA_VALUECurrent output signal; the table prints the unit as ampere.FLOAT32CDABRead
HR108CURRENTOUT_07_MA_VALUECurrent output signal; the table prints the unit as ampere.FLOAT32CDABRead
HR110CURRENTOUT_08_MA_VALUECurrent output signal; the table prints the unit as ampere.FLOAT32CDABRead
HR112MODBUS_DI_1_VALUE8-bit value in one register.UINT16Read0 = False; 1 = True
HR113MODBUS_DI_1_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR114MODBUS_DI_2_VALUE8-bit value in one register.UINT16Read0 = False; 1 = True
HR115MODBUS_DI_2_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR116MODBUS_DI_3_VALUE8-bit value in one register.UINT16Read0 = False; 1 = True
HR117MODBUS_DI_3_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR118MODBUS_DI_4_VALUE8-bit value in one register.UINT16Read0 = False; 1 = True
HR119MODBUS_DI_4_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR120MODBUS_DI_5_VALUE8-bit value in one register.UINT16Read0 = False; 1 = True
HR121MODBUS_DI_5_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR122MODBUS_DI_6_VALUE8-bit value in one register.UINT16Read0 = False; 1 = True
HR123MODBUS_DI_6_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR124MODBUS_DI_7_VALUE8-bit value in one register.UINT16Read0 = False; 1 = True
HR125MODBUS_DI_7_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR126MODBUS_DI_8_VALUE8-bit value in one register.UINT16Read0 = False; 1 = True
HR127MODBUS_DI_8_STATUSQuality of the value.UINT16Read0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR250MODBUS_DO_01_VALUE0 = false, 1-255 = true.UINT16Read/write
HR251MODBUS_DO_01_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR252MODBUS_DO_02_VALUE0 = false, 1-255 = true.UINT16Read/write
HR253MODBUS_DO_02_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR254MODBUS_DO_03_VALUE0 = false, 1-255 = true.UINT16Read/write
HR255MODBUS_DO_03_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR256MODBUS_DO_04_VALUE0 = false, 1-255 = true.UINT16Read/write
HR257MODBUS_DO_04_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR258MODBUS_DO_05_VALUE0 = false, 1-255 = true.UINT16Read/write
HR259MODBUS_DO_05_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR260MODBUS_DO_06_VALUE0 = false, 1-255 = true.UINT16Read/write
HR261MODBUS_DO_06_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR262MODBUS_DO_07_VALUE0 = false, 1-255 = true.UINT16Read/write
HR263MODBUS_DO_07_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR264MODBUS_DO_08_VALUE0 = false, 1-255 = true.UINT16Read/write
HR265MODBUS_DO_08_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR266MODBUS_AO_01_VALUEFLOAT32CDABRead/write
HR268MODBUS_AO_01_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR269MODBUS_AO_01_UNITUnit code (table 7.1 of the guide).UINT16Read/write
HR270MODBUS_AO_02_VALUEFLOAT32CDABRead/write
HR272MODBUS_AO_02_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR273MODBUS_AO_02_UNITUnit code (table 7.1 of the guide).UINT16Read/write
HR274MODBUS_AO_03_VALUEFLOAT32CDABRead/write
HR276MODBUS_AO_03_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR277MODBUS_AO_03_UNITUnit code (table 7.1 of the guide).UINT16Read/write
HR278MODBUS_AO_04_VALUEFLOAT32CDABRead/write
HR280MODBUS_AO_04_STATUSWrite 0 (good) or the device ignores the value.UINT16Read/write0 = Good; 1 = Uncertain; 2 = Bad; 3 = Not assigned
HR281MODBUS_AO_04_UNITUnit code (table 7.1 of the guide).UINT16Read/write

Notes on this map

Check it against the live device. A register map tells you what the controller can expose; the unit in front of you may differ by firmware or options. Read a few registers and compare them with the unit’s own display; for a 32-bit value the float and byte-order decoder shows all four word orders at once, so a wrong word order or scale is obvious. New to this? Read what a register map is and why addresses are off by one.

Questions about the Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP)

What Modbus registers does the Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP) have?

This map lists 221 registers: 97 input registers, 124 holding registers. 28 of them can be written. Registers vary with firmware, options and configuration, so confirm against the device.

Are the Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP) Modbus addresses 0-based or 1-based?

The manufacturer's document numbers registers 0-based (the number printed is the address on the wire). Every address on this page is the 0-based wire address you put in the request; the documented number is shown under it where it differs. The guide states that its register numbers start at 0 exactly as sent on the wire, and that Modicon-style PLCs must add 1. So AI 1 value (register 0) is 40001 in Modicon terms.

What byte order does the Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP) use for 32-bit values?

Multi-register values in this map use ABCD (high word first (big-endian)) and CDAB (low word first (word-swapped)). Each register's order is in the Order column.

What are the minimum Modbus registers to integrate the Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP)?

Start with HR 0 MODBUS_AI_1_VALUE (AI 1 value (e.g. main measured value)); HR 2 MODBUS_AI_1_STATUS (AI 1 status); HR 3 MODBUS_AI_1_UNIT (AI 1 unit code); HR 4 MODBUS_AI_2_VALUE (AI 2 value (e.g. temperature)); HR 6 MODBUS_AI_2_STATUS (AI 2 status); HR 112 MODBUS_DI_1_VALUE (DI 1 value (e.g. limit contact)).

How do I enable Modbus control of the Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP)?

Menu Setup/General settings/Extended setup/Modbus/Settings: bus address (default 247, or set by the module's address switch), RTU/ASCII, baud (default 19200), parity (default even, 1 stop bit), byte order and watchdog. For TCP set the IP under .../Ethernet/Settings (DHCP by default) and use port 502. Assign sources to the device variables under Setup/Outputs.

What happens when the Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP) loses Modbus communication?

Modbus watchdog default 5 s (0-999 s, 0 = off), set in the Modbus settings menu. If no valid request arrives in that time the device raises diagnostic S969 and lights the red BF LED until the next valid request.

Where is the official Endress+Hauser Liquiline CM44x multichannel transmitter (Modbus RS485 / TCP) Modbus documentation?

The source is Liquiline CM44x / Liquistation CSFxx / Liquiline System CA80xx - Data transmission via Modbus (Special Documentation) (SD01189C/07/EN/07.22), revision 2022-07, published free by Endress+Hauser at https://bdih-download.endress.com/files/DLA/005056A500261EECB9FE4A363A4B4FD5/SD01189CEN_07.22.pdf.