- HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU)
Sensor (H2O2 / humidity probe) · Modbus register map
Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU) Modbus register map
Updated 19 November 2026
The HPP272 is a Vaisala PEROXCAP probe for vaporized hydrogen peroxide bio-decontamination that also measures humidity and temperature. It is a Modbus RTU slave on RS-485. This map covers H2O2, humidity and temperature readings, status codes, sensor vitality, compensation, purge, low-H2O2 threshold and analog-output settings.
This page lists 68 Modbus registers for the Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU) — 31 writable — with the table, 0-based wire address, data type, word order, scale, units and access of each, re-typed from Vaisala’s own free documentation (linked below) and checked against it.
- Manufacturer
- Vaisala
- Model
- HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU)
- Equipment type
- Sensor (H2O2 / humidity probe)
- Registers
- 68 (31 writable)
- Transport
- Modbus RTU
- Default unit ID
- 240
- Default serial
- 19200 baud, 8N2
- Documented addressing
- 1-based (subtract 1 for the wire address)
- 32-bit word order
- CDAB low word first (word-swapped)
- Also known as
- HPP272, PEROXCAP HPP272
- Source document
- HPP272 User Guide · M211972EN-G · rev. 2023-04
Integration notes: Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU)
Read holding registers 1-28 in one block for all H2O2, humidity and temperature values (CDAB floats). Watch 513 and the status code pair 514/516.
Minimum registers to map
Addresses are 0-based wire addresses. Enough to monitor and run the unit; map more as the job needs.
| Register | Name | Type | Why | Access |
|---|---|---|---|---|
| HR 0 | H2O2 concentration by volume | FLOAT32 CDAB | H2O2 ppm | Read |
| HR 6 | Relative humidity | FLOAT32 CDAB | Relative humidity | Read |
| HR 22 | Temperature from temperature probe | FLOAT32 CDAB | Probe temperature | Read |
| HR 512 | Device status | BITMAP16 | Device status | Read |
| HR 546 | Sensor vitality | FLOAT32 CDAB | Sensor vitality | Read |
Enable Modbus control & writes
Digital (RS-485 Modbus) or analog output is chosen at power-up by pin 5 of the M12 connector: leave pin 5 floating for Modbus, tie it to ground for analog. Power-cycle after changing it.
Byte order & addressing
- Floats and 32-bit integers are low word first (CDAB); pick "word-swapped" in masters that default to high word first.
- Read or write each 32-bit value in one transaction.
Network setup
- Defaults: address 240, 19200 baud, no parity, 8 data bits, 2 stop bits.
- Change address, bit rate, parity/stop bits and response delay with registers 1537-1540; writing 1 to 1541 restarts the probe.
Watch out for
- Use the test registers (7937 reads -12345, 7938 reads -123.45) to confirm float format and addressing before trusting readings.
- After starting a purge by writing 1 to 1283, only read that register until the purge has finished.
Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU) Modbus registers
Address = 0-based wire address (the documented number is shown under it when it differs) · Order: ABCD = high word first, CDAB = low word first · Scale: engineering value = raw × scale.
| Table | Address | Name | Description | Type | Order | Scale | Units | Access | Values / bits |
|---|---|---|---|---|---|---|---|---|---|
| HR | 0doc 1 | H2O2 concentration by volume | FLOAT32 | CDAB | ppm | Read | |||
| HR | 2doc 3 | H2O+H2O2 relative saturation | FLOAT32 | CDAB | %RS | Read | |||
| HR | 4doc 5 | Temperature used for calculation | Probe T, or the setpoint when T compensation is on. | FLOAT32 | CDAB | °C | Read | ||
| HR | 6doc 7 | Relative humidity | FLOAT32 | CDAB | %RH | Read | |||
| HR | 8doc 9 | Absolute hydrogen peroxide | FLOAT32 | CDAB | mg/m3 | Read | |||
| HR | 10doc 11 | H2O+H2O2 dew point | FLOAT32 | CDAB | °C | Read | |||
| HR | 14doc 15 | Water concentration by volume | FLOAT32 | CDAB | ppm | Read | |||
| HR | 16doc 17 | Water vapor pressure | FLOAT32 | CDAB | hPa | Read | |||
| HR | 18doc 19 | Absolute humidity (water) | FLOAT32 | CDAB | g/m3 | Read | |||
| HR | 20doc 21 | Water vapor saturation pressure | FLOAT32 | CDAB | hPa | Read | |||
| HR | 22doc 23 | Temperature from temperature probe | FLOAT32 | CDAB | °C | Read | |||
| HR | 24doc 25 | H2O2 vapor pressure | FLOAT32 | CDAB | hPa | Read | |||
| HR | 26doc 27 | Saturation pressure (H2O+H2O2) | FLOAT32 | CDAB | hPa | Read | |||
| HR | 256doc 257 | H2O2 concentration (integer) | INT16 | ppm | Read | ||||
| HR | 257doc 258 | Relative saturation (integer) | INT16 | × 0.01 | %RS | Read | |||
| HR | 258doc 259 | Temperature used (integer) | INT16 | × 0.01 | °C | Read | |||
| HR | 259doc 260 | Relative humidity (integer) | INT16 | × 0.01 | %RH | Read | |||
| HR | 260doc 261 | Absolute H2O2 (integer) | INT16 | mg/m3 | Read | ||||
| HR | 261doc 262 | H2O+H2O2 dew point (integer) | INT16 | × 0.01 | °C | Read | |||
| HR | 263doc 264 | Water concentration (integer) | INT16 | ppm | Read | ||||
| HR | 264doc 265 | Water vapor pressure (integer) | INT16 | hPa | Read | ||||
| HR | 265doc 266 | Absolute humidity (integer) | INT16 | g/m3 | Read | ||||
| HR | 266doc 267 | Water vapor saturation pressure (integer) | INT16 | hPa | Read | ||||
| HR | 267doc 268 | Probe temperature (integer) | INT16 | × 0.01 | °C | Read | |||
| HR | 268doc 269 | H2O2 vapor pressure (integer) | INT16 | hPa | Read | ||||
| HR | 269doc 270 | Saturation pressure H2O+H2O2 (integer) | INT16 | hPa | Read | ||||
| HR | 512doc 513 | Device status | 0 = OK. | BITMAP16 | Read | bit 0 = Critical error, maintenance needed; bit 1 = Error, may recover automatically; bit 2 = Warning; bit 3 = Notification | |||
| HR | 513doc 514 | Status code low | Sum of active codes. | UINT32 | CDAB | Read | bit 0 = Firmware checksum mismatch; bit 1 = Device settings corrupted; bit 3 = Supply voltage too high; bit 4 = Internal voltage too low; bit 5 = Internal voltage too high; bit 6 = Humidity sensor failure (open circuit); bit 10 = Humidity sensor failure (open circuit); bit 14 = Temperature sensor failure (open circuit); bit 15 = Temperature sensor failure (short circuit); bit 18 = Temperature sensor failure (open circuit); bit 19 = Temperature sensor failure (short circuit); bit 22 = Temperature sensor failure (open circuit); bit 23 = Temperature sensor failure (short circuit) | ||
| HR | 515doc 516 | Status code high | Sum of active codes. | UINT32 | CDAB | Read | bit 8 = A/D converter failure (external); bit 9 = Non-volatile memory failure; bit 12 = Humidity sensor failure (short circuit); bit 13 = Purge in progress; bit 14 = A/D converter failure (internal); bit 15 = Calibration about to expire; bit 16 = Calibration expired | ||
| HR | 528doc 529 | H2O2 concentration status | 0 = OK. | BITMAP16 | Read | bit 1 = Not reliable; bit 5 = Locked during purge; bit 7 = Sensor failure; bit 8 = Not ready | |||
| HR | 529doc 530 | Relative saturation status | BITMAP16 | Read | bit 1 = Not reliable; bit 5 = Locked during purge; bit 7 = Sensor failure; bit 8 = Not ready | ||||
| HR | 530doc 531 | Temperature status | BITMAP16 | Read | bit 1 = Not reliable; bit 5 = Locked during purge; bit 7 = Sensor failure; bit 8 = Not ready | ||||
| HR | 531doc 532 | Relative humidity status | BITMAP16 | Read | bit 1 = Not reliable; bit 5 = Locked during purge; bit 7 = Sensor failure; bit 8 = Not ready | ||||
| HR | 546doc 547 | Sensor vitality | 1-100 %; 100 = new sensor. In demanding use, Vaisala advises service at 40 %. | FLOAT32 | CDAB | % | Read | ||
| HR | 768doc 769 | Volatile pressure compensation | 813.25-1213.25; cleared at reset. | FLOAT32 | CDAB | hPa | Read/write | ||
| HR | 770doc 771 | Volatile temperature setpoint | -40...+80; cleared at reset. | FLOAT32 | CDAB | °C | Read/write | ||
| HR | 772doc 773 | Power-up pressure compensation | Default 1013.25. | FLOAT32 | CDAB | hPa | Read/write | ||
| HR | 774doc 775 | Power-up temperature setpoint | FLOAT32 | CDAB | °C | Read/write | |||
| HR | 776doc 777 | Purge interval | 60-10080, default 1440. | INT16 | min | Read/write | |||
| HR | 778doc 779 | Measurement filtering factor | 0.01-1; 1 = off. | FLOAT32 | CDAB | Read/write | |||
| HR | 782doc 783 | Low H2O2 threshold on/off | UINT16 | Read/write | 0 = Off; 1 = On | ||||
| HR | 784doc 785 | Low H2O2 threshold activation level | Default 3. | FLOAT32 | CDAB | ppm | Read/write | ||
| HR | 786doc 787 | Low H2O2 threshold activation delay | Default 1800. | INT32 | CDAB | s | Read/write | ||
| HR | 788doc 789 | Low H2O2 threshold deactivation level | Default 10. | FLOAT32 | CDAB | ppm | Read/write | ||
| HR | 790doc 791 | Low H2O2 threshold deactivation delay | Default 10. | INT32 | CDAB | s | Read/write | ||
| HR | 1282doc 1283 | Purge status / manual start | Write 1 to start; reads 1-100 progress, 0 done, -1 failed. | INT16 | Read/write | ||||
| HR | 1284doc 1285 | Pressure compensation on/off | UINT16 | Read/write | 0 = Off; 1 = On | ||||
| HR | 1285doc 1286 | Temperature compensation mode | UINT16 | Read/write | 0 = Use probe temperature; 1 = Use setpoint temperature | ||||
| HR | 1286doc 1287 | Manual purge trigger in analog mode | UINT16 | Read/write | 0 = Off; 1 = On | ||||
| HR | 1287doc 1288 | Allow interval purge during H2O2 exposure | UINT16 | Read/write | 0 = Off; 1 = On | ||||
| HR | 1536doc 1537 | Modbus address | 1-255, default 240. | INT16 | Read/write | ||||
| HR | 1537doc 1538 | Bit rate | UINT16 | Read/write | 5 = 9600; 6 = 19200; 7 = 38400; 8 = 57600; 9 = 115200 | ||||
| HR | 1538doc 1539 | Parity, data, stop bits | Default 1 (None, 8, 2). | UINT16 | Read/write | 0 = None, 8, 1; 1 = None, 8, 2; 2 = Even, 8, 1; 3 = Even, 8, 2; 4 = Odd, 8, 1; 5 = Odd, 8, 2 | |||
| HR | 1539doc 1540 | Response delay | Extra delay 0-1000 ms. | INT16 | ms | Read/write | |||
| HR | 1540doc 1541 | Restart device | Write 1 to restart. | INT16 | Write | ||||
| HR | 1792doc 1793 | AOUT1 output mode | UINT16 | Read/write | 0 = Off; 2 = 4-20 mA | ||||
| HR | 1793doc 1794 | AOUT1 output parameter | Register address of the parameter, 0x0000-0x001A. | UINT16 | Read/write | ||||
| HR | 1794doc 1795 | AOUT1 scale low end | FLOAT32 | CDAB | Read/write | ||||
| HR | 1796doc 1797 | AOUT1 scale high end | FLOAT32 | CDAB | Read/write | ||||
| HR | 1798doc 1799 | AOUT1 error output | 0-25 mA. | FLOAT32 | CDAB | mA | Read/write | ||
| HR | 2048doc 2049 | AOUT2 output mode | UINT16 | Read/write | 0 = Off; 2 = 4-20 mA | ||||
| HR | 2049doc 2050 | AOUT2 output parameter | Register address of the parameter, 0x0000-0x001A. | UINT16 | Read/write | ||||
| HR | 2050doc 2051 | AOUT2 scale low end | FLOAT32 | CDAB | Read/write | ||||
| HR | 2052doc 2053 | AOUT2 scale high end | FLOAT32 | CDAB | Read/write | ||||
| HR | 2054doc 2055 | AOUT2 error output | 0-25 mA. | FLOAT32 | CDAB | mA | Read/write | ||
| HR | 7936doc 7937 | Signed integer test | Always reads -12345. | INT16 | Read | ||||
| HR | 7937doc 7938 | Floating point test | Always reads -123.45. | FLOAT32 | CDAB | Read | |||
| HR | 7939doc 7940 | Text string test | Reads the text "-123.45". | STRING[8] | Read |
Notes on this map
- All registers are holding registers: read with function 03, write with function 16.
- 32-bit floats and 32-bit integers put the least significant word at the listed register (CDAB). The guide states this and its worked read example confirms it.
- Unavailable data reads as quiet NaN (float) or 0x8000 (16-bit integer) rather than an exception.
- Left out: the restore-factory-settings register, because the guide prints its number (7937) and hex address (1E00) inconsistently and 7937 is also the signed-integer test register.
- Measurement status registers 533-542 use the same coding as 529-532 and are left out to keep the map short.
- Clipping and error-limit registers for the analog outputs (1801-1807, 2057-2063) and the service agreement block (2817-2846) are left out.
- Status code registers 514 and 516 are 32-bit values, so they follow the low-word-first rule for 32-bit integers.
Check it against the live device. A register map tells you what the controller can expose; the unit in front of you may differ by firmware or options. Read a few registers and compare them with the unit’s own display; for a 32-bit value the float and byte-order decoder shows all four word orders at once, so a wrong word order or scale is obvious. New to this? Read what a register map is and why addresses are off by one.
Questions about the Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU)
What Modbus registers does the Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU) have?
This map lists 68 registers: 68 holding registers. 31 of them can be written. Registers vary with firmware, options and configuration, so confirm against the device.
Are the Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU) Modbus addresses 0-based or 1-based?
The manufacturer's document numbers registers 1-based (subtract 1 for the wire address). Every address on this page is the 0-based wire address you put in the request; the documented number is shown under it where it differs. The guide numbers registers in decimal starting at 1 and also prints the 0-based hex address used in the request; register number 1 is address 0x0000 on the wire.
What byte order does the Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU) use for 32-bit values?
Multi-register values in this map use CDAB (low word first (word-swapped)). Set your client to this order for every 32/64-bit value in the table.
What are the minimum Modbus registers to integrate the Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU)?
Start with HR 0 H2O2 concentration by volume (H2O2 ppm); HR 6 Relative humidity (Relative humidity); HR 22 Temperature from temperature probe (Probe temperature); HR 512 Device status (Device status); HR 546 Sensor vitality (Sensor vitality).
How do I enable Modbus control of the Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU)?
Digital (RS-485 Modbus) or analog output is chosen at power-up by pin 5 of the M12 connector: leave pin 5 floating for Modbus, tie it to ground for analog. Power-cycle after changing it.
Where is the official Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU) Modbus documentation?
The source is HPP272 User Guide (M211972EN-G), revision 2023-04, published free by Vaisala at https://docs.vaisala.com/r/M211972EN-G/en-US.
More Vaisala Modbus register maps
- Sensor (dew point probe)
Vaisala DMP Series dew point probes (DMP5, DMP6, DMP7, DMP8) (Modbus RTU)
Vaisala DMP5, DMP6, DMP7 and DMP8 dew point and temperature probes, Modbus RTU interface (user guide M212357EN-F).
Also known as: DMP5, DMP6, DMP7, DMP8, DRYCAP DMP series - Sensor (dew point transmitter)
Vaisala DMT143 miniature dew point transmitter (Modbus RTU)
DMT143 dew point transmitters ordered with the Modbus RTU option (user guide M211435EN-L).
Also known as: DMT143, DRYCAP DMT143 - Sensor (dew point transmitter)
Vaisala DMT153 dew point transmitter (Modbus RTU)
DMT153 trace-moisture dew point transmitter, Modbus RTU interface (user guide M213149EN-B).
Also known as: DMT153, DRYCAP DMT153 - Sensor (CO2 duct transmitter)
Vaisala GMD110 duct CO2 transmitter (Modbus RTU)
GMD110 duct-mount CO2 transmitter, Modbus RTU interface (user guide M212885EN-B).
Also known as: GMD110, CARBOCAP GMD110 - Sensor (CO2 probe)
Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)
GMP251 probe in Modbus RTU mode, user guide M211799EN-N.
Also known as: GMP251, CARBOCAP GMP251, Vaisala GMP251 - Sensor (CO2 probe)
Vaisala GMP252 CARBOCAP carbon dioxide probe (Modbus RTU)
GMP252 probe in Modbus RTU mode, user guide M211897EN-G.
Also known as: GMP252, CARBOCAP GMP252, Vaisala GMP252 - Sensor (CO2 wall transmitter)
Vaisala GMW87 CO2 and temperature wall transmitter (Modbus RTU)
GMW87, the RS-485 Modbus RTU model of the GMW80 series (user guide M211748EN-H).
Also known as: GMW87, GMW80 series Modbus - Sensor (humidity and temperature transmitter)
Vaisala HMDW110 series humidity and temperature transmitters with TMI110 (Modbus RTU)
HMDW110-series duct, wall and outdoor transmitters and TMI110 ordered with the Modbus RTU option (user guide M211726EN-L).
Also known as: HMD110, HMD112, HMW110, HMW112, HMW112D, HMS110, HMS112, TMW110, TMW112, TMW112D, TMI110 - Sensor (humidity and temperature module)
Vaisala HMM170 humidity module (Modbus RTU)
HMM170 OEM humidity and temperature module with the Modbus RTU interface (user guide M212259EN-D).
Also known as: HMM170, HUMICAP HMM170 - Sensor (humidity and temperature probe)
Vaisala HMP Series (HMP1-HMP9) with MMP8 and TMP1 probes (Modbus RTU)
Vaisala HMP1/3/4/5/7/8/9, TMP1 and MMP8 probes with the Modbus RTU interface (user guide M212022EN-M).
Also known as: HMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9, TMP1, MMP8 - Sensor (humidity and temperature probe)
Vaisala HMP60 / HMP110 series humidity and temperature probes (Modbus RTU)
HMP60/HMP110-series probes ordered with the Modbus RTU option (user guide M211060EN-N).
Also known as: HMP110, HMP113, HMP115, TMP115, HMP60, HMP63, HUMICAP HMP110, HMP110T, HMP63T, HMP115T - Sensor (humidity and temperature transmitter)
Vaisala HMW87 / HMW87D wall humidity and temperature transmitters (Modbus RTU)
HMW87 and HMW87D wall transmitters of the HMDW80 series (the Modbus models in user guide M211510EN-J).
Also known as: HMW87, HMW87D, HMDW80 Modbus - Sensor (H2O2 probe)
Vaisala HPP271 hydrogen peroxide probe (Modbus RTU)
HPP271 vaporized hydrogen peroxide probe, Modbus RTU interface (user guide M211888EN-E).
Also known as: HPP271, PEROXCAP HPP271 - Sensor transmitter (host for Vaisala probes)
Vaisala Indigo510 / Indigo520 transmitters (Modbus TCP)
Indigo510 and Indigo520 transmitter registers (unit ID 240) over Modbus TCP (user guide M212287EN-AJ).
Also known as: Indigo510, Indigo520, Indigo500 series - Sensor (transformer oil monitor)
Vaisala MHT410 moisture, hydrogen and temperature transmitter for transformer oil (Modbus RTU)
MHT410 transformer oil monitor, Modbus RTU on the RS-485 screw terminals (user guide M211737EN-J).
Also known as: MHT410 - Sensor (moisture in oil transmitter)
Vaisala MMT143 moisture in oil transmitter (Modbus RTU)
MMT143 moisture-in-oil transmitter ordered with the Modbus RTU option (user guide M213308EN-B).
Also known as: MMT143 - Sensor (HVAC humidity, temperature and CO2 transmitter)
Vaisala Origo10 transmitters for HVAC (Origo10A, Modbus RTU)
Origo10A duct and wall transmitters ordered with the Modbus RTU output (user guide M213293EN-B). Origo10L has no Modbus.
Also known as: Origo10, Origo10A, Origo 10, Vaisala Origo10 duct transmitter, Vaisala Origo10 wall transmitter - Sensor (precipitation)
Vaisala RM60 PRECICAP radar precipitation sensor (Modbus RTU)
RM60 sensors ordered with communication option 1 (RS-485 Modbus RTU) or switched to Modbus, per user guide M213218EN-B.
Also known as: RM60, PRECICAP RM60, Vaisala RM60 - Sensor (ultrasonic wind)
Vaisala WM80 ultrasonic wind sensor (Modbus RTU)
WM80 sensors with RS-485 in Modbus mode (e.g. ordering option C), per user guide M212995EN-F.
Also known as: WM80, Vaisala WM80, Ultrasonic Wind Sensor WM80 - Sensor (weather transmitter)
Vaisala WXT530 series weather transmitters (Modbus RTU)
WXT531-WXT536 weather transmitters with firmware 3.85 or later in Modbus RTU mode (user guide M211840EN-G).
Also known as: WXT531, WXT532, WXT533, WXT534, WXT535, WXT536 - Sensor (humidity and temperature probe)
Vaisala XMP10 humidity and temperature probe (Modbus RTU)
XMP10 probes ordered with (or switched to) Modbus RTU output, user guide M213279EN-A.
Also known as: XMP10, HUMICAP XMP10, Vaisala XMP10 - Sensor (CO2/humidity wall transmitter)
Vaisala XMW85 CO2, humidity and temperature wall transmitter (Modbus RTU)
XMW85 wall transmitter, Modbus RTU interface (user guide M213257EN-B).
Also known as: XMW85
Other valves, actuators & sensors register maps
- Water treatment controller (Sensor / controller)
Walchem WebMaster WIND series Modbus TCP/IP option (s825)
Walchem WebMaster WIND-series controllers with the Modbus TCP/IP option, software s825v008 and higher, per manual 180413 Rev A1.
Also known as: WIND, WebMaster WIND, s825 - Actuator / valve
Armstrong International Brain DRV25 / DRV40 / DRV80 digital recirculation valve (Modbus RTU)
Armstrong International Brain digital recirculation (thermostatic mixing) valves DRV25, DRV40 and DRV80 on their RS-485 Modbus RTU port, per IOM-776 Brain Modbus Networking (2024).
Also known as: DRV25, DRV40, DRV80, Brain, IOM-776 - Sensor (humidity / temperature probe)
E+E Elektronik EE260 humidity and temperature probe for meteorology (RS-485 Modbus RTU)
EE260 meteorological RH/T probe with RS-485 Modbus RTU, per Quick Guide QG_EE260 v1.3 (10-2025).
Also known as: EE260, E+E EE260, EE 260 - Sensor (disinfectant / chlorine)
JUMO tecLine digital disinfection sensors (Cl2, TC, ClO2/O3, H2O2/PAA, Br, Cl2 OM) (Modbus RTU)
JUMO tecLine 202630, 202631, 202634, 202636, 202637 and 202681 sensors with RS-485 Modbus RTU (interface description V2.00, 2022-03-24).
Also known as: 202630, 202631, 202634, 202636, 202637, 202681, tecLine Cl2, tecLine TC, tecLine ClO2, tecLine O3, tecLine H2O2, tecLine PAA, tecLine Br, tecLine Cl2 OM - Sensor (weighing indicator)
Radwag PUE C32 / PUE HX7 / PUE HX5.EX weighing indicators
Radwag PUE C32, PUE HX7 and PUE HX5.EX weighing indicators per the May 2024 Modbus manual; platform 2 registers exist on PUE HX7 only.
Also known as: PUE C32, PUE HX7, PUE HX5.EX - Sensor (chart recorder)
Shinko Technos HR-700 hybrid recorder (Modbus RTU, firmware 4.00+)
Shinko HR-700 hybrid recorders (multipoint and pen types), body version 4.00 or later, Modbus map version 01; manual No. HR7CE4.
Also known as: HR-700
All valves, actuators & sensors Modbus register maps →
Source: HPP272 User Guide · M211972EN-G · rev. 2023-04. Register numbers, types, scaling and units are reproduced as published; descriptions are our own.