Sensor (CO2 probe) · Modbus register map

Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) Modbus register map

Updated 19 November 2026

Short answer

The GMP251 is a Vaisala CARBOCAP probe for percent-level CO2 (incubators, life science). Over RS-485 it is a Modbus RTU slave. This map covers the CO2 and temperature readings, the compensation registers, serial settings and status/error registers.

This page lists 25 Modbus registers for the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) — 17 writable — with the table, 0-based wire address, data type, word order, scale, units and access of each, re-typed from Vaisala’s own free documentation (linked below) and checked against it.

Manufacturer
Vaisala
Model
GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)
Equipment type
Sensor (CO2 probe)
Registers
25 (17 writable)
Transport
Modbus RTU
Default unit ID
240
Default serial
19200 baud, 8N2
Documented addressing
1-based (subtract 1 for the wire address)
32-bit word order
CDAB low word first (word-swapped)
Firmware
Error code register needs probe firmware 1.4.3 or later
Also known as
GMP251, CARBOCAP GMP251, Vaisala GMP251
Source document
GMP251 and GMP80P User Guide · M211799EN-N · rev. 2026-11
Download CSV all columns JSON Manufacturer PDF ↗

Integration notes: Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)

Integration in brief

Read holding registers 1-6 in one request (function 03) for CO2 and temperature as CDAB floats. Write compensation values to the volatile registers 521-528 if they change often.

Minimum registers to map

Addresses are 0-based wire addresses. Enough to monitor and run the unit; map more as the job needs.

RegisterNameTypeWhyAccess
HR 0Measured CO2FLOAT32 CDABCO2 concentrationRead
HR 4Measured TFLOAT32 CDABProbe temperatureRead
HR 2048Device statusUINT16Device statusRead
HR 2051Error codeUINT32 CDABActive error bitsRead
HR 520Volatile pressure compensationFLOAT32 CDABLive pressure compensationRead/write

Enable Modbus control & writes

Modbus RTU over RS-485 is one of the probe's output modes, chosen at order or later by configuration. From Vaisala Industrial Protocol, switch with the serial command smode modbus. If pin 5 is tied to GND (pin 3) the probe is locked in analog mode and Modbus is not available.

Byte order & addressing

Network setup

Watch out for

Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) Modbus registers

Address = 0-based wire address (the documented number is shown under it when it differs) · Order: ABCD = high word first, CDAB = low word first · Scale: engineering value = raw × scale.

25 of 25 shown
Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) Modbus register map
TableAddressNameDescriptionTypeOrderScaleUnitsAccessValues / bits
HR0doc 1Measured CO2FLOAT32CDABppmRead
HR2doc 3Compensation TTemperature value the probe is using for compensation.FLOAT32CDAB°CRead
HR4doc 5Measured TTemperature measured by the probe.FLOAT32CDAB°CRead
HR256doc 257Measured CO2 (integer)Integer copy, valid up to 32000 ppm.INT16ppmRead
HR257doc 258Measured CO2 (integer, x10)Integer copy scaled down by 10; multiply raw by 10.INT16× 10ppmRead
HR512doc 513Power-up pressure compensationStored in EEPROM; 500-1100 hPa, default 1013.25.FLOAT32CDABhPaRead/write
HR514doc 515Power-up temperature compensationStored in EEPROM; -40...+60, default 25.FLOAT32CDAB°CRead/write
HR516doc 517Power-up humidity compensationStored in EEPROM; 0-100, default 0.FLOAT32CDAB%RHRead/write
HR518doc 519Power-up oxygen compensationStored in EEPROM; 0-100, default 0.FLOAT32CDAB%O2Read/write
HR520doc 521Volatile pressure compensationCleared at reset; use for frequent updates.FLOAT32CDABhPaRead/write
HR522doc 523Volatile temperature compensationCleared at reset; use for frequent updates.FLOAT32CDAB°CRead/write
HR524doc 525Volatile humidity compensationCleared at reset; use for frequent updates.FLOAT32CDAB%RHRead/write
HR526doc 527Volatile oxygen compensationCleared at reset; use for frequent updates.FLOAT32CDAB%O2Read/write
HR768doc 769Modbus address1-247, default 240.UINT16Read/write
HR769doc 770Serial speedUINT16Read/write0 = 4800; 1 = 9600; 2 = 19200; 3 = 38400; 4 = 57600; 5 = 115200
HR770doc 771Serial parityUINT16Read/write0 = None; 1 = Even; 2 = Odd
HR771doc 772Serial stop bits1 or 2, default 2.UINT16Read/write
HR772doc 773Pressure compensation modeUINT16Read/write0 = Off; 1 = On
HR773doc 774Temperature compensation modeUINT16Read/write0 = Off; 1 = Given; 2 = Internal
HR774doc 775Humidity compensation modeUINT16Read/write0 = Off; 1 = On
HR775doc 776Oxygen compensation modeUINT16Read/write0 = Off; 1 = On
HR776doc 777CO2 filtering factor0-100; 100 = no filtering, 50 = factor 0.5.UINT16Read/write
HR2048doc 2049Device statusUINT16Read0 = OK; 1 = Critical error; 2 = Error; 4 = Warning
HR2049doc 2050CO2 statusUINT16Read0 = OK; 2 = CO2 reading not reliable (start-up); 256 = Measurement not ready (start-up)
HR2051doc 2052Error codeSum of active error bits; probe firmware 1.4.3 and later.UINT32CDABReadbit 0 = Program memory CRC critical error; bit 1 = Parameter memory CRC critical error; bit 4 = Low supply voltage; bit 5 = Internal 30 V error; bit 6 = Low RX signal; bit 7 = Internal 8 V error; bit 8 = RX signal cut; bit 12 = Out of measurement range; bit 13 = Sensor heater error; bit 14 = IR temperature error; bit 15 = FPI slope error; bit 16 = Internal 2.5 V error; bit 17 = Internal 1.7 V error; bit 18 = Low IR current

Notes on this map

Check it against the live device. A register map tells you what the controller can expose; the unit in front of you may differ by firmware or options. Read a few registers and compare them with the unit’s own display; for a 32-bit value the float and byte-order decoder shows all four word orders at once, so a wrong word order or scale is obvious. New to this? Read what a register map is and why addresses are off by one.

Questions about the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)

What Modbus registers does the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) have?

This map lists 25 registers: 25 holding registers. 17 of them can be written. Registers vary with firmware, options and configuration, so confirm against the device.

Are the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) Modbus addresses 0-based or 1-based?

The manufacturer's document numbers registers 1-based (subtract 1 for the wire address). Every address on this page is the 0-based wire address you put in the request; the documented number is shown under it where it differs. The guide numbers registers in decimal starting at 1 and also prints the 0-based hex address used in the request; register number 1 is address 0x0000 on the wire.

What byte order does the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) use for 32-bit values?

Multi-register values in this map use CDAB (low word first (word-swapped)). Set your client to this order for every 32/64-bit value in the table.

What are the minimum Modbus registers to integrate the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)?

Start with HR 0 Measured CO2 (CO2 concentration); HR 4 Measured T (Probe temperature); HR 2048 Device status (Device status); HR 2051 Error code (Active error bits); HR 520 Volatile pressure compensation (Live pressure compensation).

How do I enable Modbus control of the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)?

Modbus RTU over RS-485 is one of the probe's output modes, chosen at order or later by configuration. From Vaisala Industrial Protocol, switch with the serial command smode modbus. If pin 5 is tied to GND (pin 3) the probe is locked in analog mode and Modbus is not available.

Where is the official Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) Modbus documentation?

The source is GMP251 and GMP80P User Guide (M211799EN-N), revision 2026-11, published free by Vaisala at https://docs.vaisala.com/r/M211799EN-N/en-US.

More Vaisala Modbus register maps

Other valves, actuators & sensors register maps

All valves, actuators & sensors Modbus register maps →

Source: GMP251 and GMP80P User Guide · M211799EN-N · rev. 2026-11. Register numbers, types, scaling and units are reproduced as published; descriptions are our own.