- GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)
Sensor (CO2 probe) · Modbus register map
Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) Modbus register map
Updated 19 November 2026
The GMP251 is a Vaisala CARBOCAP probe for percent-level CO2 (incubators, life science). Over RS-485 it is a Modbus RTU slave. This map covers the CO2 and temperature readings, the compensation registers, serial settings and status/error registers.
This page lists 25 Modbus registers for the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) — 17 writable — with the table, 0-based wire address, data type, word order, scale, units and access of each, re-typed from Vaisala’s own free documentation (linked below) and checked against it.
- Manufacturer
- Vaisala
- Model
- GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)
- Equipment type
- Sensor (CO2 probe)
- Registers
- 25 (17 writable)
- Transport
- Modbus RTU
- Default unit ID
- 240
- Default serial
- 19200 baud, 8N2
- Documented addressing
- 1-based (subtract 1 for the wire address)
- 32-bit word order
- CDAB low word first (word-swapped)
- Firmware
- Error code register needs probe firmware 1.4.3 or later
- Also known as
- GMP251, CARBOCAP GMP251, Vaisala GMP251
- Source document
- GMP251 and GMP80P User Guide · M211799EN-N · rev. 2026-11
Integration notes: Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)
Read holding registers 1-6 in one request (function 03) for CO2 and temperature as CDAB floats. Write compensation values to the volatile registers 521-528 if they change often.
Minimum registers to map
Addresses are 0-based wire addresses. Enough to monitor and run the unit; map more as the job needs.
| Register | Name | Type | Why | Access |
|---|---|---|---|---|
| HR 0 | Measured CO2 | FLOAT32 CDAB | CO2 concentration | Read |
| HR 4 | Measured T | FLOAT32 CDAB | Probe temperature | Read |
| HR 2048 | Device status | UINT16 | Device status | Read |
| HR 2051 | Error code | UINT32 CDAB | Active error bits | Read |
| HR 520 | Volatile pressure compensation | FLOAT32 CDAB | Live pressure compensation | Read/write |
Enable Modbus control & writes
Modbus RTU over RS-485 is one of the probe's output modes, chosen at order or later by configuration. From Vaisala Industrial Protocol, switch with the serial command smode modbus. If pin 5 is tied to GND (pin 3) the probe is locked in analog mode and Modbus is not available.
Byte order & addressing
- Float word order is low word first (CDAB); use "word-swapped" in masters that expect high word first.
- Read each 32-bit value in a single transaction.
Network setup
- Defaults: address 240, 19200 baud, no parity, 8 data bits, 2 stop bits.
- Change address and serial settings with registers 769-772, or in Vaisala Insight PC software.
Watch out for
- Register 258 must be multiplied by 10.
- Relative humidity compensation only works with temperature and pressure compensation also enabled.
Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) Modbus registers
Address = 0-based wire address (the documented number is shown under it when it differs) · Order: ABCD = high word first, CDAB = low word first · Scale: engineering value = raw × scale.
| Table | Address | Name | Description | Type | Order | Scale | Units | Access | Values / bits |
|---|---|---|---|---|---|---|---|---|---|
| HR | 0doc 1 | Measured CO2 | FLOAT32 | CDAB | ppm | Read | |||
| HR | 2doc 3 | Compensation T | Temperature value the probe is using for compensation. | FLOAT32 | CDAB | °C | Read | ||
| HR | 4doc 5 | Measured T | Temperature measured by the probe. | FLOAT32 | CDAB | °C | Read | ||
| HR | 256doc 257 | Measured CO2 (integer) | Integer copy, valid up to 32000 ppm. | INT16 | ppm | Read | |||
| HR | 257doc 258 | Measured CO2 (integer, x10) | Integer copy scaled down by 10; multiply raw by 10. | INT16 | × 10 | ppm | Read | ||
| HR | 512doc 513 | Power-up pressure compensation | Stored in EEPROM; 500-1100 hPa, default 1013.25. | FLOAT32 | CDAB | hPa | Read/write | ||
| HR | 514doc 515 | Power-up temperature compensation | Stored in EEPROM; -40...+60, default 25. | FLOAT32 | CDAB | °C | Read/write | ||
| HR | 516doc 517 | Power-up humidity compensation | Stored in EEPROM; 0-100, default 0. | FLOAT32 | CDAB | %RH | Read/write | ||
| HR | 518doc 519 | Power-up oxygen compensation | Stored in EEPROM; 0-100, default 0. | FLOAT32 | CDAB | %O2 | Read/write | ||
| HR | 520doc 521 | Volatile pressure compensation | Cleared at reset; use for frequent updates. | FLOAT32 | CDAB | hPa | Read/write | ||
| HR | 522doc 523 | Volatile temperature compensation | Cleared at reset; use for frequent updates. | FLOAT32 | CDAB | °C | Read/write | ||
| HR | 524doc 525 | Volatile humidity compensation | Cleared at reset; use for frequent updates. | FLOAT32 | CDAB | %RH | Read/write | ||
| HR | 526doc 527 | Volatile oxygen compensation | Cleared at reset; use for frequent updates. | FLOAT32 | CDAB | %O2 | Read/write | ||
| HR | 768doc 769 | Modbus address | 1-247, default 240. | UINT16 | Read/write | ||||
| HR | 769doc 770 | Serial speed | UINT16 | Read/write | 0 = 4800; 1 = 9600; 2 = 19200; 3 = 38400; 4 = 57600; 5 = 115200 | ||||
| HR | 770doc 771 | Serial parity | UINT16 | Read/write | 0 = None; 1 = Even; 2 = Odd | ||||
| HR | 771doc 772 | Serial stop bits | 1 or 2, default 2. | UINT16 | Read/write | ||||
| HR | 772doc 773 | Pressure compensation mode | UINT16 | Read/write | 0 = Off; 1 = On | ||||
| HR | 773doc 774 | Temperature compensation mode | UINT16 | Read/write | 0 = Off; 1 = Given; 2 = Internal | ||||
| HR | 774doc 775 | Humidity compensation mode | UINT16 | Read/write | 0 = Off; 1 = On | ||||
| HR | 775doc 776 | Oxygen compensation mode | UINT16 | Read/write | 0 = Off; 1 = On | ||||
| HR | 776doc 777 | CO2 filtering factor | 0-100; 100 = no filtering, 50 = factor 0.5. | UINT16 | Read/write | ||||
| HR | 2048doc 2049 | Device status | UINT16 | Read | 0 = OK; 1 = Critical error; 2 = Error; 4 = Warning | ||||
| HR | 2049doc 2050 | CO2 status | UINT16 | Read | 0 = OK; 2 = CO2 reading not reliable (start-up); 256 = Measurement not ready (start-up) | ||||
| HR | 2051doc 2052 | Error code | Sum of active error bits; probe firmware 1.4.3 and later. | UINT32 | CDAB | Read | bit 0 = Program memory CRC critical error; bit 1 = Parameter memory CRC critical error; bit 4 = Low supply voltage; bit 5 = Internal 30 V error; bit 6 = Low RX signal; bit 7 = Internal 8 V error; bit 8 = RX signal cut; bit 12 = Out of measurement range; bit 13 = Sensor heater error; bit 14 = IR temperature error; bit 15 = FPI slope error; bit 16 = Internal 2.5 V error; bit 17 = Internal 1.7 V error; bit 18 = Low IR current |
Notes on this map
- All values are holding registers read with function 03 and written with function 16.
- 32-bit floats and the 32-bit error code put the least significant word at the listed register (CDAB); the guide confirms this with a worked read frame (bytes D4 7A 43 E8 = 465.66 ppm).
- Missing data returns quiet NaN (float) or 0x8000 (integer) instead of an exception.
- Power-up compensation registers are stored in EEPROM (about 30000 writes); write the volatile registers for frequent updates.
- The same guide covers GMP80P, which is not included here; the register list in the Modbus chapter is written for the GMP251 probe.
- The guide lists the CO2 registers in ppm.
Check it against the live device. A register map tells you what the controller can expose; the unit in front of you may differ by firmware or options. Read a few registers and compare them with the unit’s own display; for a 32-bit value the float and byte-order decoder shows all four word orders at once, so a wrong word order or scale is obvious. New to this? Read what a register map is and why addresses are off by one.
Questions about the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)
What Modbus registers does the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) have?
This map lists 25 registers: 25 holding registers. 17 of them can be written. Registers vary with firmware, options and configuration, so confirm against the device.
Are the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) Modbus addresses 0-based or 1-based?
The manufacturer's document numbers registers 1-based (subtract 1 for the wire address). Every address on this page is the 0-based wire address you put in the request; the documented number is shown under it where it differs. The guide numbers registers in decimal starting at 1 and also prints the 0-based hex address used in the request; register number 1 is address 0x0000 on the wire.
What byte order does the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) use for 32-bit values?
Multi-register values in this map use CDAB (low word first (word-swapped)). Set your client to this order for every 32/64-bit value in the table.
What are the minimum Modbus registers to integrate the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)?
Start with HR 0 Measured CO2 (CO2 concentration); HR 4 Measured T (Probe temperature); HR 2048 Device status (Device status); HR 2051 Error code (Active error bits); HR 520 Volatile pressure compensation (Live pressure compensation).
How do I enable Modbus control of the Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU)?
Modbus RTU over RS-485 is one of the probe's output modes, chosen at order or later by configuration. From Vaisala Industrial Protocol, switch with the serial command smode modbus. If pin 5 is tied to GND (pin 3) the probe is locked in analog mode and Modbus is not available.
Where is the official Vaisala GMP251 CARBOCAP carbon dioxide probe (Modbus RTU) Modbus documentation?
The source is GMP251 and GMP80P User Guide (M211799EN-N), revision 2026-11, published free by Vaisala at https://docs.vaisala.com/r/M211799EN-N/en-US.
More Vaisala Modbus register maps
- Sensor (dew point probe)
Vaisala DMP Series dew point probes (DMP5, DMP6, DMP7, DMP8) (Modbus RTU)
Vaisala DMP5, DMP6, DMP7 and DMP8 dew point and temperature probes, Modbus RTU interface (user guide M212357EN-F).
Also known as: DMP5, DMP6, DMP7, DMP8, DRYCAP DMP series - Sensor (dew point transmitter)
Vaisala DMT143 miniature dew point transmitter (Modbus RTU)
DMT143 dew point transmitters ordered with the Modbus RTU option (user guide M211435EN-L).
Also known as: DMT143, DRYCAP DMT143 - Sensor (dew point transmitter)
Vaisala DMT153 dew point transmitter (Modbus RTU)
DMT153 trace-moisture dew point transmitter, Modbus RTU interface (user guide M213149EN-B).
Also known as: DMT153, DRYCAP DMT153 - Sensor (CO2 duct transmitter)
Vaisala GMD110 duct CO2 transmitter (Modbus RTU)
GMD110 duct-mount CO2 transmitter, Modbus RTU interface (user guide M212885EN-B).
Also known as: GMD110, CARBOCAP GMD110 - Sensor (CO2 probe)
Vaisala GMP252 CARBOCAP carbon dioxide probe (Modbus RTU)
GMP252 probe in Modbus RTU mode, user guide M211897EN-G.
Also known as: GMP252, CARBOCAP GMP252, Vaisala GMP252 - Sensor (CO2 wall transmitter)
Vaisala GMW87 CO2 and temperature wall transmitter (Modbus RTU)
GMW87, the RS-485 Modbus RTU model of the GMW80 series (user guide M211748EN-H).
Also known as: GMW87, GMW80 series Modbus - Sensor (humidity and temperature transmitter)
Vaisala HMDW110 series humidity and temperature transmitters with TMI110 (Modbus RTU)
HMDW110-series duct, wall and outdoor transmitters and TMI110 ordered with the Modbus RTU option (user guide M211726EN-L).
Also known as: HMD110, HMD112, HMW110, HMW112, HMW112D, HMS110, HMS112, TMW110, TMW112, TMW112D, TMI110 - Sensor (humidity and temperature module)
Vaisala HMM170 humidity module (Modbus RTU)
HMM170 OEM humidity and temperature module with the Modbus RTU interface (user guide M212259EN-D).
Also known as: HMM170, HUMICAP HMM170 - Sensor (humidity and temperature probe)
Vaisala HMP Series (HMP1-HMP9) with MMP8 and TMP1 probes (Modbus RTU)
Vaisala HMP1/3/4/5/7/8/9, TMP1 and MMP8 probes with the Modbus RTU interface (user guide M212022EN-M).
Also known as: HMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9, TMP1, MMP8 - Sensor (humidity and temperature probe)
Vaisala HMP60 / HMP110 series humidity and temperature probes (Modbus RTU)
HMP60/HMP110-series probes ordered with the Modbus RTU option (user guide M211060EN-N).
Also known as: HMP110, HMP113, HMP115, TMP115, HMP60, HMP63, HUMICAP HMP110, HMP110T, HMP63T, HMP115T - Sensor (humidity and temperature transmitter)
Vaisala HMW87 / HMW87D wall humidity and temperature transmitters (Modbus RTU)
HMW87 and HMW87D wall transmitters of the HMDW80 series (the Modbus models in user guide M211510EN-J).
Also known as: HMW87, HMW87D, HMDW80 Modbus - Sensor (H2O2 probe)
Vaisala HPP271 hydrogen peroxide probe (Modbus RTU)
HPP271 vaporized hydrogen peroxide probe, Modbus RTU interface (user guide M211888EN-E).
Also known as: HPP271, PEROXCAP HPP271 - Sensor (H2O2 / humidity probe)
Vaisala HPP272 hydrogen peroxide, humidity and temperature probe (Modbus RTU)
HPP272 vaporized hydrogen peroxide probe, Modbus RTU interface (user guide M211972EN-G).
Also known as: HPP272, PEROXCAP HPP272 - Sensor transmitter (host for Vaisala probes)
Vaisala Indigo510 / Indigo520 transmitters (Modbus TCP)
Indigo510 and Indigo520 transmitter registers (unit ID 240) over Modbus TCP (user guide M212287EN-AJ).
Also known as: Indigo510, Indigo520, Indigo500 series - Sensor (transformer oil monitor)
Vaisala MHT410 moisture, hydrogen and temperature transmitter for transformer oil (Modbus RTU)
MHT410 transformer oil monitor, Modbus RTU on the RS-485 screw terminals (user guide M211737EN-J).
Also known as: MHT410 - Sensor (moisture in oil transmitter)
Vaisala MMT143 moisture in oil transmitter (Modbus RTU)
MMT143 moisture-in-oil transmitter ordered with the Modbus RTU option (user guide M213308EN-B).
Also known as: MMT143 - Sensor (HVAC humidity, temperature and CO2 transmitter)
Vaisala Origo10 transmitters for HVAC (Origo10A, Modbus RTU)
Origo10A duct and wall transmitters ordered with the Modbus RTU output (user guide M213293EN-B). Origo10L has no Modbus.
Also known as: Origo10, Origo10A, Origo 10, Vaisala Origo10 duct transmitter, Vaisala Origo10 wall transmitter - Sensor (precipitation)
Vaisala RM60 PRECICAP radar precipitation sensor (Modbus RTU)
RM60 sensors ordered with communication option 1 (RS-485 Modbus RTU) or switched to Modbus, per user guide M213218EN-B.
Also known as: RM60, PRECICAP RM60, Vaisala RM60 - Sensor (ultrasonic wind)
Vaisala WM80 ultrasonic wind sensor (Modbus RTU)
WM80 sensors with RS-485 in Modbus mode (e.g. ordering option C), per user guide M212995EN-F.
Also known as: WM80, Vaisala WM80, Ultrasonic Wind Sensor WM80 - Sensor (weather transmitter)
Vaisala WXT530 series weather transmitters (Modbus RTU)
WXT531-WXT536 weather transmitters with firmware 3.85 or later in Modbus RTU mode (user guide M211840EN-G).
Also known as: WXT531, WXT532, WXT533, WXT534, WXT535, WXT536 - Sensor (humidity and temperature probe)
Vaisala XMP10 humidity and temperature probe (Modbus RTU)
XMP10 probes ordered with (or switched to) Modbus RTU output, user guide M213279EN-A.
Also known as: XMP10, HUMICAP XMP10, Vaisala XMP10 - Sensor (CO2/humidity wall transmitter)
Vaisala XMW85 CO2, humidity and temperature wall transmitter (Modbus RTU)
XMW85 wall transmitter, Modbus RTU interface (user guide M213257EN-B).
Also known as: XMW85
Other valves, actuators & sensors register maps
- Water treatment controller (Sensor / controller)
Walchem W900 water treatment controller (Modbus/TCP)
Walchem W900 controllers (software 3.01 and up) with the Modbus/TCP feature, per Modbus manual 180676.K. W600, Intuition-6 and Intuition-9 use the same type maps but different block start addresses.
Also known as: W900 series - Sensor (Ethernet analog output module)
Acromag BusWorks XT1531-000 4-channel current output + 4 digital I/O (Modbus TCP)
BusWorks XT1531-000 per its user's manual (file revision H, 18 Nov 2022).
Also known as: XT1531-000, XT1531, BusWorks XT1531 - Sensor (humidity / temperature transmitter)
Dwyer Series RHPX communicating humidity/temperature transmitter
Dwyer Series RHPX humidity and temperature transmitters with the BACnet MS/TP / Modbus RTU communication board, set to Modbus with DIP switch 9 (bulletin AQ-RHPX-C).
Also known as: RHPX, Dwyer RHPX BACnet/Modbus, 607xxx - Sensor (turbidity)
JUMO ecoLine NTU digital turbidity sensor (Modbus RTU)
JUMO ecoLine NTU type 202670 with RS-485 Modbus RTU (interface description V1.01).
Also known as: 202670, JUMO ecoLine NTU, ecoLine NTU type 202670 - Sensor (vibration / temperature)
Machine Saver TriVibe vibration and temperature sensor (Modbus RTU)
Machine Saver TriVibe triaxial vibration sensors on RS-485 Modbus RTU, per the 'Modbus Registers Map' page of the Machine Saver online library (revision #12).
Also known as: TriVibe, Machine Saver TriVibe - Sensor (conductivity controller)
Shinko Technos AER-102-ECH conductivity indicating controller (C5 option)
Shinko AER-102-ECH electrical conductivity controller with the C5 RS-485 option; manual No. AER11CECE5.
Also known as: AER-102-ECH
All valves, actuators & sensors Modbus register maps →
Source: GMP251 and GMP80P User Guide · M211799EN-N · rev. 2026-11. Register numbers, types, scaling and units are reproduced as published; descriptions are our own.